5) Signa HDx (K052293 Signa HDx and HDxt MR Systems) recallZ-1428-2011 · GE Healthcare, LLC · initiated Nov 9, 2010

Terminated All recalls posted in February 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety. If the gradient cables are inadvertently swapped while servicing the gradients, and the Service Engineer does not perform the geometry check as required in the service procedure, images m

FDA root cause category: Nonconforming Material/Component

What the company is doing

On November 9, 2010, the firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter dated November 5, 2010 to all Consignees/Customers. The letter described the product, problem and action to be taken by the firm. GE instructed the customers to contact their local GEHC support Engineer, if they have any questions about the image orientation, or the procedures for geometry checks on thier system. Additionally, a GE Healthcare representative will install a device, which will reduce the potiential for the gradient cables to be swapped. Contact your local GEHC Support Engineer if you have any questions concerning this notification or phone: 262-548-2731.

Recall details

Recall number
Z-1428-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 9, 2010
Posted by the FDA
Feb 24, 2011
Recalling firm
GE Healthcare, LLC
Quantity
2417
Product code
LNH
FDA submissions
K052293
Distribution
Worldwide distribution: USA including DC and PR; and countries including: VIET NAM, VENEZUELA, URUGUAY, UNITED KINGDOM, UNITED ARAB EMIRATES, UKRAINE, TURKEY, TUNISIA, FORMER YUGOSLAV REPUBLIC OF MACEDONIA, THAILAND, TAIWAN, SYRIAN ARAB REPUBLIC, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SAUDI ARABIA, RUSSIAN FEDERATION, ROMANIA, REUNION, REPUBLIC OF KOREA, PORTUGAL POLAND, PHILIPPINES, PERU,PARAGUAY, PANAMA, PAKISTAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MEXICO, MAURITIUS, MARTINIQUE,MALTA, MALAYSIA, LUXEMBOURG,LITHUANIA, LIBYAN ARAB JAMAHIRIYA, LEBANON, LATVIA, KUWAIT, KENYA, KAZAKHSTAN, JORDAN, JAPAN, JAMAICA, ITALY, ISRAEL, ISLAMIC REPUBLIC OF IRAN, IRELAND, INDONESIA, INDIA, ICELAND, HUNGARY, HONG KONG, HONDURAS, GUYANA, GUATEMALA, GUADELOUPE, GREECE, GERMANY, GEORGIA, FRENCH POLYNESIA, FRANCE, FINLAND, EL SALVADOR, EGYPT, ECUADOR, DOMINICAN REPUBLIC, DENMARK, CECH REPUBLIC, CYPRUS, CROATIA, COLOMBIA, CHINA, CHILE, CANDA, BULGARIA, BRAZIL, BOSNIA AND HERZEGOVINA, BOLIVIA, BELGIUM, BELARUS, AZERBAIJAN, AUSTRIA, AUSTRALIA, ARGENTINA, ALGERIA.

Product description

5) Signa HDx (K052293 Signa HDx and HDxt MR Systems) Signa HDx or Signa HDxt (K052293 Signa HDx and HDxt MR Systems) Signa HDxt (K052293 Signa HDx and HDxt MR Systems ) The GE Signa HDe MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa HDe MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa HDe system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Code information (models, lots, serial numbers)

Show the full code information (36k characters)

SERIAL  000000HDMR3855 00000290191MR6 21 000000HDMR5594 000000HDMR3972 000000HDMR5909 00000HDMR10275 000000HDMR4367 000000HDMR2622 1568421 00000288797MR4 000000HDMR2901 000000HDMR2799 00000011875YR2 00000289023MR4 000000HDMR4087 000000HDMR5809 000000HDMR6277 00000HDMR27758 000000HDMR5774 000000HDMR3407 00000DDDM10670 000000HDMR3649 00000289786MR6 000000HDMR3464 00000GE0743594 00000HDMR20044 00000HDMR20171 000000HDMR5401 000000HDMR3861 000000HDMR1063 000000HDMR3724 00000011927YR1 00000280253MR6 00000HDMR10554 000000HDMR4015 00000HDMR10788 00000HDMR20157 000000HDMR3261 00000HDMR10459 000000HDMR3470 000000HDMR3330 000000HDMR2936 00000011926YR3 00000HDMR20298 000000HDMR3010 000000HDMR5633 000000HDMR2377 000000HDMR3008 000000HDMR3391 000000HDMR5224 000000HDMR2924 000000HDMR6012 000000HDMR3844 00000HDML10300 000000HDMR3679 00000011571YR7 000000HDMR5228 00000012239YR0 000000HDMR2531 00000012078YR2 000000HDMR1148 00000011867YR9 00000011595YR6 000000HDMR3885 000000HDMR2962 000000HDMR4242 000000HDMR3150 00000280251MR0 000000HDMR3141 00000011786YR1 00000011887YR7 000000HDMR4062 000000HDMR6263 000000HDMR3152 000000HDMR3811 00000HDMR20219 00000HDMR20258 00000HDMR10630 00000012241YR6 000000HDMR5593 00000278041MR9 000000HDMR3514 000000HDMR5765 000000HDMR3513 000000HDMR5956 000000HDMR2590 000000HDMR2892 00000HDMR20018 00000HDMR10743 00000HDMR11028 000000HDMR2967 00000011746YR5 000000HDMR6031 00000277916MR3 000000HDMR4299 000000HDMR3987 00000HDMR20256 00000012180YR6 00000011586YR5 000000HDMR6029 000000HDMR6292 00000011878YR6 000000HDMR3529 000000HDMR3142 000000HDMR2508 00000011502YR2 000000HDMR4149 000000HDMR4106 00000HDMR20142 00000011742YR4 00000HDMR20050 000000HDMR4154 00000011588YR1 000000HDMR5584 000000HDMR5844 000000HDMR4081 000000HDMR3648 00000012042YR8 000000HDMR3329 000000HDMR3248 000000HDMR3755 000000HDMR3766 000000HDMR3979 000000HDMR3633 00000HDMR20151 000000HDMR4204 000000HDMR3916 000000HDMR3262 00000011798YR6 