Skip to content

Integra Radionics HRAIM Intubation Head Ring Assembly recallZ-1400-2011 · Integra LifeSciences Corp. · initiated Jan 12, 2011

Terminated All recalls posted in February 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Overall length of the intubation hoop in the HRAIM Intubation Head Ring Assembly is too long and will not allow a device to to attach

FDA root cause category: Manufacturing material removal

What the company is doing

Integra sent a Medical Device Field Corrective Action letter dated January 12, 2011, to all affected customers via Federal Express. Integra Sales Specialists were provided with the replacement part and written instructions on how to replace the affected intubation hoops from customer inventory identified in the traceability report. Upon completion of the field correction each sales specialist was instructed to return the affected hoop along with a completed copy the Recall Acknowledgement and Return Form. If additional information is required contact me at (609) 936-2485 or [email protected].

Recall details

Recall number
Z-1400-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jan 12, 2011
Posted by the FDA
Feb 23, 2011
Recalling firm
Integra LifeSciences Corp.
Quantity
8 units
Product code
HAW
FDA submissions
K944463
Distribution
Worldwide Distribution - USA, Canada, and Australia

Product description

Integra Radionics HRAIM Intubation Head Ring Assembly Ref: HRAIM Head Rings serve as the general stereotactic treatment platform. Head Rings are used to provide a reference frame for instrumentation used for precise spatial localization and treatment of physiologic targets for stereotactic neurosurgical procedures such as craniotomies, biopsies, functional neurosurgery, and radiation therapy. Head Rings are delivered to the user non-sterile, and are reusable.

Code information (models, lots, serial numbers)

Lot/Serial Number(s): 0189463, 0189289, 0186642, 0184885

Other devices under the same FDA submission

2 devices cleared under K944463 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.