Medical device recalls: January 2006
The FDA posted 117 medical device recall records in January 2006, from 72 recall events by 58 companies. 117 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Roche Diagnostics Corp. · 3 events, 15 records
- Baxter Healthcare Corp. · 3 events, 4 records
- Becton Dickinson & Co. · 3 events, 3 records
- Boston Scientific Corporation · 2 events, 6 records
- Becton Dickinson & Company · 2 events, 5 records
- Smiths Medical ASD, Inc. · 2 events, 4 records
- Boston Scientific Scimed · 2 events, 3 records
- Cook, Inc. · 2 events, 3 records
FDA root cause
- Other · 112
- Device Design · 2
- Software design · 1
- Software change control · 1
- Process control · 1
Not yet classified (117)
The FDA has posted these records but has not published a class for them yet.
- Z-0409-06 · Boston Scientific Corporation · posted Jan 31, 2006 · Terminated
The distal shaft of the catheter may separate during withdrawal of the device from the patient. If the shaft separates, the surgical procedure may be prolonged or require more in-depth surgery to physically remove the broken-off shaft from the blood vessel.
- Z-0444-06 · Baxter Healthcare Corp. · posted Jan 31, 2006 · Terminated
Interruption of therapy due to battery undercharging, the generation of air-in-line alarms due to IV administration set tugging, and gearbox wear; and underinfusion due to misalignment of the pump head components.
- Z-0445-06 · Baxter Healthcare Corp. · posted Jan 31, 2006 · Terminated
Interruption of therapy due to battery undercharging, the generation of air-in-line alarms due to IV administration set tugging, and gearbox wear; and underinfusion due to misalignment of the pump head components.
- Z-0461-06 · AGA Medical Corporation · posted Jan 31, 2006 · Terminated
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (PTFE) plastic lining of the sheath. Those shavings could potentially migrate into a patient's bloodstream.
- Z-0462-06 · AGA Medical Corporation · posted Jan 31, 2006 · Terminated
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (PTFE) plastic lining of the sheath. Those shavings could potentially migrate into a patient's bloodstream.
- Z-0463-06 · AGA Medical Corporation · posted Jan 31, 2006 · Terminated
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (PTFE) plastic lining of the sheath. Those shavings could potentially migrate into a patient's bloodstream.
- Z-0464-06 · AGA Medical Corporation · posted Jan 31, 2006 · Terminated
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (PTFE) plastic lining of the sheath. Those shavings could potentially migrate into a patient's bloodstream.
- Z-0465-06 · AGA Medical Corporation · posted Jan 31, 2006 · Terminated
The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (PTFE) plastic lining of the sheath. Those shavings could potentially migrate into a patient's bloodstream.
- Z-0411-06 · Baxter Healthcare Corp. · posted Jan 28, 2006 · Terminated
Possible shorting of some printed circuit boards could result in a possible interruption of therapy or cause the pump to fail the Power-On Self Test.
- Z-0412-06 · Baxter Healthcare Corp. · posted Jan 28, 2006 · Terminated
Failures within the PCA profile as well as incidents resulting in interruptions of therapy in various profiles.
- Z-0446-06 · Depuy Mitek, a Johnson & Johnson Co. · posted Jan 26, 2006 · Terminated
The device may contain one strand of suture instead of two strands
- Z-0447-06 · Wahl Clipper Corp · posted Jan 26, 2006 · Terminated
The massager may develop a crack in the cord insulation where it exits out of the strain relief. A crack in the insulation may expose live conductors resulting in a possible electric shock and injury to the user.
- Z-0448-06 · Wahl Clipper Corp · posted Jan 26, 2006 · Terminated
The massager may develop a crack in the cord insulation where it exits out of the strain relief. A crack in the insulation may expose live conductors resulting in a possible electric shock and injury to the user.
- Z-0449-06 · Ethicon, Inc. · posted Jan 26, 2006 · Terminated
Lots of Proceed Surgical Mesh may delaminate from the polypropylene mesh during certain hernia repairs. Exposure of the polypropylene layer to the bowel prior to reperitonealization could increase the risk of adhesions and bowel fistulization.