000000HDMR3810 000000HDMR5525 000000HDMR2515 00000012163YR2 000000HDMR2653 00000275647MR6 000000HDMR2615 00000011579YR0 000000HDMR3732 00000011596YR4 000000HDMR2641 000000HDMR4293 000000HDMR5605 000000HDMR5442 000000HDMR6311 000000HDMR3681 000000HDMR6297 00000HDMR10722 000000HDMR3543 000000HDMR2852 000000HDMR3099 000000HDMR2998 000000HDMR2328 000000HDMR5941 000000HDMR3108 000000HDMR3190 00000HDMR20276 00000HDMR20065 00000HDMR20053 000000HDMR6080 00000HDMR10944 00000HDMR10622 000000HDMR3909 000000HDMR5486 000000HDMR2451 000000HDMR4021 000000HDMR5949 000000HDMR5846 000000HDMR2463 000000HDMR4379 000000HDMR5532 000000HDMR2691 000000HDMR2556 000000HDMR3243 00000011938YR8 00000012196YR2 000000HDMR2399 00000286785MR1 000000HDMR4064 000000HDMR2588 000000HDMR3777 1516299 00000HDMR10891 000000HDMR2482 00000012050YR1 000000HDMR2741 00000012008YR9 00000012229YR1 000000HDMR2562 000000HDMR4040 157024 00000HDMR20070 000000HDMR3273 000000HDMR3401 000000HDMR3823 000000HDMR2949 00000HDMR20214 000000HDMR5855 00000011774YR7 000000HDMR1151 11797 00000277534MR4 00000HDMR10291 000000HDMR3155 000000HDMR3609 00000278851MR1 00000012222YR6 000000HDMR5735 000000HDMR5396 00000011584YR0 00000289395MR6 000000HDMR5273 000000HDMR3847 000000HDMR3265 000000HDMR2460 000000HDMR2495 000000HDMR3396 000000HDMR4102 1578896 000000HDMR4027 1583350 000000HDMR3806 000000HDMR3115 00000011724YR2 00000011950YR3 000000HDMR3620 1612018 00000011784YR6 00000HDMR10398 000000HDMR3966 000000HDMR5573 000000HDMR2903 000000HDMR3799 000000HDMR3798 000000HDMR3953 000000HDMR2855 000000HDMR2596 000000HDMR4249 00000011897YR6 000000HDMR3096 00000012020YR4 000000HDMR6325 000000HDMR3833 00000HDMR10901 000000HDMR3059 000000HDMR4090 00000012158YR2 000000HDMR2693 000000HDMR2651 000000HDMR4305 000000HDMR5503 00000011920YR6 00000273822MR7 000000HDMR2365 000000HDMR2375 000000HDMR3534 000000HDMR5255 000000HDMR3270 00000011900YR8 000000HDMR5472 00000HDMR10482 000000HDMR3748 000000HDMR3209 000000HDMR3927 000000HDMR3678 000000HDMR2794 000000HDMR4006 00000288608MR3 000000HDMR5352 000000HDMR3917 00000012244YR0 00000HDMR10832 00000HDMR10625 00000HDMR10334 000000HDMR5739 000000HDMR2879 000000HDMR5373 00000HDMR10548 000000HDMR5237 00000273632MR0 00000276045MR2 000000HDMR3590 00000HDMR10212 000000HDMR3232 000000HDMR3257 000000HDMR2765 000000HDMR5313 000000HDMR3669 00000011958YR6 00000012236YR6 000000HDMR2346 000000HDMR3421 00000HDMR20235 00000011892YR7 000000HDMR4303 000000HDMR5872 00000HDMR20224 000000HDMR4130 00000011885YR1 000000HDMR6183 000000HDMR4300 00000HDMR10739 000000HDMR5988 1 00000012063YR4 000000HDMR3721 000000HDMR2993 00000273127MR1 000000HDMR3751 000000HDMR3682 000000HDMR5498 000000HDMR3607 000000HDMR5837 00000012181YR4 00000HDMR20128 000000HDMR3088 000000HDMR3462 000000HDMR3292 000000HDMR5885 1543949 000000HDMR4181 00000011901YR6 00000HDMR10489 00000HDMR10846 000000HDMR5362 00000HDMR10981 00000HDMR20084 00000HDMR10986 000000HDMR5397 00000HDMR20059 00000011941YR2 000000HDMR5583 000000HDMR2742 000000HDMR2684 00000HDMR10297 000000HDMR3765 000000HDMR5664 000000HDMR5398 000000HDMR4255 3767 00000012240YR8 00000011953YR7 00000276245MR8 000000HDMR3130 000000HDMR2726 00000HDMR20069 000000HDMR4121 000000HDMR5465 000000HDMR3037 000000HDMR3782 000000HDMR3996 000000HDMR3947 000000HDMR5685 000000HDMR3224 000000HDMR2880 000000HDMR3160 000000HDMR4332 000000HDMR3639 00000011791YR1 00000012189YR7 000000HDMR3029 000000HDMR5721 00000HDMR20242 000000HDMR5318 00000HDMR10440 000000HDMR2560 00000278391MR8 000000HDMR3255 000000HDMR2822 00000HDMR20071 000000HDMR4061 000000HDMR2775 000000HDMR3355 000000HDMR2448 12345 000000HDMR1086 000000HDMR2609 00000287596MR1 000000HDMR4046 00000011787YR9 000000HDMR2904 00000012162YR4 000000HDMR2817 000000HDMR2660 000000HDMR5502 000000HDMR4114 000000HDMR5329 000000HDMR2692 00000280000MR1 000000HDMR5609 000000HDMR2871 00000VALUE1083 00000HDMR20046 000000HDMR2813 000000HDMR3848 000000HDMR2465 00000011886YR9 000000HDMR5759 000000HDMR5761 000000HDMR2397 000000HDMR2523 00000012080YR8 000000HDMR5323 00000HDMR20073 1537489 00000HDMR20289 00000012023YR8 000000HDMR6134 000000HDMR4306 000000HDMR3933 00000HDMR20279 00000HDMR10789 000000HDMR4012 00000HDMR20248 00000011589YR9 00000011581YR6 00000HDMR10559 000000HDMR4036 00000HDMR10758 000000HDMR3440 000000HDMR5999 000000HDMR2623 000000HDMR3183 000000HDMR5322 000000HDMR3729 000000HDMR2756 00000280775MR8 000000HDMR5343 000000HDMR3908 000000HDMR4056 000000HDMR3677 000000HDMR3952 000000HDMR3659 000000HDMR5453 000000HDMR3910 000000HDMR5644 000000HDMR5779 0 000000HDMR3200 000000HDMR3161 000000HDMR5623 00000012133YR5 000000HDMR3092 000000HDMR2517 00000012230YR9 000000HDMR3109 00000290130MR4 000000HDMR2827 000000HDMR4165 1555378 000000HDMR4292 000000HDMR2977 000000HDMR2804 1541119 000000HDMR5628 000000HDMR3466 00000HDMR10351 000000HDMR3646 00000290141MR1 000000HDMR4376 000000HDMR3272 000000HDMR5654 000000HDMR5907 