Skip to content

i-STAT CHEM 8+ Cartridge, List Number 03P91-25 recallZ-2022-2011 · Abbott Point Of Care Inc. · initiated Feb 15, 2011

Terminated All recalls posted in April 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

i-STAT CHEM 8+ cartridge may generate false results for sodium, chloride and hematocrit.

FDA root cause category: Other

What the company is doing

The firm, Abbott Point of Care (APOC), sent an "Urgent Recall Notice" letter dated February 2011 with business reply cards attached via Fed Ex overnight to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to discontinue use of cartridges from the box numbers of identified lots; discontinue use as a precaution, if their cartridges are no longer traceable to a specific box; return all unused cartridges as instructed on the enclosed Business Reply Card; if they've provided another institution with the cartridges, provide the institution with a copy of this letter, and complete and return the business reply card to APOC. If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 800-336-8020, Option 1 or your Abbott Point of Care representative.

Recall details

Recall number
Z-2022-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 15, 2011
Posted by the FDA
Apr 18, 2011
Recalling firm
Abbott Point Of Care Inc.
Quantity
6575 cartirdges
Product code
JGS
FDA submissions
K053110
Distribution
Nationwide distribution.

Product description

i-STAT CHEM 8+ Cartridge, List Number 03P91-25; 510 K053110 Useful for monitoring a variety of conditions. The panel of tests is used in the hospital environment to assess kidney function, electrolyte status, acid/base balance, and blood sugar level. This test panel is also used to assess hypertension and hypokalemia.

Code information (models, lots, serial numbers)

510 K053110; Lot # C10247 boxes 0532 through 0593

Other devices under the same FDA submission

1 device cleared under K053110 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page