- Z-0438-06 · Siemens Medical Solutions USA, Inc · posted Jan 25, 2006 · Terminated
patient information can be stored in another patient's file.
- Z-0439-06 · Siemens Medical Solutions USA, Inc · posted Jan 25, 2006 · Terminated
patient information can be stored in another patient's file.
- Z-0440-06 · Boston Scientific Scimed · posted Jan 25, 2006 · Terminated
One lot of one catalog number of the Sentinol Nitinol Biliary Stent System may have been packaged in a Sentinol nitinol Vascular outer carton. The product, pouch label and carton label are all correct and the correct DFU is in the package.
- Z-0441-06 · Cook Endoscopy · posted Jan 25, 2006 · Terminated
The dilator size does not match the label on the product and product packaging.
- Z-0442-06 · Ev3, Inc · posted Jan 25, 2006 · Terminated
A Primus device was labeled as a BXB 35 09-37-75 but contained a BXB35 07-27-75 stent.
- Z-0443-06 · Becton Dickinson & Co. · posted Jan 25, 2006 · Terminated
A operator's manual for a diagnostic test kit is incorrect and may cause inaccurate results for bacterial identification and antibiotic susceptibility in patient samples.
- Z-0413-06 · GE Healthcare · posted Jan 24, 2006 · Terminated
In certain cases, one of the four collimator locking mechanisms that attach the collimator to the head may work itself free if collimator exchange has not been performed for an extended period of time. This situation, if neglected, could eventually cause…
- Z-0414-06 · GE Healthcare · posted Jan 24, 2006 · Terminated
In certain cases, one of the four collimator locking mechanisms that attach the collimator to the head may work itself free if collimator exchange has not been performed for an extended period of time. This situation, if neglected, could eventually cause…
- Z-0415-06 · GE Healthcare · posted Jan 24, 2006 · Terminated
In certain cases, one of the four collimator locking mechanisms that attach the collimator to the head may work itself free if collimator exchange has not been performed for an extended period of time. This situation, if neglected, could eventually cause…
- Z-0416-06 · Abbott Molecular · posted Jan 24, 2006 · Terminated
The DNA-Probe has a target size of ~300 Kb region instead of the labeled ~190 Kb region.
- Z-0417-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0418-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0419-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0420-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0421-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0422-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0423-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0424-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0425-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0426-06 · Roche Diagnostics Corp. · posted Jan 24, 2006 · Terminated
Incorrect patient blood glucose test results may be reported due to a software problem.
- Z-0427-06 · Instrumentation Industries Inc · posted Jan 24, 2006 · Terminated
potential for the flexible tube portion of the assembly to become brittle and break
- Z-0428-06 · Hitachi Medical Systems America Inc · posted Jan 24, 2006 · Terminated
The bolts holding the lift mechanism may be damaged causing them to fail and the table top to drop several inches.
- Z-0429-06 · Hitachi Medical Systems America Inc · posted Jan 24, 2006 · Terminated
The bolts holding the lift mechanism may be damaged causing them to fail and the table top to drop several inches.
- Z-0430-06 · Becton Dickinson & Co. · posted Jan 24, 2006 · Terminated
A calibrator for an in vitro diagnostic kit is not standarized correctly and may cause a calibration error when measuring microbiological cultures for turbidity.
- Z-0431-06 · Integra LifeSciences Corp. · posted Jan 24, 2006 · Terminated
Additional Model 24kHz Tip Sets were inadvertently packaged with the incorrect 35kHz Neuro Flues. (2nd recall, same problem see Z-141-6)
- Z-0432-06 · Boston Scientific Scimed · posted Jan 24, 2006 · Terminated
Some Mach 1 6F Coronary Guide Catheter could be mislabeled in that a few units labeled as Mach 1 6F Kimny Curve style, could contain a Mach 1 6F JR 3.5 curve style and a few units labeled as Mach 1 6F JR 3.5 curve style, could contain a Mach 1 6F Kimny…
- Z-0433-06 · Boston Scientific Scimed · posted Jan 24, 2006 · Terminated
Some Mach 1 6F Coronary Guide Catheter could be mislabeled in that a few units labeled as Mach 1 6F Kimny Curve style, could contain a Mach 1 6F JR 3.5 curve style and a few units labeled as Mach 1 6F JR 3.5 curve style, could contain a Mach 1 6F Kimny…
- Z-0434-06 · Tnco, Inc. · posted Jan 24, 2006 · Terminated
The tip may not engage with the handle and may dislodge
- Z-0435-06 · Lifescan Inc · posted Jan 24, 2006 · Terminated
Product may be defective and may give low inaccurate glucose results.