000000HDMR6283 00000HDMR10719 000000HDMR2504 000000HDMR5556 000000HDMR5206 00000011934YR7 000000HDMR4362 000000HDMR4140 000000HDMR3519 000000HDMR6122 000000HDMR4132 00000HDMR10263 000000HDMR3712 000000HDMR2795 000000HDMR6171 000000HDMR3943 2102 000000HDMR3299 000000HDMR4216 000000HDMR4355 000000HDMR3568 00000HDMR10557 000000HDMR4241 00000HDMR20172 00000HDMR20293 000000HDMR3218 00000012003YR0 000000HDMR3263 000000HDMR5657 000000HDMR3517 00000HDMR10822 00000HDMR10979 00000HDMR20047 000000HDMR2592 000000HDMR5697 000000HDMR2568 00000012224YR2 000000HDMR2548 000000HDMR5575 000000HDMR5440 00000HDMR10475 000000HDMR4295 00000HDMR20223 000000HDMR5673 00000011591YR5 000000HDMR5402 000000HDMR5798 00000HDMR20036 000000HDMR3791 000000HDMR5363 000000HDMR4386 000000HDMR4323 00000289347MR7 000000HDMR3050 000000HDMR2981 00000HDMR10853 000000HDMR2987 000000HDMR2639 000000HDMR2758 000000HDMR3375 000000HDMR2789 000000HDMR2524 00000276266MR4 00000011946YR1 000000HDMR3560 000000HDMR3942 000000HDMR3509 000000HDMR3635 00000012232YR5 000000HDMR2890 000000HDMR3762 000000HDMR4253 000000HDMR3280 000000HDMR5492 000000HDMR2773 000000HDMR3558 000000HDMR3337 000000HDMR4347 00000011884YR4 000000HDMR2873 1555399 00000011578YR2 000000HDMR2797 000000HDMR4353 00000011585YR7 00000011567YR3 00000008550YR6 000000HDMR4007 000000HDMR4155 000000HDMR4177 000000HDMR5538 000000HDMR5388 00000HDMR20129 00000275967MR8 00000HDMR20237 000000HDMR6116 000000HDMR5998 00000HDMR10356 000000HDMR6020 000000HDMR2751 000000HDMR2948 00000HDMR20291 000000HDMR3753 000000HDMR3316 000000HDMR5334 00000HDMR20269 00000HDMR10841 000000HDMR4272 00000HDMR20189 00000HDMR10479 000000HDMR6265 000000HDMR2418 000000HDMR2755 000000HDMR2695 000000HDMR5536 000000HDMR3801 000000HDMR6024 000000HDMR6023 00000279415MR4 00000011883YR6 000000HDMR5679 000000HDMR3372 000000HDMR5245 000000HDMR5375 000000HDMR5598 000000HDMR2710 00000011711YR9 000000HDMR4337 000000HDMR1001 00000011721YR8 00000011801YR8 000000HDMR3362 000000HDMR3199 000000HDMR5945 000000HDMR1160 000000HDMR5244 000000HDMR2824 000000HDMR5830 000000HDMR3824 00000012065YR9 00000012188YR9 1556043 00000275648MR4 000000HDMR5424 00000012047YR7 000000HDMR3000 000000HDMR5422 00000012223YR4 00000012225YR9 000000HDMR5827 000000HDMR2958 000000HDMR3459 00000HDMR20178 00000HDMR10880 00000HDMR20089 000000HDMR5274 000000HDMR5932 00000012017YR0 000000HDMR4311 000000HDMR6001 000000HDMR3783 000000HDMR5905 000000HDMR3238 000000HDMR5272 000000HDMR5236 00000011882YR8 00000HDMR10886 000000HDMR3293 00000HDMR10793 00000HDMR20099 000000HDMR6145 000000HDMR5285 000000HDMR5285 000000HDMR3207 000000HDMR2553 000000HDMR2917 000000HDMR3254 000000HDMR5513 000000HDMR5777 000000HDMR4050 000000HDMR5931 000000HDMR5651 000000HDMR5303 00000HDMR20205 00000HDMR20292 000000HDMR5546 00000012193YR9 000000HDMR6222 000000HDMR2100 00000HDMR19996 00000012167YR3 000000HDMR5308 000000HDMR3978 000000HDMR3851 000000HDMR2395 000000HDMR5268 00000285490MR9 00000012198YR8 00000286587MR1 000000HDMR3284 00000012185YR5 00000HDMR10967 00000012157YR4 000000HDMR2847 000000HDMR3825 00000001206YR5 000000HDMR3545 000000HDMR3948 00000288342MR9 000000HDMR4335 000000HDMR3524 000000HDMR5600 1574694 000000HDMR4151 000000HDMR3342 000000HDMR3719 000000HDMR6155 000000HDMR3672 000000HDMR5958 000000HDMR3998 00000HDMR10483 00000HDMR10574 000000HDMR4115 000000HDMR4123 000000HDMR2921 00000011956YR0 00000012161YR6 00000280328MR6 000000HDMR5215 00000011796YR0 00000HDMR55647 000000HDMR3900 00000012192YR1 000000HDMR5707 000000HDMR2313 000000HDMR6302 000000HDMR4333 000000HDMR6211 000000HDMR4108 00000011743YRZ 000000HDMR5457 000000HDMR3350 000000HDMR3080 000000HDMR1162 00000274744MR2 000000HDMR3887 00000012132YR7 00000012070YR9 000000HDMR5299 00000011931YR3 00000012231YR7 000000HDMR3549 000000HDMR2572 00000280001MR9 000000HDMR2600 000000HDMR1110 000000HDMR3331 000000HDMR5789 00000HDMR20022 000000HDMR3619 000000HDMR3333 1515938 00000HDMR10862 000000HDMR5324 000000HDMR3937 000000HDMR4317 000000HDMR3436 000000HDMR4158 1586023 00000287391MR7 00000012006YR3 000000HDMR4172 000000HDMR3645 000000HDMR3623 1589432 000000HDMR5710 000000HDMR5627 000000HDMR5580 000000HDMR4289 57413 000000HDMR1077 00000011955YR2 000000HDMR4095 00000HDMR10969 00000011717YR6 000000HDMR5327 000000HDMR6123 00000011551YR9 000000HDMR4182 00000HDMR10343 000000HDMR3452 00000011804YR2 000000HDMR4039 00000289265MR1 000000HDMR3354 00000277444MR6 000000HDMR3713 000000HDMR2484 000000HDMR3875 000000HDMR4183 000000HDMR4372 000000HDMR2927 000000HDMR3145 000000HDMR4157 00000280672MR7 00000HDMR10835 00000287441MR0 00000278393MR4 000000HDMR3604 00000280690MR9 00000280578MR6 00000HDMR10930 00000HDMR10289 000000HDMR3412 00000278312MR4 00000HDMR10401 00000HDMR10710 1541135 00000279870MR0 000000HDMR3717 000000HDMR4197 00000HDMR10938 000000HDMR6063 00000HDMR10799 000000HDMR5746 000000HDMR5883 000000HDMR1178 000000HDMR2932 000000HDMR3707 000000HDMR6113 000000HDMR3860 