Medical device recalls: November 2008
The FDA posted 138 medical device recall records in November 2008, from 46 recall events by 36 companies. 138 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Medical Solutions USA, Inc · 4 events, 29 records
- Terumo Cardiovascular Systems Corporation · 3 events, 6 records
- Animas Corporation · 2 events, 6 records
- Abbott Laboratories · 2 events, 2 records
- Advanced Sterilization Products · 2 events, 2 records
- Biomet Microfixation, Inc. · 2 events, 2 records
- Biomet, Inc. · 2 events, 2 records
- Kerr Corp · 1 event, 42 records
FDA root cause
- Component design/selection · 45
- Software design · 29
- Device Design · 17
- Nonconforming Material/Component · 15
- Process control · 8
- Other · 5
- Material/Component Contamination · 4
- Labeling mix-ups · 3
Not yet classified (138)
The FDA has posted these records but has not published a class for them yet.
- Z-0326-2009 · Siemens Medical Solutions USA, Inc · posted Nov 28, 2008 · Terminated
Issue may occur where an intended movement of the joystick may unintentionally initiate an x-ray exposure.
- Z-0327-2009 · Siemens Medical Solutions USA, Inc · posted Nov 28, 2008 · Terminated
Issue may occur where an intended movement of the joystick may unintentionally initiate an x-ray exposure.
- Z-0328-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0329-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0330-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0331-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0332-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0333-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0334-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0335-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0336-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0337-2009 · Boston Scientific Corporation · posted Nov 28, 2008 · Terminated
Boston Scientific Corporation initiated a recall of its easy Core" BIOPSY SYSTEM due to difficulty cocking or arming the cannula latch on the device. This difficulty may result in an inability to use the device.
- Z-0415-2009 · Biomet Microfixation, Inc. · posted Nov 28, 2008 · Terminated
Fiber optic cable is missing from the Frederick's Converse Retractor Lot # 042408D08.
- Z-0120-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0121-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0122-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0123-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0124-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0125-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0126-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0127-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0128-2009 · Siemens Medical Solutions USA, Inc · posted Nov 25, 2008 · Terminated
Incorrect contour finding: Image may be calibrated to the wrong size catheter.
- Z-0229-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0230-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0231-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0232-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0233-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0234-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0235-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0236-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0237-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0238-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0239-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0240-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0241-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0242-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0243-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0244-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0245-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0246-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0247-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0248-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0249-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0250-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0251-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0252-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0253-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0254-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0255-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0256-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0257-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0258-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0259-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0260-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0261-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0262-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0263-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0264-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0265-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0266-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0267-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0268-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0269-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0270-2009 · Kerr Corp · posted Nov 25, 2008 · Terminated
Material stiff: Material in several lots of Premise Unidose appears to stiffen and become difficult to extrude over time.
- Z-0290-2009 · Animas Corporation · posted Nov 25, 2008 · Terminated
Unintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure of the device to…
- Z-0291-2009 · Animas Corporation · posted Nov 25, 2008 · Terminated
Unintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure of the device…
- Z-0292-2009 · Animas Corporation · posted Nov 25, 2008 · Terminated
Unintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure of the device to…
- Z-0293-2009 · Animas Corporation · posted Nov 25, 2008 · Terminated
Unintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure of the device…
- Z-0294-2009 · Animas Corporation · posted Nov 25, 2008 · Terminated
Unintentional rebooting: Pump products exhibit an intermittent loss of power due intermittent loss of contact between battery cap and battery canister resulting in the device resetting. The failure of the battery cap may result in failure of the device…
- Z-0308-2009 · Stelkast Co · posted Nov 25, 2008 · Terminated
Five (5) Three Peg Patella SC1591-29 were packaged in containers labeled for SC1591-38 sharing the same lot number.
- Z-0309-2009 · Terumo Cardiovascular Systems Corporation · posted Nov 25, 2008 · Terminated
The pumps may fail to power up or experience unplanned pump stops.
- Z-0310-2009 · Terumo Cardiovascular Systems Corporation · posted Nov 25, 2008 · Terminated
The pumps may fail to power up or experience unplanned pump stops.
- Z-0322-2009 · Arjo, Inc. · posted Nov 24, 2008 · Terminated
The pivot bolt that attaches the hanger bar to the scale of the patient lift can break, resulting in the hanger bar falling.
- Z-0323-2009 · Arjo, Inc. · posted Nov 24, 2008 · Terminated
The pivot bolt that attaches the hanger bar to the scale of the patient lift can break, resulting in the hanger bar falling.
- Z-0004-2009 · Lifescan Inc · posted Nov 21, 2008 · Terminated
Products exceed inaccuracy threshold , with many complaints resulting from inaccurate low results.
- Z-0289-2009 · Biomet, Inc. · posted Nov 21, 2008 · Terminated
The package is properly labeled, but the laser etch for size incorrectly reads "16 X 63/63" instead of the correct size, "16 X 63/67".
