Medical device recalls: October 2017
The FDA posted 8 medical device recall records in October 2017, from 6 recall events by 6 companies. 3 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Permobil, Ab · 1 event, 2 records
- St Jude Medical Inc. · 1 event, 2 records, 2 Class I
- OriGen Biomedical, Inc. · 1 event, 1 record, 1 Class I
- Sysmex America, Inc. · 1 event, 1 record
- Teleflex Medical · 1 event, 1 record
- W L Gore & Associates, Inc. · 1 event, 1 record
FDA root cause
- Device Design · 4
- Unknown/Undetermined by firm · 1
- Under Investigation by firm · 1
- Process control · 1
- Nonconforming Material/Component · 1
Class I recalls (3)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0021-2018 · OriGen Biomedical, Inc. · posted Oct 20, 2017 · Terminated
Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss).
- Z-0003-2018 · St Jude Medical Inc. · posted Oct 10, 2017 · Terminated
The device may exhibit premature battery depletion.
- Z-0004-2018 · St Jude Medical Inc. · posted Oct 10, 2017 · Terminated
The device may exhibit premature battery depletion.
Class II recalls (5)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0048-2018 · W L Gore & Associates, Inc. · posted Oct 25, 2017 · Terminated
Incomplete and/or partial deployment of the CTAG Device during the endovascular procedure.
- Pilling Knife Handles: ↗Class IIZ-0039-2018 · Teleflex Medical · posted Oct 16, 2017 · Completed
The knife handles slot depth is out of specification, so blades may not fit on the handles properly.
- Sysmex Fluorocell(R) PLT ↗Class IIZ-0045-2018 · Sysmex America, Inc. · posted Oct 16, 2017 · Completed
There is potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%, #) results obtained using the lots listed above may exhibit lower than actual PLT-F and IPF values. The issue may be recognized by a large discrepancy between the…
- Electric Wheelchairs: ↗Class IIZ-0005-2018 · Permobil, Ab · posted Oct 3, 2017 · Terminated
There is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
- Electric Wheelchair bases: ↗Class IIZ-0006-2018 · Permobil, Ab · posted Oct 3, 2017 · Terminated
There is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.