00000286585MR5 00000277641MR7 00000HDMR20158 000000HDMR3715 000000HDMR4268 00000294510MR3 000000HDMR3137 00000287389MR1 000000HDMR5693 000000HDMR3958 000000HDMR5768 000000HDMR4076 000000HDMR2997 000000HDMR6045 000000HDMR3334 000000HDMR3488 000000HDMR3794 000000HDMR5634 000000HDMR5716 000000HDMR5292 000000HDMR5433 000000HDMR6166 000000HDMR5784 000000HDMR6288 00000012059YR2 1531438 000000HDMR5387 000000HDMR2497 000000HDMR2606 000000HDMR6190 000000HDMR6069 000000HDMR5372 000000HDMR2587 00000HDMR10724 000000HDMR2825 000000HDMR5590 000000HDMR3003 00000027810MR1 000000HDMR2782 00000HDMR10978 000000HDMR2697 00000289212MR3 000000HDMR3745 000000HDMR3644 000000HDMR6189 000000HDMR3235 000000HDMR5881 000000HDMR3181 00000280781MR6 000000HDMR5393 000000HDMR3542 000000HDMR5714 000000HDMR5581 00000VALUE1177 000000HDMR5968 00000280327MR8 00000287847MR8 000000HDMR5734 00000011503YR0 000000HDMR4360 00000290128MR8 00000HDMR10462 00000HDMR20020 00000012049YR3 000000HDMR2875 000000HDMR3208 00000286687MR9 000000HDMR3936 000000HDMR3768 000000HDMR5595 000000HDMR6148 000000HDMR3367 00000HDMR20109 000000HDMR6193 000000HDMR5971 000000HDMR2776 000000HDMR3973 000000HDMR2989 000000HDMR3643 00000289314MR7 00000278549MR1 000000HDMR1138 000000HDMR4263 000000HDMR5976 000000HDMR3840 000000HDMR6293 000000HDMR2956 00000011559YR2 00000288387MR4 000000HDMR3491 00000280923MR4 000000HDMR3320 000000HDMR5894 000000HDMR3443 000000HDMR4358 000000HDMR5353 000000HDMR5676 00000HDMR10474 000000HDMR2642 000000HDMR2471 000000HDMR1154 000000HDMR3651 000000HDMR3476 000000HDMR3456 000000HDMR4161 000000HDMR4287 00000HDMR10493 000000HDMR5669 000000HDMR5799 00000011712YR7 000000HDMR5992 000000HDMR5719 000000HDMR5728 000000HDMR4205 000000HDMR4018 000000HDMR5366 000000HDMR3376 000000HDMR3419 000000HDMR5856 000000HDMR2792 000000HDMR4399 000000HDMR4315 00000VALUE1166 000000HDMR6087 00000280778MR2 00000282865MR5 00000011733YR3 000000HDMR3191 000000HDMR3356 000000HDMR6114 000000HDMR2930 000000HDMR2872 000000HDMR3818 000000HDMR2477 00000HDMR20034 000000HDMR5914 00000276265MR6 00000282383MR9 00000HDMR20149 000000HDMR3789 000000HDMR3179 000000HDMR2329 000000HDMR3018 000000HDMR5420 000000HDMR4254 00000011779YR6 000000HDMR3838 000000HDMR2637 000000HDMR1130 000000HDMR3660 000000HDMR2521 00000289914MR4 000000HDMR3097 000000HDMR3685 00000284710MR1 00000278389MR2 00000283743MR3 000000HDMR5962 00000290194MR0 000000HDMR2321 00000288823MR8 000000HDMR2718 00000HDMR10659 00000279563MR1 000000HDMR3341 00000VALUE1294 00000VALUE1281 000000HDMR3515 000000HDMR4237 000000HDMR3920 000000HDMR5277 000000HDMR3188 000000HDMR3868 000000HDMR3506 000000HDMR3406 000000HDMR3867 000000HDMR3566 000000HDMR6093 000000HDMR5630 000000HDMR5825 000000HDMR3932 000000HDMR3399 000000HDMR3361 000000HDMR2519 000000HDMR3773 000000HDMR5280 00000011737YR4 00000HDMR10867 000000HDMR4167 000000HDMR5982 000000HDMR1145 000000HDMR4334 000000HDMR5840 000000HDMR5775 000000HDMR3647 000000HDMR5425 00000276244MR1 00000HDMR10771 00000HDMR10623 000000HDMR3598 00000278801MR6 00000278556MR6 00000281375MR6 000000HDMR6356 00000280366MR6 000000HDMR2366 000000HDMR2959 000000HDMR2616 000000HDMR1183 000000HDMR3223 000000HDMR1157 1524447 00000278674MR7 000000HDMR5978 000000HDMR3016 000000HDMR3862 000000HDMR2452 000000HDMR3930 000000HDMR3744 00000287269MR5 000000HDMR4035 000000HDMR2851 000000HDMR3556 000000HDMR4345 000000HDMR4365 000000HDMR6096 000000HDMR5851 000000HDMR6083 00000288160MR5 000000HDMR6191 000000HDMR4145 00000011729YR1 00000011557YR6 000000HDMR6346 000000HDMR3180 000000HDMR5646 00000275968MR6 00000HDMR10838 00000278397MR5 00000HDMR10647 000000HDMR2525 00000HDMR20162 00000281341MR8 000000HDMR2624 000000HDMR2808 000000HDMR3688 000000HDMR4384 00000276434MR8 00000HDMR10421 000000HDMR6336 00000HDMR10714 00000289259MR4 00000289191MR9 00000289040MR8 00000288875MR8 00000289655MR3 000000HDMR3135 00000HDMR10813 00000HDMR20160 00000288359MR3 000000HDMR2779 000000HDMR2331 00000282550MR3 000000HDMR6224 000000HDMR3336 000000HDMR4144 000000HDMR4134 000000HDMR4104 000000HDMR4393 000000HDMR2570 00000278799MR2 000000HDMR2823 00000288711MR5 00000284493MR4 000000HDMR2772 000000HDMR1107 00000011576YR6 000000HDMR2833 000000HDMR2621 00000011522YR0 000000HDMR5900 000000HDMR3271 00000278103MR7 1512388 000000HDMR3389 000000HDMR4129 00000288437MR7 00000288358MR5 000000HDMR2582 00000012069YR1 000000HDMR3297 1503843 000000HDMR3555 000000HDMR2483 000000HDMR2887 00000HDMR10392 00000VALUE1181 000000HDMR2845 00000289821MR1 00000HDMR10819 000000HDMR5776 000000HDMR4243 000000HDMR2969 000000HDMR3788 000000HDMR5234 00000HDMR20156 00000HDMR10877 000000HDMR5549 00000HDMR10878 000000HDMR6251 00000HDMR10201 00000HDMR20270 000000HDMR4143 00000HDMR20027 000000HDMR5700 00000011930YR5 00000012128YR5 00000012051YR9 00000011933YR9 00000011924YR8 00000011893YR5 000000HDMR3965 00000012159YR0 00000012120YR2 00000276459MR5 000000HDMR3757 00000296147MR2 