- Z-0297-2009 · Abbott Diagnostic International, Ltd. · posted Nov 21, 2008 · Terminated
AxSYM TDM/ Transplant Assay Disks 3D53-03 and higher do not contain the AxSYM Digoxin II Assay file. The instrument might require installation of the AxSYM Digoxin II assay file.
- Z-0299-2009 · Zimmer Inc. · posted Nov 21, 2008 · Terminated
When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.
- Z-0300-2009 · Zimmer Inc. · posted Nov 21, 2008 · Terminated
When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.
- Z-0301-2009 · Zimmer Inc. · posted Nov 21, 2008 · Terminated
When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.
- Z-0302-2009 · Zimmer Inc. · posted Nov 21, 2008 · Terminated
When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.
- Z-0303-2009 · Zimmer Inc. · posted Nov 21, 2008 · Terminated
When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.
- Z-0304-2009 · Zimmer Inc. · posted Nov 21, 2008 · Terminated
When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.
- Z-0305-2009 · Datex Ohmeda, Inc. dba GE Healthcare · posted Nov 21, 2008 · Terminated
Manufacturer of neonatal incubators changed the device labeling to account for new use and care instructions.
- Z-0306-2009 · Abbott Laboratories · posted Nov 21, 2008 · Terminated
Exposed wire resulted in minor electrical shock and burn to the service technician
- Z-0340-2009 · Advanced Sterilization Products · posted Nov 21, 2008 · Terminated
1) Inability of the sterilizer to detect when an injection takes place without hydrogen peroxide being transferred to the vaporizer bowl. This situation can result in a cycle completion with the injection of insufficient hydrogen peroxide sterilant and 2)…
- Z-0217-2009 · Stryker Spine · posted Nov 20, 2008 · Terminated
The Reflex Hybrid 4.0 x 14 mm Variable, Self Tapping Bone Screw, Lot NSZ, was determined to have been anodized with the incorrect color for that screw.
- Z-0272-2009 · Theken Spine LLC · posted Nov 18, 2008 · Terminated
The kits were manufactured in 2005 with syringes that expire on 10/2009 and labeled with an overall expiration on 10/2010.
- Z-0216-2009 · Abbott Laboratories · posted Nov 13, 2008 · Terminated
The values listed in the Architect LH MasterCheck, Lot 82520 data sheet are incorrect. When MasterChecks do not perform as intended, the integrity of the system cannot be verified and patient results would not be generated. Other commercially available…
- Z-0195-2009 · Cardinal Health, Inc · posted Nov 12, 2008 · Terminated
GSI Audera systems requires a system software update due to issues with the Split-Screen and the Vestibular Evoked Myogenic Potential (VEMP) functions. A. Split Screen Field Correction: GSI Audera units with version 2.6 software may mislabel Auditory…
- Z-0214-2009 · Sunquest Information Systems, Inc. · posted Nov 12, 2008 · Terminated
While performing in-house testing for the Sunquest Laboratory System and under certain conditions, there have been instances in which an error that occurred at the database level or originated in the operating system (Cache)was not communicated or…
- Z-0218-2009 · Gebauer Company · posted Nov 12, 2008 · Terminated
Some units of Fluro-Ethyl containing the defective valve which could malfunction and spray refrigerant out from the side of the valve in addition to or instead of spraying in the normal inverted position from the product's actuator. The product can also…
- Z-0287-2009 · Navilyst Medical · posted Nov 9, 2008 · Terminated
Firm became aware that a specific batch of the product may contain loose plastic particulates in the fluid pathway.
- Z-0288-2009 · Biomet, Inc. · posted Nov 9, 2008 · Terminated
The outer package is properly labeled, but the inner package may be labeled as a femoral component.
- Z-0197-2009 · DeRoyal Industries Inc · posted Nov 7, 2008 · Terminated
Surgical kits contained a recalled Endopath ETX 35 mm Endoscopic Linear Cutter White Reload/Cartridges.
- Z-0198-2009 · DeRoyal Industries Inc · posted Nov 7, 2008 · Terminated
Surgical kits contained a recalled Endopath ETX 35 mm Endoscopic Linear Cutter White Reload/Cartridges.
- Z-0208-2009 · Drummond Scientific Co · posted Nov 7, 2008 · Terminated
presence of OSCS contaminant
- Z-0209-2009 · Drummond Scientific Co · posted Nov 7, 2008 · Terminated
presence of OSCS contaminant
- Z-0210-2009 · Drummond Scientific Co · posted Nov 7, 2008 · Terminated
presence of OSCS contaminant
- Z-0211-2009 · Drummond Scientific Co · posted Nov 7, 2008 · Terminated
presence of OSCS contaminant
- Z-0215-2009 · Greiner Bio-One North America, Inc. · posted Nov 7, 2008 · Terminated
Labeling Error: Incorrect tube label reads "Z Serum Clot Activator" instead of "LH Lithium Heparin" tube. All other information on the tube label including color coding of tube was correct. Case and rack label information showed the correct information…
- Z-0275-2009 · Hologic, Inc. · posted Nov 7, 2008 · Terminated
The distal tip of the needle may become detached and remain in the patient, requiring surgical removal.
