Skip to content

INTRAVIA 2B8019Recalls & reported problems

Baxter International Inc. · INTRAVIA Container, 50 mL Capacity · Class 2

Quick summary

This Baxter INTRAVIA container is used to hold IV fluids. Reports describe it leaking, disconnecting or shedding small pieces or particulate matter into the fluid path, often noticed during compounding or before patient use. There are 39 reports and no injuries or deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 00085412003504), received between Mar 23, 2018 and Sep 11, 2026 — classified as 3 injuries, 42 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (20 reports).

3 other recalls cover devices cleared under the same FDA submission (K964853); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 6 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material20
  • Fluid/Blood Leak17
  • Disconnection3
  • Loose or Intermittent Connection3
  • Crack2
  • Detachment of Device or Device Component2
  • Device Dislodged or Dislocated2
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions34
  • No Patient Involvement10
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • InjurySep 11, 2026Loose or Intermittent Connection; Device Contamination with Chemical or Other Material

    Describe Event, Problem, or Product Use Error: Multiple INTRAVENOUS bags, once spiked, had large membranes come loose from the port and floating in the INTRAVENOUS medication. 3 occurred in the same day and I had been informed that other bags had the same flaw from previous days.ANNEX A: 1371,2944 ANNEX E: 4582 ANNEX F: 4652 ANNEX G: 4755.REFERENCE REPORTS:MW6002230, MW6002231.THIS REPORT CAPTURES IV BAG 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50740251 ↗
  • InjurySep 11, 2026Loose or Intermittent Connection; Device Contamination with Chemical or Other Material

    Describe Event, Problem, or Product Use Error: Multiple INTRAVENOUS bags, once spiked, had large membranes come loose from the port and floating in the INTRAVENOUS medication. 3 occurred in the same day and I had been informed that other bags had the same flaw from previous days.ANNEX A: 1371,2944 ANNEX E: 4582 ANNEX F: 4652 ANNEX G: 4755.REFERENCE REPORTS:MW6002230, MW6002232.THIS REPORT CAPTURES IV BAG 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50740190 ↗
  • InjurySep 11, 2026Loose or Intermittent Connection; Device Contamination with Chemical or Other Material