00000288821MR2 00000280858MR2 00000286601MR0 00000011935YR4 00000012166YR5 00000HDMR10354 00000HDMR10273 00000HDMR10642 00000VALUE1176 000000HDMR2608 000000HDMR2738 000000HDMR2354 00000284686MR3 00000290255MR9 00000011567YR5 000000HDMR3891 000000HDMR3127 000000HDMR2931 00000011594YR9 000000HDMR3242 000000HDMR2787 000000HDMR5286 000000HDMR4048 00000289316MR2 000000HDMR5696 000000HDMR3157 000000HDMR4001 1528473 000000HDMR3599 000000HDMR3028 00000012044YR4 000000HDMR6119 000000HDMR5428 000000HDMR6280 000000HDMR2990 000000HDMR3076 000000HDMR6212 000000HDMR3993 000000HDMR4336 000000HDMR2835 000000HDMR2544 000000HDMR3444 000000HDMR2576 000000HDMR2862 00000278105MR2 000000HDMR6082 000000HDMR5933 000000HDMR3380 00000HDMR20218 00000284648MR3 000000HDMR3125 000000HDMR5902 00000HDMR10327 000000HDMR5980 000000HDMR3319 000000HDMR2664 000000HDMR2713 000000HDMR2512 00000275538MR7 000000HDMR2882 000000HDMR3121 000000HDMR3826 000000HDMR2676 000000HDMR2766 000000HDMR2902 000000HDMR3621 000000HDMR5252 000000HDMR6015 000000HDMR5928 000000HDMR5951 000000HDMR3040 000000HDMR2541 00000HDMR10427 00000284688MR9 000000HDMR2884 000000HDMR5379 00000279616MR7 00000HDMR10535 000000HDMR4025 000000HDMR3415 00000HDMR20167 000000HDMR6004 000000HDMR3940 000000HDMR3122 00000HDMR20012 000000HDMR3522 000000HDMR3779 00000275168MR3 000000HDMR4349 000000HDMR5917 000000HDMR5336 00000283350MR7 000000HDMR2712 00000HDMR10419 00000HDMR10460 000000HDMR2351 000000HDMRXXXX 1542144 000000HDMR3866 000000HDMR5462 000000HDMR5887 000000HDMR5954 000000HDMR2459 00000276163MR3 00000280777MR4 000000HDMR2677 000000HDMR4304 000000HDMR4044 000000HDMR3301 00000HDMR10998 000000HDMR5915 00000HDMR10094 000000HDMR4071 00000HDMR20117 00000HDMR10780 000000HDMR2747 00000HDMR10678 00000HDMR10417 00000279461MR8 000000HDMR4342 000000HDMR5622 00000HDMR20290 000000HDMR3652 000000HDMR5668 000000HDMR3858 000000HDMR5528 000000HDMR5413 000000HDMR2785 000000HDMR5895 00000012165YR7 00000012186YR3 00000HDMR10281 000000HDMR5530 00000011894YR3 00000012183YR0 00000011997YR4 000000HDMR5382 00000VALUE1162 000000HDMR4361 00000012184YR8 00000011590YR7 000000HDMR5834 00000011952YR9 000000HDMR5452 00000011866YR1 000000HDMR5477 00000011928YR9 000000HDMR3637 000000HDMR3872 000000HDMR5356 00000011574YR1 00000011871YR1 000000HDMR6160 000000HDMR3086 000000HDMR6256 000000HDMR6030 000000HDMR5330 000000HDMR5474 00000011876YR0 00000012190YR5 00000VALUE1280 00000VALUE1179 00000012125YR1 000000HDMR6153 000000HDMR5783 000000HDMR5342 000000HDMR5504 000000HDMR3612 000000HDMR5242 000000HDMR5599 000000HDMR3373 000000HDMR2348 00000HDMR10221 000000HDMR4070 000000HDMR2905 000000HDMR1111 000000HDMR3849 000000HDMR3295 00000HDMR10260 000000HDMR2675 00000HDMR10827 000000HDMR4184 00000011861YR2 000000HDMR3563 00000011889YR3 000000HDMR2745 00000HDMR10368 000000HDMR3730 000000HDMR5419 000000HDMR5222 000000HDMR3905 00000011944YR6 000000HDMR4074 00000278606MR9 00000HDMR20272 000000HDMR3734 000000HDMR2549 00000277779MR5 000000HDMR2492 000000HDMR3305 00000HDMR10477 000000HDMR5929 000000HDMR4245 00000HDMR20134 00000HDMR10216 00000280921MR8 000000HDMR2513 000000HDMR3143 000000HDMR3489 00000HDMR10528 000000HDMR3393 00000HDMR10870 00000280673MR5 00000HDMR10943 00000HDMR10888 000000HDMR3212 000000HDMR3800 00000011790YR3 000000HDMR6007 000000HDMR4301 000000HDMR6308 000000HDMR5737 000000HDMR4135 00000012009YR7 000000HDMR4324 000000HDMR5432 000000HDMR6192 000000HDMR5795 000000HDMR5868 00000011898YR4 000000HDMR6124 000000HDMR5854 00000011598YR0 00000011599YR8 00000011895YR0 00000277642MR5 000000HDMR3981 000000HDMR5561 000000HDMR3907 000000HDMR5560 00000288883MR2 00000287655MR5 000000HDMR5973 00000011580YR8 000000HDMR6139 00000283517MR1 000000HDMR2891 000000HDMR6272 000000HDMR4093 000000HDMR3955 00000012005YR5 000000HDMR4173 000000HDMR3070 000000HDMR2614 000000HDMR2796 000000HDMR3068 00000011896YR8 000000HDMR2889 000000HDMR3286 000000HDMR3631 000000HDMR2438 00000276184MR9 000000HDMR3539 000000HDMR5404 00000HDMR20181 000000HDMR2699 000000HDMR2324 000000HDMR2441 000000HDMR3837 000000HDMR5213 000000HDMR5448 000000HDMR4256 000000HDMR3001 00000HDMR20228 000000HDMR5483 000000HDMR4222 00000012019YR6 00000290369MR8 000000HDMR1005 000000HDMR6142 000000HDMR5358 00000010726YR8 00000294274MR6 00000009568YR7 00000294106MR0 00000290690MR7 00000296303MR1 00000290326MR8 00000295233MR1 00000008539YR9 00000007871YR7 00000294554MR1 00000005992YR3 00000293343MR0 00000008349YR3 00000007556YR4 00000010294YR7 00000007498YR9 00000007878YR2 00000007565YR5 00000007877YR4 00000007100YR1 00000290641MR0 00000006740YR5 00000007097YR9 00000293937MR9 00000943585YMO 00000294102MR9 00000009462YR3 00000295058MR2 00000006736YR3 00000292262MR3 00000292636MR8 00000295218MR2 00000291359MR8 00000292466MR0 00000295593MR8 00000009849YR1 00000293631MR8 00000296113MR4 00000006713YR2 00000008396YR4 00000009038YR1 00000292520MR4 00000292908MR1 00000292980MR0 00000008352YR7 