Philips Microstream Model 989803159581 VITALINE H SET… recallZ-2248-2011 · Philips Healthcare Inc. · initiated Apr 7, 2011
Terminated All recalls posted in May 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Healthcare Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Fine plastic strands on inner surface of the infant/ neonatal airway adapter may become dislodged and inhaled by the patient.
FDA root cause category: Equipment maintenance
What the company is doing
The firm, Philips, sent a Field Safety Notice titled "URGENT MEDICAL DEVICE RECALL" letter dated April 2011 to its customers. Customer notifications outside the US will be managed by Philips representatives in each affected geography. The letter described the product, problem and actions to be taken. The customers were instructed to immediately identify all products from affected lots; remove from inventory and dispose in accordance with local regulations (new products will be sent proactively to each affected customer). Direct accounts were instructed to notify additional customers to whom the product has been distributed. These customers are instructed to follow the instructions in the "Action to be taken by Customer/User" section of the Notice. Questions regarding the recall can be directed towards the customer's local Philips representative at 1-800-722-9377.
Recall details
- Recall number
- Z-2248-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 7, 2011
- Posted by the FDA
- May 26, 2011
- Recalling firm
- Philips Healthcare Inc.
- Quantity
- 75 units (3 boxes)
- Product code
- CCK
- FDA submissions
- K980327
- Distribution
- Worldwide distribution.
Product description
Philips Microstream Model 989803159581 VITALINE H SET INFANT/NEONATAL 25 UN PHILIPS These products are used with Philips Monitors for the measurement of endtidal CO2 in infants and neonates. They provide a means to transport samples of patient respiratory gases from the patient's airway to the Microstream capnograph. This by performed by placing the airway adapter portion of the FilterLine H Set Infant/Neonate an VitaLine H Set Infant/Neonatal in the patient airway system of a respirator or an anesthesia machine.
Code information (models, lots, serial numbers)
Manufactured from October 2010 through February 2011. Product Number: 989803159581; Lot Code: M8409P10
Other devices under the same FDA submission
6 devices cleared under K980327 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Microstream 006330Oridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream 010808Oridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream 015707_RMS1Oridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream XS04624COV-05Oridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream 006324COV-05Oridion Medical 1987 Ltd. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall0 reports
- Microstream 011108Oridion Medical 1987 Ltd. · Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 58471).
- Z-2247-2011 · TerminatedPhilips Microstream Model M1923A FILTERLINE H SET INF/NEO 25UN PHILIPS These products are used with Philips Monitors for the measurement of endtidal CO2 in infants and neonates. …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page