Enzymatic Creatinine (ECREA) Flex(R) reagent cartridges… recallZ-2154-2011 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 10, 2010

Terminated All recalls posted in May 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Results may be falsely depressed up to 0.4 mg/dL or falsely elevated up to 0.6 mg/dL across the assay range.

FDA root cause category: Device Design

What the company is doing

Siemens sent an Urgent Field Safety Notice letter dated June 2010 to all Dimension Vista(R) ECREA customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to: Discontinue use of ECREA method and transition to the Dimension(R) Jaffe creatinine (CREA) method. To notify anyone to whom they may have distributed the product. Complete and return the Effectiveness Check Questionnaire attached to the letter by fax to (302) 631-8467. For questions regarding this recall call (302) 631-6299.

Recall details

Recall number
Z-2154-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jun 10, 2010
Posted by the FDA
May 9, 2011
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1,399
Product code
JFY
FDA submissions
K090330
Distribution
Nationwide Distribution - including AL, AZ, CA, CO, DE, FL, GA, ID, IL, IA, MD, MA, MI, MT, NH, NJ, NY, NC, OH, OR, PA, SC, TX, VA, WA, and WI.

Product description

Enzymatic Creatinine (ECREA) Flex(R) reagent cartridges, K1270 for the Dimension Vista(R) 500 System The ECREA method is an in vitro diagnostic test for the quantitative measurement of creatinine in human serum, plasma and urine on the Dimension Vista(R) System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for other urine analytes.

Code information (models, lots, serial numbers)

Catalog number K1270 - lot numbers 09224AA, 09286AA, 09350AC, 10011AF, 10047AA, 10089AC, 10089AD, 10131AC and 10131AD.

Other devices under the same FDA submission

1 device cleared under K090330 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.