00000011140YR1 00000006707YR4 00000011156YR7 00000010973YR6 00000011055YR1 00000007582YR0 00000007501YR0 00000007860YR0 00000009057YR1 00000007882YR4 00000007887YR3 00000010511YR4 00000006558YR1 00000008197YR6 00000297488MR9 00000293218MR4 00000011011YR4 00000007087YR0 00000008379YP0 00000297698MR3 00000008387YR3 00000008342YR8 00000293432MR1 00000295743MR9 00000006735YR5 00000006700YR9 00000006675YR3 00000005564YRO 00000296144MR9 00000005547YR5 00000007133YR2 00000005537YR6 00000006737YR1 00000005232YR4 00000006757YR9 00000005564YR0 00000009052YR2 00000008997YR9 00000006731YR4 00000006724YR9 00000293007MR1 00000007532YR5 00000007858YR4 00000007856YR8 00000295025MR1 00000009834YR3 00000010599YR9 00000008338YR6 00000295027MR7 00000011057YR7 00000005927YR5 00000007549YR9 00000293316MR6 00000007122YR5 00000006696YR9 00000292823MR2 00000010435YR6 00000291805MR0 00000009754YR3 00000295004MR6 00000006666YR2 00000297419MR4 00000007866YR7 00000007865YR9 00000029785MR6 00000293670MR6 00000009986YR1 00000009603YR2 00000009766YR7 00000294015MR3 00000295512MR8 00000296104MR3 00000006600YR1 00000006727YR2 00000295518MR1 00000029787MR9 00000009853YR3 00000007544YR0 00000293712MR6 00000006572YR2 00000006673YR8 00000006571YR4 00000008332YR9 00000005974YR1 00000006669YR6 00000293658MR1 00000010722YR7 00000290655MR0 00000292792MR9 00000007585YR3 00000293324MR0 00000001060YR5 00000011089YR0 00000010617YR9 00000010596YR5 00000008378YR2 00000008391YR5 00000011064YR3 00000005216YR7 00000290735MR0 00000293657MR2 00000295741MR3 00000009007YR6 00000005536YR8 00000007572YR1 00000009599YR2 00000007862YR6 00000011117YR9 00000007507YR7 00000006570YR6 00000006708YR2 00000295235MR6 00000011126YR0 00000007591YR1 00000005976YR6 00000011339YR9 00000008335YR2 00000009438YR3 00000293614MR4 00000294144MR1 00000294990MR7 00000007579YR6 00000294667MR1 00000008061YR4 00000006667YR0 00000011217YR7 00000011086YR6 00000008541YR5 00000295229MR9 00000009078YR7 00000292820MR8 00000006681YR1 00000006556YR5 00000006568YR0 00000297700MR7 00000008359YR2 00000009571YR1 00000006743YR9 00000006733YR0 00000007884YR0 00000011190YR6 00000295789MR2 00000011224YR3 00000008380YR8 00000008385YR7 00000006703YR3 00000007116YR7 00000010167YR5 00000292290MR4 00000007519YRO 00000293315MR8 00000009062YR1 00000007109YR2 00000009771YR7 00000006705YR8 00000009031YR6 00000295899MR9 00000007864YR2 00000007863YR4 00000005990YR7 00000010805YR0 00000008372YR5 00000006753YR8 00000294402MR3 00000008530YR8 00000009529YR9 00000007554YR9 00000009847YR5 00000008318YR8 00000007524YR2 00000007533YR3 00000010429YR9 00000009026YR6 00000294003MR9 00000005551YR7 00000006739YR7 00000007892YR3 00000007540YR8 00000008063YR0 00000006656YR3 00000010140YR2 00000298044MR9 00000006726YR4 00000293959MR3 00000006583YR9 00000296441MR9 00000296301MR5 00000291361MR4 00000008314YR7 00000007531YR7 00000008519YR1 00000006595YR3 00000007879YR0 00000011201YR1 00000011223YR5 00000292964MR4 00000009763YR4 00000009873YR1 00000293510MR4 00000297445MR9 00000008344YR4 00000011328YR2 00000007849YR3 00000291749MR0 00000010922YR3 00000029135MR9 00000007120YR9 00000007557YR2 00000009855YR8 00000290695MR6 00000291974MR4 00000007496YR3 00000293870MR2 00000007852YR7 00000006565YR6 00000008329YR5 00000007104YR3 00000290683MR2 00000293876MR9 00000008544YR9 00000006716YR5 00000290543MR8 00000296236MR3 00000008337YR8 00000008120YR8 00000294435MR3 00000290614MR7 00000296360MR1 00000294271MR2 00000292435MR5 00000294441MR1 00000010796YR1 00000011026YR2 00000295056MR6 00000295477MR4 00000294607MR7 00000010785YR4 00000011084YR1 00000011415YR7 00000295014MR5 00000010308YR5 00000291780MR5 00000293951MR0 00000293936MR1 00000294514MR5 00000290356MR5 00000295481MR6 00000294907MR1 00000295174MR7 00000294660MR6 00000007584YR6 00000293217MR6 00000007114YR2 00000007847YR7 00000008325YR3 00000290613MR9 00000296238MR9 00000294344MR7 00000006723YR1 00000290812MR7 00000006659YR7 00000007539YR0 00000007520YR0 00000006665YR4 00000296047MR4 00000295902MR1 00000292077MR5 00000290724MR4 00000293672MR2 00000005989YR9 00000011221YR9 00000006555YR7 00000294285MR2 00000293935MR3 00000294396MR7 00000010917YR3 00000293498MR2 00000010972YR8 00000292258MR1 00000291462MR0 00000008311YR3 00000011232YR6 00000294329MR8 00000291976MR9 00000010804YR3 00000010049YR5 00000007550YR7 00000009592YR7 00000011063YR5 00000007562YR2 00000005539YR2 00000010067YR7 00000294392MR6 00000295700MR9 00000295877MR5 00000295762MR9 00000007528YR3 00000294130MR0 00000290611MR3 00000293644MR1 00000296091MR2 00000011307YR6 00000009582YR8 00000293591MR4 00000290656MR8 00000290694MR9 00000294693MR7 00000294895MR8 00000295090MR5 00000290732MR7 00000292941MR2 00000008365YR9 00000010943YR9 00000006662YR1 00000007568YR9 00000007497YR1 00000008116YR6 00000008322YR0 00000011260YR7 00000294506MR1 00000294961MR8 00000295089MR7 00000295051MR7 00000008384YR0 00000297697MR5 00000290728MR5 