- Z-0436-06 · Smiths Medical PM, Inc. · posted Jan 24, 2006 · Terminated
A potential safety related problem associated with the Pnuepac babyPac Ventilator model 100 has been discovered in that the interface between the air input hose and the air input fitting may leak, and if tightened too much, the air input fitting may loosen…
- Z-0437-06 · Datascope Corp · posted Jan 24, 2006 · Terminated
Software anomaly that affects the Panorama Telepack where if communications are lost during standby, certain alarms which appear set could be actually off.
- Z-0408-06 · Diagnostica Stago, Inc. · posted Jan 20, 2006 · Terminated
The mean levels (%) of normal and abnormal control plasmas (STA Liatest Control N and P) obtained with STA Liatest Free Protein S are found in the lower part of the range.
- Z-0360-06 · Vapotherm, Inc. · posted Jan 19, 2006 · Terminated
A medical device firm changed the labeling of its respiratory gas humidifier by revising the protocol for higher level disinfectants to be used by healthcare practitioners.
- Z-0407-06 · Sonosite, Inc. · posted Jan 19, 2006 · Terminated
22 gauge needle guides were intermixed with 21 gauge guides and labeled as 21 gauge needle guides.
- Z-0393-06 · Varian Medical Systems Inc · posted Jan 18, 2006 · Terminated
The export of multi-static segment BrainLab M 3 MLC from VARiS Vision may produce undesirable result, which could lead to mistreatment.
- Z-0394-06 · Smiths Medical ASD, Inc. · posted Jan 18, 2006 · Terminated
Sterility is compromised by incomplete package seal.
- Z-0395-06 · Smiths Medical ASD, Inc. · posted Jan 18, 2006 · Terminated
Sterility is compromised by incomplete package seal.
- Z-0396-06 · Smiths Medical ASD, Inc. · posted Jan 18, 2006 · Terminated
Sterility is compromised by incomplete package seal.
- Z-0397-06 · Plus Orthopedics USA · posted Jan 18, 2006 · Terminated
The adjustment knob tower on one device was found to have broken off from the body of the instrument. This did not happend during a surgery and is believed to have been caused by the instrument being dropped or otherwise mishandled during hospital…
- Z-0398-06 · Abbott Laboratories, Inc · posted Jan 18, 2006 · Terminated
System software assigns a calibrator default volume of 2.0uL when field is left empty by operator at time assay parameters set. Patient results could be affected if the volume required is not 2.0 uL.
- Z-0399-06 · Stentor Inc · posted Jan 18, 2006 · Terminated
When using a third party dictation accessory with the iSite PACS, a malfunction may occur that would result in the petient/exam selected and shown may be different on the radiologist console and diagnostic display screens
- Z-0400-06 · Abbott Laboratories MPG · posted Jan 18, 2006 · Terminated
Some of the kits were shipped at the incorrect shipping temperature. They were shipped ambient, but should have been shipped frozen.
- Z-0401-06 · CooperSurgical, Inc. · posted Jan 18, 2006 · Terminated
Potential for the plunger to separate from the piston
- Z-0402-06 · Becton Dickinson & Company · posted Jan 18, 2006 · Terminated
Safety mechanisms on the syringes are rotating too freely. This can result in the safety mechanism not holding its rotational position and increasing risk of an accidental needle-stick.
- Z-0403-06 · Becton Dickinson & Company · posted Jan 18, 2006 · Terminated
Safety mechanisms on the syringes are rotating too freely. This can result in the safety mechanism not holding its rotational position and increasing risk of an accidental needle-stick.
- Z-0404-06 · Becton Dickinson & Company · posted Jan 18, 2006 · Terminated
Safety mechanisms on the syringes are rotating too freely. This can result in the safety mechanism not holding its rotational position and increasing risk of an accidental needle-stick.