- Z-0276-2009 · Barcoview · posted Nov 7, 2008 · Terminated
The front protective cover may loosen and completely fall off.
- Z-0280-2009 · C. R.Bard, Inc./BardElectrophysiology Division · posted Nov 7, 2008 · Terminated
Failure to insert the guidewire through the introducer needle.
- Z-0281-2009 · Advanced Sterilization Products · posted Nov 7, 2008 · Terminated
Residual high-level disinfectant solution remaining in endoscopes that have been reprocessed in the ASP Automatic Endoscope Reprocessor. If high-level disinfectant, or other fluids, remains in the endoscope after reprocessing, contact with the mucous…
- Z-0283-2009 · Animas Corporation · posted Nov 7, 2008 · Terminated
display old, inaccurate values for bolus amount delivered and amount of planned bolus totals
- Z-0295-2009 · Terumo Cardiovascular Systems Corporation · posted Nov 7, 2008 · Terminated
The level sensor may not properly couple to the reservoir, resulting in a sensor "not attached" message, an alert or alarm condition, or a failure to detect a low level condition.
- Z-0296-2009 · Terumo Cardiovascular Systems Corporation · posted Nov 7, 2008 · Terminated
The level sensor may not properly couple to the reservoir, resulting in a sensor "not attached" message, an alert or alarm condition, or a failure to detect a low level condition.
- Z-0203-2009 · Terumo Cardiovascular Systems Corporation · posted Nov 6, 2008 · Terminated
The pumps may exhibit blank or distorted displays, loss of local control and/or may have local control knobs that are difficult to use.
- Z-0204-2009 · Terumo Cardiovascular Systems Corporation · posted Nov 6, 2008 · Terminated
The units may exhibit blank or distorted displays, loss of local control and/or may have local control knobs that are difficult to use.
- Z-0274-2009 · Alcon Laboratories, Inc. · posted Nov 6, 2008 · Terminated
Alcon had identified that indications for use unapproved by the Food and Drug Administration are included in the PurePoint Laser System Operator's Manual, Catalog Number 8065751131, Revision B.
- Z-0188-2009 · MAQUET Inc. · posted Nov 5, 2008 · Terminated
Table tilts: The Alphastar Surgical Table piston rod head may break and cause a full inclination of the table top during Trendelenberg positioning if there is insufficient lubrication.
- Z-0213-2009 · GE Healthcare · posted Nov 5, 2008 · Terminated
Issue 1 -Possible failure of CIC Pro audible alarms: Requests for telemetry defaults from Mobile Care Server could cause excessive CIC Pro memory consumption. This excessive memory consumption may lead to potential loss of audio at the CIC or a system…
- Z-0273-2009 · Gyrus ACMI Corporation · posted Nov 5, 2008 · Terminated
Sterility may be compromised
- Z-0284-2009 · Covidien LP · posted Nov 5, 2008 · Terminated
Mislabeled: Package labeled as an insulin syringe for use with U-100 insulin contains an insulin syringe for use with U-40 insulin. Risk of overdose of insulin.
- Z-0133-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0134-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0135-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0136-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0137-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0138-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0139-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0140-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0141-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0142-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0143-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0144-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0145-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0146-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0147-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0148-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Unintended movement of the system table and/or C arm.
- Z-0193-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Patients could possibly fall off the table if not properly secured when: 1) moving them onto the patient table, 2) moving them on the patient table or 3) removing them from the patient table.
- Z-0194-2009 · Siemens Medical Solutions USA, Inc · posted Nov 4, 2008 · Terminated
Patients could possibly fall off the table if not properly secured when: 1) moving them onto the patient table, 2) moving them on the patient table or 3) removing them from the patient table.
- Z-0278-2009 · Nerl Diagnostics LLC · posted Nov 4, 2008 · Terminated
Beverage may contain glass particles.
- Z-0282-2009 · Biomet Microfixation, Inc. · posted Nov 4, 2008 · Terminated
The packaging for the 91-2416, 2.4X16MM HT X-DRIVE SCREW, Lot 493200 actually contained the 99-6577, 2.0X7MM FOSSA X-DRIVE SCREW.
- Z-0131-2009 · Thomas Medical Products Inc · posted Nov 3, 2008 · Terminated
Difficulty inserting guidewire through introducer needle. Needle hub is not tapered down to the needle cannula.
- Z-0201-2009 · VIBE Technologies · posted Nov 3, 2008 · Terminated
Medical device marketed without marketing approval for claims that include cures cancer, infections, depression.
- Z-0212-2009 · Inverness Medical BioStar Inc. · posted Nov 3, 2008 · Terminated
Diagnostic kits for flu were distributed with incorrect components.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.