00000290723MR6 00000292241MR7 00000293523MR7 00000293875MR1 00000291166MR7 00000010783YR9 00000011226YR8 00000010985YR0 00000009839YR2 00000007512YR7 00000290977MR8 00000291968MR6 00000293248MR1 00000293615MR1 00000293023MR8 00000296362MR7 00000009539YR8 00000009612YR3 00000007105YR0 00000292033MR8 00000010143YR6 00000008095YR2 00000005981YR6 00000005982YR4 00000295015MR2 00000291238MR4 00000011318YR3 00000010169YR1 00000294722MR4 00000010524YR7 00000006702YR5 00000006589YR6 00000011406YR6 00000009569YR5 00000009583YR6 00000291358MR0 00000295540MR9 00000011228YR4 00000294520MR2 00000291610MR4 00000294894MR1 00000290692MR3 00000296358MR5 00000292032MR0 00000294324MR9 00000293525MR2 00000294321MR5 00000293883MR5 00000293426MR3 00000291195MR6 00000296361MR9 00000295745MR4 00000294960MR0 00000292925MR5 00000295095MR4 00000296072MR2 00000294105MR2 00000293953MR6 00000292264MR9 00000296213MR2 00000296405MR4 00000292242MR5 00000291249MR1 00000292434MR8 00000294736MR4 00000291847MR2 00000009864YR0 00000010888YR6 00000295012MR9 00000009619YR8 00000295571MR4 00000291939MR7 00000010947YR0 00000006725YR6 00000007088YR8 00000007123YR3 00000008122YR4 00000006672YR0 00000008350YR1 00000008527YR4 00000009032YR4 00000008371YR7 00000010037YR0 00000010514YR8 00000010518YR9 00000007504YR4 00000010031YR3 00000010782YR1 00000010956YR1 00000291229MR3 00000291673MR2 00000291807MR6 00000291170MR9 00000291972MR8 00000291781MR3 00000291884MR5 00000290693MR1 00000292035MR3 00000008557YR1 00000292473MR6 00000292905MR7 00000292550MR1 00000293266MR3 00000291306MR9 00000011261YR5 00000290615MR4 00000291913MR2 00000292288MR8 00000291450MR5 00000291165MR9 00000293884MR3 00000292642MR6 00000293680MR5 00000006706YR6 00000294709MR1 00000295189MR5 00000292243MR3 00000295192MR9 00000293626MR8 00000292940MR4 00000009870YR7 00000293879MR3 00000294002MR1 00000293881MR9 00000292088MR2 00000292547MR7 00000007574YR7 00000007503YR6 00000009883YR0 00000010441YR4 00000295896MR5 00000295809MR8 00000295234MR9 00000295237MR2 00000295542MR5 00000295702MR5 00000296092MR0 00000295895MR7 00000296202MR5 00000296304MR9 00000296411MR2 00000294555MR8 00000295897MR3 00000297204MR0 00000292688MR9 00000296644MR8 00000296426MR0 00000296306MR4 00000295480MR8 00000295901MR3 00000293952MR8 00000291228MR5 00000291488MR5 00000011193YR0 00000010710YR2 00000006562YR3 00000291497MR6 00000293877MR7 00000290412MR6 00000011319YR1 00000007115YR9 00000009846YR7 00000295442MR8 00000011225YR0 0000SGH7460N5W 00000293641MR7 00000005566YR5 00000009464YR9 00000008364YR2 00000006661YR3 00000011247YR4 00000295484MR0 00000296110MR0 00000008118YR2 00000294407MR2 00000294443MR7 00000294723MR2 00000293878MR5 00000010984YR3 00000292211MR0 00000010488YR5 00000010718YR5 00000290696MR4 00000009079YR5 00000290637MR8 00000009018YR3 00000008103YR4 00000010821YR7 00000011119YR5 00000292467MR8 00000294307MR4 00000294145MR8 00000010041YR2 00000010877YR9 00000009616YR4 00000011017YR1 00000296215MR7 00000294159MR9 00000010916YR5 00000010513YR0 00000296204MR1 00000009002YR7 00000005565YR7 00000009417YR7 00000008102YR6 00000294004MR7 00000009025YR8 00000007558YR0 00000294989MR9 00000294146MR6 00000005226YR6 0000000B523YR3 00000006697YR7 00000294659MR8 00000293730MR8 00000011207YR8 00000007093YR8 00000291472MR9 00000291416MR6 00000295507MR8 00000295699MR3 00000292821MR6 00000297628MR0 00000008377YR4 00000296403MR9 00000007134YR0 00000293165MR7 00000295508MR6 00000296632MR3 00000292433MR0 00000294419MR7 00000007515YR0 00000007566YR3 00000007570YR5 00000293242MR4 00000293524MR5 00000292921MR4 00000008528YR2 00000005527YR7 00000009024YR1 00000006718YR1 00000291714MR4 00000290471MR2 00000294343MR9 00000292086MR6 00000009047YR2 00000291438MR0 00000295193MR7 00000290733MR5 00000293999MR9 00000005574YR9 00000294735MR6 00000294919MR6 00000007845YR1 00000011101YR3 00000009454YR0 00000293584MR9 00000291640MR1 00000294552MR5 00000295804MR9 00000297660MR3 00000293656MR5 00000290544MR6 00000007107YR6 00000005980YR8 00000291197MR2 00000291498MR4 00000291463MR8 00000008323YR8 00000294658MR0 00000294114MR4 00000295182MR0 00000290697MR2 00000007505YR1 00000290699MR8 00000009549YR7 00000294117MR7 00000293318MR2 00000294661MR4 00000293683MR9 00000293643MR3 00000005234YR0 00000292089MR0 00000006567YR2 00000009869YR9 00000005534YR3 00000006658YR9 00000291712MR8 00000296833MR7 00000292819MR0 00000292969MR3 00000010039YR6 00000290721MR0 00000290657MR6 00000293258MR0 00000292673MR1 00000294529MR3 00000293681MR3 00000293526MR0 00000006586YR2 00000297444MR2 00000007089YR6 00000293970MR0 00000011120YR3 00000009010YR0 00000010714YR4 00000010970YR2 00000293732MR4 00000010058YR6 00000011335YR7 00000294306MR6 00000009431YR8 00000006573YR0 00000294628MR3 00000005220YR9 00000290974MR5 00000290727MR7 00000291609MR6 00000007577YR0 00000290474MR6 00000294626MR7 00000296267MR8 00000296302MR3 00000291247MR5 00000293321MR6 00000292591MR5 