Skip to content

ES0101, AEM Disposable Curved Scissors, 1/2", 35cm, Sterile… recallZ-2395-2011 · Encision, Inc. · initiated Apr 7, 2011

Terminated All recalls posted in May 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Encision, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sterility may be compromised due to holes in packaging.

FDA root cause category: Package design/selection

What the company is doing

Encision, Inc. sent an Important Safety Notice letter dated April 7, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to either carefully inspect any remaining pouches for readily visible holes in the Tyvek pouch, or return the affected lots to Encision for inspection. Customers were also asked to fill in and fax back the attached Confirmation to 303-444-2693.

Recall details

Recall number
Z-2395-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 7, 2011
Posted by the FDA
May 31, 2011
Recalling firm
Encision, Inc.
Quantity
574 Ten-Count Boxes, 24 Kits
Product code
GEI
FDA submissions
K912780
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia and New Zealand

Product description

ES0101, AEM Disposable Curved Scissors, 1/2", 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical procedures

Code information (models, lots, serial numbers)

Lot # QJC, QKD, QLE, RAG, RBD

Other devices under the same FDA submission

3 devices cleared under K912780 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 58518).