- Z-0405-06 · Becton Dickinson & Company · posted Jan 18, 2006 · Terminated
Safety mechanisms on the syringes are rotating too freely. This can result in the safety mechanism not holding its rotational position and increasing risk of an accidental needle-stick.
- Z-0406-06 · Becton Dickinson & Co. · posted Jan 18, 2006 · Terminated
Microbiological based test system may be contaminated with bacteria and cause false clinical test results in patient samples.
- Z-0390-06 · Cook, Inc. · posted Jan 13, 2006 · Terminated
The side of the box gives the size of the stent as 10 x 40 instead of the actual size of 7 x 40. The label front is correct.
- Z-0391-06 · Cook, Inc. · posted Jan 13, 2006 · Terminated
The side of the box gives the size of the stent as 8 x 60 instead of the actual size of 6 x 60. The label front is correct.
- Z-0392-06 · Medtronic Gastroenterology / Urology · posted Jan 13, 2006 · Terminated
Certain lots have a manufacturing defect that allows the water pushed through the catheter to travel back into the connector of the Digitrapper pH recorder. Medtronic has determined that patient safety is not compromised, but catheter function may be…
- Z-0388-06 · Ossur · posted Jan 12, 2006 · Terminated
After one year devices assembled with a new type of O-rings could exhibit enough wear to allow a gradual oil leak which would increase play in the ankle sub-assembly of the foot.
- Z-0389-06 · Hospira Inc · posted Jan 12, 2006 · Terminated
The pump may shut-off prematurely during battery operation without providing adequate response time when 'E320' service warning message appears.
- Z-0371-06 · Smith & Nephew, Inc., Endoscopy Div. · posted Jan 11, 2006 · Terminated
Drill does not contain laser etched depth marks to assure proper depth of inserted device
- Z-0372-06 · Becton Dickinson & Company · posted Jan 11, 2006 · Terminated
Cracked finger grip luer adapter components on the Push Button Blood Collection Set.
- Z-0373-06 · Kimberly-Clark Corporation · posted Jan 11, 2006 · Terminated
The integrity of the sterile unit package may become compromised.
- Z-0374-06 · Kimberly-Clark Corporation · posted Jan 11, 2006 · Terminated
The integrity of the sterile unit package may become compromised.
- Z-0375-06 · Kimberly-Clark Corporation · posted Jan 11, 2006 · Terminated
The integrity of the sterile unit package may become compromised.
- Z-0376-06 · Boston Scientific Corp · posted Jan 11, 2006 · Terminated
The patient sample may become contaminated because the wax plug, intended to protect the sample from contamination during collection, may be missing.
- Z-0377-06 · Boston Scientific Corp · posted Jan 11, 2006 · Terminated
The patient sample may become contaminated because the wax plug, intended to protect the sample from contamination during collection, may be missing.
- Z-0378-06 · Instrumentation Laboratory Co. · posted Jan 11, 2006 · Terminated
Software error may report higher results for glucose and lactate
- Z-0379-06 · Kinetic Concepts, Inc · posted Jan 11, 2006 · Terminated
Possible mislabeling of shipping cartons.
- Z-0380-06 · Beckman Coulter Inc · posted Jan 11, 2006 · Terminated
Under certain conditions an incorrect Tube ID (Barcode) can be displayed and printed on the Runtime Panel Report in CXP Cytometer Software Version 2.0
- Z-0381-06 · Roche Diagnostics Corp. · posted Jan 11, 2006 · Terminated
There is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer.
- Z-0382-06 · Roche Diagnostics Corp. · posted Jan 11, 2006 · Terminated
There is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer.
- Z-0383-06 · Abbott Laboratories Diagnostic Div · posted Jan 11, 2006 · Terminated
These specific protein assays are not meeting the 57-day on board stability claim
- Z-0384-06 · Abbott Laboratories Diagnostic Div · posted Jan 11, 2006 · Terminated
These specific protein assays are not meeting the 57-day on board stability claim
- Z-0385-06 · Abbott Laboratories Diagnostic Div · posted Jan 11, 2006 · Terminated
These specific protein assays are not meeting the 57-day on board stability claim
- Z-0386-06 · General Electric Med Systems LLC · posted Jan 11, 2006 · Terminated
The Estimated Fetal Weight (EFW) result will be incorrect if the operator fails to perform the 'end exam' or new patient registration step before starting an EFW measurement procedure on the next patient.
