Skip to content

CELL-DYN Shear Valve Assembly used on CELL-DYN 3700SL… recallZ-2015-2011 · Abbott Laboratories · initiated Mar 15, 2011

Terminated All recalls posted in April 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Nylon Washer has an outer diameter which is undersized resulting in the inability to generate patient results.

FDA root cause category: Pending

What the company is doing

Product Correction letters were sent on March 15, 2011. via Federal Express. The letter identified the affected product, explained the problem, the patient impact and the necessary action that were to be taken. Customers were to review the letter and inform their Laboratory Manager of the pending action.; complete and return the Customer Reply Form; and save a copy of the letter for their records. If there were any questions, customers were to contact Customer Support at 1-877-4ABBOTT or their local hematology customer support representative for customers outside of the US.

Recall details

Recall number
Z-2015-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 15, 2011
Posted by the FDA
Apr 29, 2011
Recalling firm
Abbott Laboratories
Quantity
174 total, all varieties
Product code
GKZ
FDA submissions
K991605
Distribution
Worldwide Distribution -- AZ, CA, FL, IA, ID, IN, MA, MI, MN, MS, ND, NY, OK, OR, SC, SD, TX, WI, WY and Germany, Singapore, Mexico, China, South Korea, Argentina, Japan, Brazil, Canada, Australia, and Chile.

Product description

CELL-DYN Shear Valve Assembly used on CELL-DYN 3700SL Analyzer and CELL-DYN 3700SL Analyzer (refurb). Abbott Diagnostics Division, Santa Clara, CA 05054 Designed for in vitro diagnostic use in clinical laboratories.

Code information (models, lots, serial numbers)

List Number 02H31-01; lot numbers: 23444AK, 23445AK, 23446AK, 23448AK to 23456AK, 23461AK, 20576AK, 21408AK, 21426AK, 21515AK, 21651AK, 21694AK, 271726AK, 21955AK, 22145AK, 22229AK, 22458AK, 22732AK, 22760AK, 23198AK, 23209AK, 23210AK, 23236AK, 22881AK, 19966AK, 20880AK, 20924AK, 21051AK, 21522AK, and 22650AK. Refurbished list number: 02H31-03; lot numbers: 20074AK96, 20334AK96, 20416AK96, 20751AK96, 20985AK96, 21167AK96, 21368AK96, 21610AK96, 21753AK96, 21842AK96, 21885AK96, 21934AK96, 22142AK96. 22340AK96, 22346AK96, 22380AK96, 22501AK96, 23056AK96, 23145AK96, 23457AK96, 23458AK96, 23459AK96, 23460AK96, and 23462AK96.

Other devices under the same FDA submission

3 devices cleared under K991605 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 58188).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page