00000294944MR4 00000297420MR2 00000290654MR3 00000294328MR0 00000293298MR6 00000294284MR5 00000295529MR2 00000008560YR5 00000297014MR3 00000011314YR2 00000295483MR2 00000006698YR5 00000007530YR9 00000006678YR7 00000006690YR2 00000007099YR5 00000010954YR6 00000007547YR3 00000007853YR5 00000007090YR4 00000006742YR1 00000007546YR5 00000006732YR2 00000007890YR7 00000010054YR5 00000008104YR2 00000008357YR6 000000008393Y1 00000007551YR5 00000006677YR9 00000007870YR9 00000007508YR5 00000007511YR9 00000008395YR6 00000009621YR4 00000011325YR8 00000290734MR3 00000295230MR7 00000296633MR1 00000007861YR8 00000290475MR3 00000292203MR7 00000292979MR2 00000292076MR7 00000011001YR1 00000294988MR1 00000293882MR7 00000008394YR9 00000006557YR3 00000006722YR3 00000008376YR6 00000008317YR0 00000008099YR4 00000007126YR6 00000007581YR2 00000009434YR2 00000010053YR7 00000293869MR4 00000295003MR8 00000007521YR8 00000010987YR6 00000010301YR0 00000290813MR5 00000290691MR5 00000914BD5A7E 00000950BD62A1 00000947BD61C0 00000930BD5D45 00000922BD5C3B 00001009BD6549 00000930BD5D58 00000912BD59AD 00001008BD6529 00000945BD6180 00000939BD5FB3 00000937BD5F04 00000922BD5C37 00001002BD6345 00001001BD633A 00001008BD6516 00001008BD6546 00001001BD6322 00000012275YR4 00000012351YR3 00000903BD55EB 00000918BD5B50 00000948BD61E8 00000931BD5D8B 00000933BD5E09 00000937BD5EE9 00000919BD5B77 00000833BD4AD4 00000952BD62F3 00000012387YR7 000S0933BD5E18 00000011910YR7 00000912BD599C 00000838BD4EA0 00000933BD5E06 00000934BD5E40 00000912BD59A8 00001008BD6519 00001001BD633B 000S1001BD6331 000S0933BD5E00 00000939BD5F72 00000937BD5F03 00000936BD5E70 00000937BD5EC9 000098318D48F0 00000836BD4D9A 00000012303YR4 00000012264YR8 00001007BD6468 00001008BD653E 00001007BD648A 00001002BD636B 000000948D61F3 00000950BD6294 000S0930BD5D48 00000914BD5A7A 000S0933BD5E17 00000937BD5EEA 00000291755MR7 00000939BD5F46 00001002BD634C 00001002BD63AA 00000930BD5D60 000S0933BD5E02 00000012402YR4 00000939BD5F4B 00000939BD5FAD 000S0933BD5DFD 00000937DB5EDO 00000291979MR3 00001009BD655E 00000915BD5AED 00001006BD645C 00000926BD5CF7 00000930BD5D3C 00000919BD5BAA 00000938BD5F15 00000012385YR1 00001001BD6329 00000937BD5F02 00000831BD4A07 00001002BD6392 00000934BD5E53 00000012406YR5 00000919BD5BC4 00000950BD62B6 00000939BD5F41 00000012355YR4 00000939BD5F48 00000939BD5F71 00001002BD6361 00000938BD5F18 000S0933BD5DF3 00000912BD59AB 00000937BD5EEC 00000939BD5F49 00000939BD5FA6 00000948BD61EC 000S0933BD5E0E 000000948D61F5 00000933BD5E10 00000934BD5E42 00000930BD5D3E 00001003BD641B 00000919BD5B98 00000291138MR6 00000919BD5BB5 00000845BD51B0 00000939BD5FAE 00000012272YR1 00001007BD6479 000S0934BD5E46 00000836BD4DA8 00000012346YR3 00001007BD6485 00000931BD5D8E 000S0919BD5BC0 00000930BD5D51 00000933BD5E08 00000914BD5A81 00001007BD6469 00001003BD6437 00000292641MR8 000000105258M0 00000293500MR5 00000292316MR7 00000293651MR6 00000294692MR9 00000012311YR7 00000845BD51D2 00000930BD5D61 00000918BD5B36 000S0934BD5E4A 00000937BD5EFF 00000903BD55DE 00000918BD5B55 00001008BD6540 00000933BD5E0C 00001009BD6562 00000297008MR5 00000937BD5EEF 00000930BD5D3A 00000948BD61EB 00001002BD6399 00000292640MR0 000S0918BD5B5F 000S0934BD5E3A 00000292078MR3 00000292080MR9 00000292028MR8 00000292942MR0 00000295240MR6 00000296036MR7 00000294927MR9 00000295697MR7 00000295239MR8 00000294926MR1 00000295594MR6 00000290324MR3 00000292984MR2 00001002BD638F 00001002BD6373 00001002BD639E 00000012359YR6 00000012319YR0 00000931BD5D90 00001001BD6323 00000012342YR2 00000948BD61EA 00000939BD5F6B 00001002BD633E 00000931BD5D7C 00000934BD5E4E 00000012373YR7 00000292524MR6 00000296050MR8 00000294636MR6 00000292029MR6 00000291969MR4 00000291850MR6 00000947BD61BE 00001002BD6360 00000291970MR2 00000012360YR4 00000939BD5F69 00000837BD4E13 00001001BD6326 00000905BD566F 00000919BD5BB3 00000937BD5EDA 00001003BD6401 00000903BD55FC 00001002BD63A7 00000931BD5D8A 00001003BD640C 00000290325MR0 00000012358YR8 00000919BD5BA8 00000950BD62BD 00001007BD6491 00000293650MR8 00000937BD5ED6 00000012315YR8 00000012304YR2 00000848BD5311 00000939BD5F52 00000938BD5F17 00000912BD59B9 00000012270YR5 00000939BD5F50 00000939BD5FAC 00000913BD5A48 00000831BD49FE 00000919BD5BA1 00001009BD655B 00000012352YR1 00000944BD6168 00000950BD6295 00000913BD5A4B 00001007BD6475 00000913BD5A47 00000030BD4960 00000952BD62FB 000002008AUG20 00000838BD4EA8 00000814BD39ED 00000937BD5ED2 00001007BD6498 00000948BD61E2 00000011909YR9 00000012383YR6 00000012268YR9 00000937BD5ED1 00000012354YR7 00001002BD636D 00000012266YR3 00001003BD641E 00000937BD5EF2 00000937BD5EFB 00000939BD5F64 00000938BD5F16 00001003BD63F9 00001003BD642D 00001003BD63FD 00000950BD6290 00000290310MR2 00000012274YR7 00000914BD5A88 00000912BD59B4 00000930BD5D38 00000093BD5F68 00000948BD61E5

Other devices under the same FDA submission

1 device cleared under K052293 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.