Skip to content

Radiesse, Injectable Implant - 1.3cc Syringe, Sterile… recallZ-2221-2011 · Merz Aesthetics, Inc. · initiated Apr 6, 2011

Terminated All recalls posted in May 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Merz Aesthetics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Merz Aesthetics, Inc. is recalling 1 lot (1034417) of Radiesse 1.3 cc Implant. The lot was supposed to be scrapped because it had failed to meet its Loss On Drying (LOD) specification. Unfortunately it had been marked as approved, released and shipped.

FDA root cause category: Employee error

What the company is doing

The firm, Merz Aesthetics, Inc., contacted its consignee/customer via e-mail on April 6, 2011. The e-mail states the product, problem and action to be taken. The customer was instructed to follow up with the concerned and inform the status time to time to the Recall Coordinator. The product was confirmed to be quarantined and has been requested to be sent back. If you need any additional information regarding this issue or if you have any questions, contact the International Regulatory Affairs Specialist at 262-835-3300 ext. 3081.

Recall details

Recall number
Z-2221-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 6, 2011
Posted by the FDA
May 19, 2011
Recalling firm
Merz Aesthetics, Inc.
Quantity
695 units
Product code
LMH
FDA submissions
P050052, P050037
Distribution
International only: Taiwan.

Product description

Radiesse, Injectable Implant - 1.3cc Syringe, Sterile, Sterilized Using Steam, Use Before 2013-01, Ref 8046M7, Lot 1024417, Manufactured by: BioForm Medical Inc, 4133 Courtney Rd #10 Franksville, WI 53126. Subdermal implantation for the correction of moderate to severe facial wrinkles and folds, such as masolabial folds and it is also intended for restoration and/or correction of the signs of facial fat loss (lipoatrophy) in people with human immunodeficiency virus.

Code information (models, lots, serial numbers)

Lot 1024417

Other devices under the same FDA submission

5 devices cleared under P050052, P050037 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.