    Describe Event, Problem, or Product Use Error: Multiple INTRAVENOUS bags, once spiked, had large membranes come loose from the port and floating in the INTRAVENOUS medication. 3 occurred in the same day and I had been informed that other bags had the same flaw from previous days.ANNEX A: 1371,2944 ANNEX E: 4582 ANNEX F: 4652 ANNEX G: 4755.REFERENCE REPORTS:MW6002231, MW6002232.THIS REPORT CAPTURES IV BAG 1.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50739998 ↗
  • MalfunctionSep 10, 2026Device Contamination with Chemical or Other Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721663 ↗
  • MalfunctionSep 10, 2026Device Contamination with Chemical or Other Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721617 ↗
  • MalfunctionSep 1, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50536356 ↗
  • MalfunctionJul 29, 2026Fluid/Blood Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50068887 ↗
  • MalfunctionMar 6, 2026Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THERE WAS PARTICULATE MATTER (PM) IN THE FLUID PATH OF AN UNSPECIFIED QUANTITY OF 50 ML INTRAVIA CONTAINERS. WHEN THE PORT IS PUNCTURED, A SMALL PIECE OF PLASTIC DISLODGES INTO THE BAG. THE FOREIGN MATERIAL APPEARS TO BE A FRAGMENT OF THE PORT INTRODUCED DURING NEEDLE INSERTION. THE PARTICULATE WAS OBSERVED WITHIN THE FLUID PATH AND WAS NOT EMBEDDED IN THE PRODUCT. IT WAS DETECTED AFTER ADDING MEDICATION OR DURING RECONSTITUTION. THE ABRAXANE BAG WAS SUBSEQUENTLY REMADE, AND THE AFFECTED BAG HAS BEEN QUARANTINED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24531391 ↗
  • MalfunctionOct 17, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (2) 50ML INTRAVIA CONTAINERS WERE LEAKING FROM WHERE THE CLEAR AND BLUE TUBING MEET ON THE RIGHT-HANDED TUBE OF THE BAG. THE FIRST LEAK WAS OBSERVED AFTER FILLING AND THE SECOND BAG WAS USED WHICH ALSO LEAKED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23330011 ↗
  • MalfunctionAug 14, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A 50ML INTRAVIA BAG WAS LEAKING. THE LEAK WAS OBSERVED ON THE SIDE OF THE DOSING PORT WHERE THE COLLAR MEETS THE BAG AS WELL AS THE SPACE THE NEEDLE TYPICALLY OCCUPIES UPON DOSING. THIS WAS OBSERVED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22804706 ↗
  • MalfunctionJul 18, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE INJECTION PORT OF FOUR (4) 50 ML CAPACITY INTRAVIA CONTAINERS HAD ISSUES. THE PORT HAD COME OFF THE BAG DURING COMPOUNDING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22546034 ↗
  • MalfunctionMar 12, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A 50ML INTRAVIA BAG LEAKED ON THE SIDE OF THE DOSING (INJECTION) PORT WHERE THE PORT TUBING FROM THE BAG MEETS THE PORT INSERT. THIS WAS OBSERVED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21586871 ↗
  • MalfunctionNov 4, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A 50 ML INTRAVIA CONTAINER LEAKED DUE TO A COMPROMISED SEAL WHERE THE DOSING PORT AND BAG MEET. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20598740 ↗
  • MalfunctionOct 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A 50 ML INTRAVIA CONTAINER LEAKED AND FLUID WAS COMING FROM SIDE PORT SEAL. THIS OCCURRED DURING SHIPPING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20413093 ↗
  • MalfunctionOct 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A 50 ML INTRAVIA CONTAINER LEAKED FROM THE SIDE OF THE INJECTION PORT. THIS WAS OBSERVED DURING INSPECTION/LABELLING/PACKAGING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20408669 ↗
  • MalfunctionOct 9, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A 50 ML INTRAVIA CONTAINER LEAKED WHERE THE PORT IS WELDED TO THE BAG. THIS WAS OBSERVED DURING SHIPPING ACTIVITIES. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20407995 ↗
  • MalfunctionSep 11, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN INTRAVIA 50 ML CAPACITY CONTAINER HAD FOREIGN PARTICULATE MATTER. FTIR (FOURIER TRANSFORM INFRARED SPECTROSCOPY) IDENTIFIED WADDING MATERIAL AND PAPER PARTICULATE INSIDE THE FINISHED PRODUCT THAT THIS DEVICE WAS USED WITH DURING THE COMPOUNDING PROCESS. THE PARTICULATES WERE OBSERVED DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20195862 ↗
  • MalfunctionSep 11, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN INTRAVIA 50 ML CAPACITY CONTAINER HAD FOREIGN PARTICULATE MATTER. FTIR (FOURIER TRANSFORM INFRARED SPECTROSCOPY) IDENTIFIED WOOL PARTICULATE INSIDE THE FINISHED PRODUCT THAT THIS DEVICE WAS USED WITH DURING THE COMPOUNDING PROCESS. THE PARTICULATES WERE OBSERVED DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20194603 ↗
  • MalfunctionApr 18, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT FOREIGN PARTICULATE MATTER (PM) WAS OBSERVED IN TWO (2) 50 ML INTRAVIA CONTAINERS. THE PM WAS FURTHER DESCRIBED AS AN EXTRINSIC PARTICLE ¿POLYESTER FIBER AND WOOL¿ AND ¿DWG OIL¿. THIS WAS IDENTIFIED VIA FOURIER TRANSFORM INFRARED SPECTROSCOPY (FTIR) DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139332 ↗
  • MalfunctionApr 18, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT FOREIGN PARTICULATE MATTER (PM) WAS OBSERVED IN A 50 ML INTRAVIA CONTAINER. THE PM WAS FURTHER DESCRIBED AS AN EXTRINSIC PARTICLE ¿GABRIELLE¿. THIS WAS IDENTIFIED VIA FOURIER TRANSFORM INFRARED SPECTROSCOPY (FTIR) DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19139183 ↗
  • MalfunctionApr 10, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT FOREIGN PARTICULATES WERE OBSERVED IN A 50 ML INTRAVIA CONTAINER. THE FOREIGN PARTICULATES WERE DESCRIBED AS EXTRINSIC PARTICLES THAT WERE DISCOVERED DURING FOURIER TRANSFORM INFRARED SPECTROSCOPY (FTIR) TESTING PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19084492 ↗
  • MalfunctionApr 10, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT FOREIGN PARTICULATES WERE OBSERVED IN A 50 ML INTRAVIA CONTAINER. THE FOREIGN PARTICULATES WERE DESCRIBED AS EXTRINSIC PARTICLES THAT WERE DISCOVERED DURING FOURIER TRANSFORM INFRARED SPECTROSCOPY (FTIR) TESTING PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19084490 ↗
  • MalfunctionDec 22, 2023Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THAT A 50 ML INTRAVIA CONTAINER HAD PUNCTURE HOLES IN THE PORT. THIS WAS OBSERVED WHEN THE BAG WAS ATTACHED DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18384672 ↗
  • MalfunctionNov 21, 2023Detachment of Device or Device Component; Material Split, Cut or Torn; Detachment of Device or Device Component; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE PORTS OF AN UNSPECIFIED QUANTITY OF 50 ML CAPACITY INTRAVIA CONTAINERS WERE STARTING TO FRAY. IT WAS FURTHER STATED THAT THERE WAS AN INCIDENT WHERE THE PORT COMPLETELY CAME OFF AFTER BEING PUNCTURED. THIS WAS DISCOVERED DURING SET-UP/PREPARATION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18183420 ↗
  • MalfunctionJul 31, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A 50ML INTRAVIA CONTAINER WAS LEAKING FROM THE TUBING. THE LEAK WAS DISCOVERED BY THE CUSTOMER UPON RECEIPT OF THE SHIPMENT DURING THEIR ORDER RECONCILIATION BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17425735 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K964853. One submission often covers many models, so check the product and lot details in each record.

  • Z-0485-2018 · Jul 27, 2017 · Terminated
    Empty IntraVia Container with Non-DEHP Fluid Path, 500mL Containers are intended for use in the preparation and administration of drug admixtures.
  • Z-0084-2015 · Sep 16, 2014 · Terminated
    Baxter 2B8011 Empty INTRA VIA Container with PVC Ports, Sterile fluid path, 150 ml, Admixture Products and Accessories. The lntraVia Empty Plastic Container is intended for use i
  • Z-0085-2015 · Sep 16, 2014 · Terminated
    Baxter 2B8013 Empty INTRA VIA Container with PVC Ports (Sterile, nonpyrogenic fluid path) ,500 ml. Admixture Products and Accessories The lntraVia Empty Plastic Container is in

Other CONTAINER, I.V. devices

All CONTAINER, I.V. devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.