MENTOR CPX 4 Breast Tissue Expander with Suture Tabs 3549315Recalls & reported problems
Mentor Texas L.P. · SILTEX Tall Height Tissue Expander Style 9300, 650cc · Class U · Implant
All models of MENTOR CPX 4 Breast Tissue Expander with Suture Tabs (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00081317006817), received between Sep 6, 2018 and Aug 17, 2025 — classified as 14 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (9 reports).
4 other recalls cover devices cleared under the same FDA submission (K130813); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture9
- Adverse Event Without Identified Device or Use Problem5
- Off-Label Use2
- Inflation Problem1
Patient problems
- Deformity/ Disfigurement6
- No Code Available3
- Seroma2
- Capsular Contracture1
- Local Reaction1
- No Clinical Signs, Symptoms or Conditions1
- Post Operative Wound Infection1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
InjuryAug 17, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST RECONSTRUCTION REVISION SURGERY WITH IMPLANTATION OF A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED WITH A DEFLATED LEFT EXPANDER USING ULTRASOUND IMAGING. AS A RESULT, THE PATIENT UNDERWENT EXCHANGE WITH A MENTOR EXPANDER ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 22824524 ↗InjuryJul 17, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST SURGERY WITH A 650 CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS AND EXPERIENCED A DEFLATION ON THE LEFT SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2025.
Patient problems: Deformity/ Disfigurement
MAUDE report 22532157 ↗InjuryMar 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT A BREAST RECONSTRUCTION PRIMARY WITH MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS 650CC AND EXPERIENCED INFECTION ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT REMOVAL ON (B)(6) 2024.
Patient problems: Post Operative Wound Infection
MAUDE report 18988488 ↗InjuryMar 14, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT A PATIENT WHO HAD A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS PLACED EXPERIENCED LEFT SIDED DEFLATION POSTOPERATIVELY. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Deformity/ Disfigurement
MAUDE report 18906655 ↗InjuryNov 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT BREAST PROSTHESIS IMPLANTATION SURGERY WITH A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS ON THE RIGHT BREAST. POST-OPERATIVELY, THE PATIENT SUFFERED RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE IV). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2023.
Patient problems: Capsular Contracture
MAUDE report 18154683 ↗InjuryJul 17, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT A BREAST RECONSTRUCTION WITH A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS AND EXPERIENCED A DEFLATION ON AN UNKNOWN SIDE POST-OPERATIVELY. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION ON (B)(6) 2023.
Patient problems: Deformity/ Disfigurement
MAUDE report 17334486 ↗MalfunctionMar 21, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS RUPTURED INTRA-OPERATIVELY DURING A BREAST IMPLANTATION SURGERY. NO REPORT OF PATIENT CONSEQUENCE OR ADVERSE EVENT. AS A RESULT, ANOTHER SET OF IMPLANTS WERE IMPLANTED: (LEFT) 550CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS CATALOG: 3549214 LOT: 9744943 SN: (B)(4) AND (RIGHT) 550CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS CATALOG: 3549214 LOT: 9744943 SN: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16584998 ↗InjuryNov 4, 2021Off-Label Use; Material Rupture; Off-Label Use; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT AN UNSPECIFIED BREAST IMPLANTATION PROCEDURE WITH A MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS 650CC AND EXPERIENCED DEFLATION ON THE RIGHT SIDE POST-OPERATIVELY, WHICH WAS CONFIRMED DURING A PHYSICAL EXAM. AS A RESULT, THE PATIENT UNDERWENT REMOVAL AND REPLACEMENT WITH A MENTOR TISSUE EXPANDER (SERIAL NUMBER (B)(4)) ON (B)(6) 2021. IN MENTOR INSTRUCTIONS FOR USE (IFU), IT IS MENTIONED THAT TISSUE EXPANDERS ARE NOT INTENDED FOR IMPLANTATION BEYOND SIX MONTHS. THEREFORE, THIS DEVICE WAS USED IN AN OFF-LABEL MANNER.
Patient problems: Deformity/ Disfigurement
MAUDE report 12753930 ↗InjurySep 29, 2021Off-Label Use; Material Rupture; Off-Label Use; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT WHO UNDERWENT BREAST AUGMENTATION WITH A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS EXPERIENCED DEFLATION POST PROCEDURE. THE DEVICE WAS NOTABLY SMALLER THAN THE CONTRALATERAL SIDE. THERE WAS A VISIBLE HOLE UPON EXPLANT. THE DEFLATION WAS DIAGNOSED THROUGH A PHYSICAL EXAMINATION. AS A RESULT, AN EXPLANT WAS PERFORMED ON (B)(6) 2021.
Patient problems: Deformity/ Disfigurement
MAUDE report 12551012 ↗InjuryJun 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A (B)(6) YEAR OLD FEMALE PATIENT, WHO'S UNDERGONE BREAST PROSTHESIS IMPLANTATION SURGERY WITH A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS ON AN UNSPECIFIED BREAST, SUFFERED PERSISTENT HIGH OUTPUT OF ACUTE INFLAMMATORY LYMPHATIC SECRETION. THE SEROMA HAS REQUIRED PUNCTURE DRAINAGE ONE TO TWO TIMES PER WEEK. THE PATIENT HAD INFLAMMATORY CLINICAL SIGNS THAT MAY CORRESPOND TO AN IMMUNE REACTION TO THE EXPANDER. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Seroma
MAUDE report 11942923 ↗InjuryJun 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A (B)(6) YEAR-OLD FEMALE PATIENT UNDERWENT A BREAST SURGERY PROCEDURE WITH A MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS 650CC AND SUFFERED FROM A LOCAL REACTION ON THE LEFT SIDE. THE PATIENT REPORTED LOCALIZED PAIN WITH A SENSATION OF PRESSURE AROUND THE LEFT TISSUE EXPANDER. AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE.
