Bipolar cord for use with the VirtuoSaph Endoscopic Vein… recallZ-1230-2011 · Terumo Cardiovascular Systems Corporation · initiated Jan 10, 2011

Terminated All recalls posted in February 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Terumo Cardiovascular Systems Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Bipolar cord product specification indicates the product has a service life of 50 uses. However the IFU does not specify an end of service life. The firm has recieved 46 complaints on the Bipolar cord. The firm investigated and found that the units involved in the complaints were often over 12 months old. The firm believes the units involved in the complaints were beyond their product life

FDA root cause category: Other

What the company is doing

Terumo Cardiovascular Systems Corporation sent an URGENT MEDICAL DEVICE CORRECTION letter dated January 10, 2011, to Operating room managers. The letter informed the customers of the Bipolar cord product life of 50 uses. The letter states that the IFU for the bipolar cord is being updated to include the information regarding the limit of 50 uses and sterlization cycles. The letter recommends that user keep a sterile back up bi-polar cord in a convienent place during endoscopic vein harvesting procedures. The firm provided a tracking sheet in each package to help users track the number of sterlization cycles for the cord. Customers were instructed to review the Medical Device Correction notice, assure that all users were aware of the notice, and confirm receipt of the communicaiton by faxing the attached Customer Response Form to the fax number indicated on the form. For questions regarding this recall call Terumo CVS Customer Service at 1-800-521-2818.

Recall details

Recall number
Z-1230-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jan 10, 2011
Posted by the FDA
Feb 14, 2011
Recalling firm
Terumo Cardiovascular Systems Corporation
Quantity
2782 cords
Product code
GCJ
FDA submissions
K083194
Distribution
Worldwide Distribution - United States including AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, , NY, MO, MS, MT, NC, NE, NH, NJ, NM, OH, OK, OR, PA, SC, TN, TX, VA, WI, AND WV and the countries of Chile, Hong Kong, Malaysia, United Arab Emirates, Belgium, Japan, Korea, The Phillipines, Singapore, Taiwan, Thailand, Germany, and United Kingdom.

Product description

Bipolar cord for use with the VirtuoSaph Endoscopic Vein Harvesting System, MCBICORD1, Terumo Cardiovascular System, Ann Arbor, MI. The Terumo Bipolar cord MCBICORD1 is a flexible, reusable power cable that is indicated for use with the Terumo VirtuoSaph Endoscopic Vein Harvesting System and a compatible generator during saphenous vein harvesting procedures.

Code information (models, lots, serial numbers)

01K, 02K, 03K, 04K, 05K, 06K, 07K, 08K, 09K, 0XK, 4ZK, 51K, 52K, 53K, 54K, 55K, 6XK, 6ZK, 71K, 72K, 73K, 74K, 75K, 76K, 77K, 78K, 79K, 7XK, 7YK, 7ZK, 81K, 82K, 83K, 84K, 85K, 86K, 87K, 87K, 88K, 89K, 8XK, 8YK, 8ZK, 91K, 95K, 96K, 97K, 99K, 9XK, 9YK and 9ZK.

Other devices under the same FDA submission

4 devices cleared under K083194 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.