- Z-0387-06 · General Electric Med Systems LLC · posted Jan 11, 2006 · Terminated
The Estimated Fetal Weight (EFW) result will be incorrect if the operator fails to perform the 'end exam' or new patient registration step before starting an EFW measurement procedure on the next patient.
- Z-0365-06 · Vygon Neuro · posted Jan 10, 2006 · Terminated
stylet in package is shorter than that claimed on the label
- Z-0366-06 · Vygon Neuro · posted Jan 10, 2006 · Terminated
stylet in package is shorter than that claimed on the label
- Z-0367-06 · Vygon Neuro · posted Jan 10, 2006 · Terminated
stylet in package is shorter than that claimed on the label
- Z-0368-06 · Cook, Inc. · posted Jan 10, 2006 · Terminated
The coil is 20 centimeters in length, but the label states the length is 20 millimeters.
- Z-0369-06 · Alsius Corporation · posted Jan 10, 2006 · Terminated
Alsius has found that fluid spills or ingress may result in an alarm condition. The potential health risk involved in this state can be the discontinuance of therapy.
- Z-0370-06 · Medtronic Inc, Neurological & Spinal Division · posted Jan 10, 2006 · Terminated
A safety concern exists regarding the Medtronic Activa Deep Brain Stimulation system instructions for performing MRI (magnetic resonance imaging). Medtronic's instructions have until now recommended a use of a displayed MRI head specific absorption rate…
- Z-0361-06 · Roche Diagnostics Corp. · posted Jan 7, 2006 · Terminated
Potential for hemoglobin derivatives which are outside of the quality control (QC) limits to not engage the auto lock out feature of the system for 02Hb, COHb, HHb, MetHb or SO2.
- Z-0362-06 · Roche Diagnostics Corp. · posted Jan 7, 2006 · Terminated
Potential for hemoglobin derivatives which are outside of the quality control (QC) limits to not engage the auto lock out feature of the system for 02Hb, COHb, HHb, MetHb or SO2.
- Z-0363-06 · Roche Diagnostics Corp. · posted Jan 7, 2006 · Terminated
Potential for hemoglobin derivatives which are outside of the quality control (QC) limits to not engage the auto lock out feature of the system for 02Hb, COHb, HHb, MetHb or SO2.
- Z-0364-06 · Siemens Medical Solutions USA, Inc · posted Jan 7, 2006 · Terminated
When using the Creatine Clearance dose calculations a software bug will allow the algorithm to assume a patients age to be in years when it is entered in months. This miscalculation may result in patients less than two years old to be seriously overdosed…
- Z-0327-06 · Ortho-Clinical Diagnostics · posted Jan 6, 2006 · Terminated
Increased background signal of the sample diluent in the affected Confirmatory Kits could cause some results to be classified as 'Not Confirmed'.
- Z-0338-06 · Smiths Medical ASD, Inc. · posted Jan 6, 2006 · Terminated
The shaft was manufactured approximately 1 to 3 mm under the length specification, and may be 105 mm in length.
- Z-0339-06 · Integra Neuro Sciences · posted Jan 6, 2006 · Terminated
Certain Model NS-P probes contained in the NeuroSensor Cerebral Blood Flow and Intracranial Pressure Monitoring Kit may allow an Intracranial Pressure (ICP) drift of up to 5 mmHg during the first 24 hours of monitoring ICP. This is outside the products…
- Z-0340-06 · Boston Scientific Corporation · posted Jan 6, 2006 · Terminated
Device kits contain B. Braun Locking Tearaway Introducer Systems labeled as 7Fr but may contain 8 Fr sheaths.
- Z-0341-06 · Boston Scientific Corporation · posted Jan 6, 2006 · Terminated
Device kits contain B. Braun Locking Tearaway Introducer Systems labeled as 7Fr but may contain 8 Fr sheaths.