Patient problems: Local Reaction
MAUDE report 11921455 ↗InjuryNov 25, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A FEMALE PATIENT WHO HAD A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS PLACED ON AN UNKNOWN DATE EXPERIENCED SEROMA POST PROCEDURE. FUTURE EXPLANT IS INDICATED, HOWEVER, AT THE TIME OF THIS REPORT, MENTOR HAS RECEIVED NO INFORMATION REGARDING EXPLANTATION OR AN EXPECTED EXPLANTATION DATE. NO ADDITIONAL EVENT DETAILS HAVE BEEN MADE AVAILABLE TO MENTOR AT THIS TIME. IF ADDITIONAL EVENT DETAILS ARE MADE AVAILABLE IN THE FUTURE, THEN A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: Seroma
MAUDE report 10900571 ↗InjuryOct 22, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT WHO UNDERWENT BREAST RECONSTRUCTION REVISION WITH A 650CC MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS EXPERIENCED DEFLATION OF A TISSUE EXPANDER POST PROCEDURE. AS A RESULT, THE PATIENT HAD TO HAVE THE DEVICE REMOVED AND REPLACED WITH THE PERMANENT DEVICE PRIOR TO THE DATE ORIGINALLY PLANNED.
Patient problems: No Code Available
MAUDE report 9218412 ↗InjuryDec 4, 2018Material Rupture; Material Rupture
IT WAS REPORTED THAT A FEMALE PATIENT UNDERWENT BREAST RECONSTRUCTION WITH MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS 650CC BREAST PROSTHESIS. SLOW LEAK ON ONE SIDE OF THE EXPANDER WHILE BEING FILLED WAS REPORTED. AS A RESULT, THE PATIENT WAS SCHEDULED FOR REMOVAL AND SECOND STAGE IMPLANTS IMPLANTATION IN (B)(6).
Patient problems: No Code Available
MAUDE report 8131603 ↗InjurySep 6, 2018Inflation Problem; Inflation Problem
IT WAS REPORTED THAT A (B)(6)-YEAR-OLD FEMALE PATIENT UNDERWENT AN UNKNOWN PROCEDURE WITH MENTOR CPX 4 BREAST TISSUE EXPANDER WITH SUTURE TABS 650CC IMPLANTS WHICH THE RIGHT SIDE FAILED TO INFLATE AFTER INITIAL INFLATION AND SIX WEEKS AFTER IMPLANTATION. AS A RESULT PATIENT HAD THE EXPANDER REMOVED AND REPLACED WITH ALLERGAN DEVICE AS FOLLOW: CATALOG NUMBER TCM-375, LOT NUMBER 2993976, SERIAL NUMBER (B)(4) ON (B)(6) 2018.
Patient problems: No Code Available
MAUDE report 7852908 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K130813. One submission often covers many models, so check the product and lot details in each record.
- Z-1681-2026 · Feb 18, 2026 · Open, ClassifiedCPX 4 US Breast Tissue Expanders with Suture Tabs Reference Numbers: 354-9113, 3549313, 354-9314 SILTEX Low Height Tissue Expander Style (9100, 450cc), (9300, 450cc), (9300, 550cc)
- Z-2781-2016 · Aug 9, 2016 · TerminatedCPX4 Breast Tissue Expander 650 cc : Product Usage: Tissue expanders can be used for breast reconstruction after mastectomy, correction of an underdeveloped breast, scar revision
- Z-2748-2016 · Aug 9, 2016 · TerminatedMENTOR 2" Tall Height TE With Sutures 450cc PN: 354-9313 The MENTOR CPX4 and CPX4 with Suture Tabs Breast Tissue Expanders are used for breast reconstruction following mastectomy
- Z-2747-2016 · Aug 9, 2016 · TerminatedMENTOR 2" Tall Height TE With Sutures 350cc PN: 354-9312 The MENTOR CPX4 and CPX4 with Suture Tabs Breast Tissue Expanders are used for breast reconstruction following mastectomy
Other EXPANDER, SKIN, INFLATABLE devices
- B.BRAUN 7B3050B. Braun Medical Inc. · EXPANDER, SKIN, INFLATABLE1 recall40 reports
- MENTOR CPX 4 Breast Tissue Expander with Suture Tabs 3549314Mentor Texas L.P. · EXPANDER, SKIN, INFLATABLE1 recall22 reports
- MENTOR CPX4 PLUS, Smooth, Medium Height SCPX146MHMentor Texas L.P. · Expander, skin, inflatable1 recall18 reports
- Natrelle® Style 133 Tissue Expanders with Suture Tabs 133MX-13-TAllergan, Inc. · Expander, Skin, Inflatable1 recall16 reports
- MENTOR ARTOURA PLUS, Smooth, Ultra High Profile SDC135UHMentor Texas L.P. · Expander, skin, inflatable1 recall14 reports
- MENTOR ARTOURA PLUS, Smooth, High Profile SDC150HMentor Texas L.P. · Expander, skin, inflatable1 recall14 reports
- MENTOR ARTOURA PLUS, Smooth, Ultra High Profile SDC120UHMentor Texas L.P. · Expander, skin, inflatable1 recall14 reports
- Natrelle® Style 133 Tissue Expanders with Suture Tabs 133MV-14-TAllergan, Inc. · Expander, Skin, Inflatable1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page