VCare Vaginal-Cervical Ahluwalia's Retractor-Elevator… recallZ-1520-2011 · ConMed Corporation · initiated Feb 2, 2011
Terminated All recalls posted in March 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ConMed Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product manufactured after 4/11/2008, but before 8/01/2009, contains the earlier Instructions For Use, IFU (PN 17416, Rev. D) which does not contain the revisions on the addition of Warning Notes 10 and 11 and the expansion of Warning Note 6 to the VCare Instructions for Use. The reason for the field correction is to provide a copy of the revised IFU to all customers who may have unexpired Vcare
FDA root cause category: Device Design
What the company is doing
ConMed Corporation sent a communication to provide its direct consignees with important supplemental information concerning the Instructions for Use (IFU) for the VCARE Vaginal-Cervical Ahluwalia's Retractor-Elevator uterine manipulators. The direct consignees were instructed to place a copy of the enclosed Instructions for Use, Part No. 14071 (Rev. A), with any inventory in your possession with an expiration date prior to August 2011 (2011-08) and replace any other Instructions for Use in thier possession. A copy of the enclosed Instructions for Use is already packaged with all devices bearing later expiration dates. Any inventory in their consignees possession with an expiration date prior to January 2011 (2011-01) is expired and should not be used. If the devices were further distributed to other facilities or distributors, or their consignees are a distributor, they were instructed to forward this information as appropriate to any known users of these devices. For questions regarding this recall call 315-624-3225. --- The IFU Warning changes are as follows: A new warning statement # 6 asks the user to verify the presence of all the components of the device upon removal of the VCARE device from the patient. Although the prior IFU contained warnings about verifying that all components of the device were properly removed following use during surgery, the new statement expands the earlier, briefer statement. 6. Upon removing VCARE, the surgeon should visually inspect the VCARE device, and the patient, to make sure that the entire VCARE device was properly removed and that no components or fragments of these components were retained in the patient. There are 5 parts/components to the VCARE Cervical Elevator Retractor. These are: 1) the balloon; 2) the forward "cervical" cup; 3) the back or vaginal cup; 4) the locking assembly with thumb screw; 5) the metal shaft and handle with balloon inflation valve. --- A new warning statement # 10 further addresses
Recall details
- Recall number
- Z-1520-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 2, 2011
- Posted by the FDA
- Mar 3, 2011
- Recalling firm
- ConMed Corporation
- Quantity
- 31,028 VCare¿ Vaginal-Cervical Ahluwalia's Retractor-Elevator devices
- Product code
- L
- FDA submissions
- K071907
- Distribution
- Worldwide Distribution - USA including NY, TX, CA, IA, AZ, PA, IN, and KS and the countries of Austria, Jordan, Turkey, China, Portugal, Chile, Belgium, Ecuador, Brazil, Romania, Japan, Malaysia, India Switzerland, Colombia, Netherlands, Spain, Saudi Arabia, France, and Canada.
Product description
VCare Vaginal-Cervical Ahluwalia's Retractor-Elevator devices. The VCare IFU is packaged one per carton of eight VCare Pouched Devices. The Instructions for Use is an 8.5" x 12" 12-page booklet and is packed loose in the carton. Only the carton will have to be opened to replace the old IFU; there is no need to open the sterile packaging of the individual VCare pouches. The ConMed VCARE Retractor/Elevator is indicated for manipulation of the uterus and injection of fluids or gases during laparoscopic procedures such as Laparoscopic Assisted Vaginal Hysterectomy (LAVH), Total Laparoscopic Hysterectomy (TLH), minilap, laparoscopic tubal occlusion, or diagnostic laparoscopy and also maintains pneumoperitoneum by sealing the vagina once colpotomy is performed.
Code information (models, lots, serial numbers)
VCare¿ Standard Cervical Cup (Catalog # 60-6085-100), VCare¿ Small Cervical Cup (Catalog # 60-6085-101) and VCare¿ Large Cervical Cup (Catalog # 60-6085-102). Lot codes for product manufactured between 01/20/2009 (Lot 090120X) and through 07/31/2009 (Lot 090731X). Lot Code Key = YYMMDDX, where X = manufacturing shift (e.g. 1, 2 or 3)
Other devices under the same FDA submission
3 devices cleared under K071907 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VCARE MEDIUM 60-6085-201Conmed Corporation · CANNULA, MANIPULATOR/INJECTOR, UTERINE9 reports
- VCARE SMALL 60-6085-200Conmed Corporation · CANNULA, MANIPULATOR/INJECTOR, UTERINE6 reports
- VCARE LARGE 60-6085-202Conmed Corporation · CANNULA, MANIPULATOR/INJECTOR, UTERINE1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page