- Z-0342-06 · Boston Scientific Corporation · posted Jan 6, 2006 · Terminated
Device kits contain B. Braun Locking Tearaway Introducer Systems labeled as 7Fr but may contain 8 Fr sheaths.
- Z-0343-06 · Boston Scientific Corporation · posted Jan 6, 2006 · Terminated
Device kits contain B. Braun Locking Tearaway Introducer Systems labeled as 7Fr but may contain 8 Fr sheaths.
- Z-0344-06 · Boston Scientific Corporation · posted Jan 6, 2006 · Terminated
Device kits contain B. Braun Locking Tearaway Introducer Systems labeled as 7Fr but may contain 8 Fr sheaths.
- Z-0345-06 · Oscor Inc. · posted Jan 6, 2006 · Terminated
Sheath size is larger than dilator size causing insertion difficulties.
- Z-0346-06 · B. Braun Medical, Inc. · posted Jan 6, 2006 · Terminated
Tear away sheath assembly may contain a sheath (8FR) that is not the correct French size for the dilator (7FR).
- Z-0347-06 · Baxter Healthcare Renal Div · posted Jan 6, 2006 · Terminated
Potential for fluid or air to be passed through the venous line clamp if the tubing is not centered in the clamp and extends beyond the edge of the clamp's pinch zone. This could result in an air emboli condition.
- Z-0348-06 · Merit Medical Systems, Inc · posted Jan 6, 2006 · Terminated
Dot matrix printer, as a component of a pressure monitoring system, may overheat and pose a fire hazard.
- Z-0349-06 · Merit Medical Systems, Inc · posted Jan 6, 2006 · Terminated
Dot matrix printer, as a component of a pressure monitoring system, may overheat and pose a fire hazard.
- Z-0350-06 · Merit Medical Systems, Inc · posted Jan 6, 2006 · Terminated
Dot matrix printer, as a component of a pressure monitoring system, may overheat and pose a fire hazard.
- Z-0351-06 · Merit Medical Systems, Inc · posted Jan 6, 2006 · Terminated
Dot matrix printer, as a component of a pressure monitoring system, may overheat and pose a fire hazard.
- Z-0352-06 · AGFA Corp. · posted Jan 6, 2006 · Terminated
Users of affected systems may experience a loss of images, the need to repeat the imaging procedure and possibly a short delay in diagnosis.
- Z-0353-06 · AGFA Corp. · posted Jan 6, 2006 · Terminated
Users of affected systems may experience one or more of four (4) problems the clinical impact could be a loss of image(s), the need to repeat the imaging procedure and, possibly, a short delay in diagnosis.
- Z-0354-06 · Respironics California Inc · posted Jan 6, 2006 · Terminated
During product testing of the PLV Continuum Ventilator's Safety Valve/Pressure Relief Valve (SV/PRV) at our manufacturing facility, Respironics found that when the PLY Continuum Ventilator is operated in the 'Face Up' orientation, the SV /PRV may stick in…
- Z-0355-06 · Siemens Medical Solutions USA, Inc · posted Jan 6, 2006 · Terminated
Incomplete installation of the optional Beam Shield may allow unexpected doses of radiation outside the radiation protected areas.
- Z-0356-06 · Sandhill Scientific, Inc · posted Jan 6, 2006 · Terminated
The product is mislabeled. The labeling on the front of the packaging states that product contains latex. However, the back of the package contains a latex- free symbol.
- Z-0357-06 · Sandhill Scientific, Inc · posted Jan 6, 2006 · Terminated
The product is mislabeled. The labeling on the front of the packaging states that product contains latex. However, the back of the package contains a latex- free symbol.
- Gyroscan NT MRI System rev 4.x, 5.x, and 6.x software. System, Nuclear Magnetic Resonance imaging. ↗Z-0358-06 · Philips Medical Systems North America Co. Phillips · posted Jan 6, 2006 · Terminated
The Gyroscan NT MRI system has a computer software problem which results in images from one patient being placed into another patient's image record.
- Z-0359-06 · Sandhill Scientific, Inc · posted Jan 6, 2006 · Terminated
Labeling does not indicate that the product contains latex.