NA 2C5641Recalls & reported problems

Baxter International Inc. · Extension Set, Male Luer Lock Adapter with Retractable Collar · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00085412003818), received between May 2, 2019 and Dec 17, 2025 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material6
  • Fluid/Blood Leak4
  • Loose or Intermittent Connection4
  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Patient Involvement4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 17, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN EXTENSION SET HAD PLASTIC FILM INSIDE THE FEMALE PORT. THIS WAS OBSERVED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23833946 ↗
  • MalfunctionNov 26, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT AN EXTENSION SET HAD A POLYPROPYLENE PARTICLE. THIS WAS OBSERVED DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23661094 ↗
  • MalfunctionAug 11, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXTENSION SET WAS LEAKING FROM THE LUER LOCK. THIS ISSUE WAS OBSERVED WHILE CONNECTED TO A BAG OF BLOOD/HUMAN CELLS PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22772399 ↗
  • MalfunctionJun 26, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THERE WAS PARTICULATE MATTER (PM) CONTAMINATION OBSERVED IN AN EXTENSION SET. THE PM WAS DESCRIBED AS, ¿POLYPROPYLENE PARTICLE WAS OBSERVED IN THE FINAL PRODUCT BAG BY USING THE GLOBAL EXTENSION SET¿. THIS WAS OBSERVED BEFORE USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22341206 ↗
  • MalfunctionJun 7, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THREE (3) EXTENSION SETS LEAKED FROM THE FEMALE LUER CONNECTION WITH THE BLUE CAP. LIQUID WAS COMING AROUND THE SIDES OF THE CAP. THIS WAS IDENTIFIED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17080571 ↗
  • MalfunctionMay 25, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF EXTENSION SETS HAD CONNECTION ISSUES; FURTHER DESCRIBED AS "WOULD NOT MAKE GOOD SEALS WHEN ATTEMPTING TO CONNECT VIA TUBE SEALERS". THE ISSUE WAS IDENTIFIED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11876403 ↗
  • MalfunctionMay 25, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF EXTENSION SETS HAD CONNECTION ISSUES; FURTHER DESCRIBED AS "WOULD NOT MAKE GOOD SEALS WHEN ATTEMPTING TO CONNECT VIA TUBE SEALERS". THE ISSUE WAS IDENTIFIED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11876389 ↗
  • MalfunctionMay 25, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF EXTENSION SETS HAD CONNECTION ISSUES; FURTHER DESCRIBED AS "WOULD NOT MAKE GOOD SEALS WHEN ATTEMPTING TO CONNECT VIA TUBE SEALERS". THE ISSUE WAS IDENTIFIED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11876319 ↗
  • MalfunctionMay 25, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF EXTENSION SETS HAD CONNECTION ISSUES; FURTHER DESCRIBED AS "WOULD NOT MAKE GOOD SEALS WHEN ATTEMPTING TO CONNECT VIA TUBE SEALERS". THE ISSUE WAS IDENTIFIED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11876299 ↗
  • MalfunctionSep 15, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A RED FIBER PARTICULATE WAS OBSERVED IN THE EXTENSION SET. IT WAS FURTHER REPORTED THAT THE PARTICULATE WAS FOUND TO BE AN ACRYLIC. THIS ISSUE WAS IDENTIFIED DURING CHECK. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Patient Involvement

    MAUDE report 10535345 ↗
  • MalfunctionSep 15, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT A RED FIBER PARTICULATE WAS OBSERVED IN THE EXTENSION SET. IT WAS FURTHER REPORTED THAT THE PARTICULATE WAS FOUND TO BE AN ACRYLIC. THIS ISSUE WAS IDENTIFIED DURING CHECK. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Patient Involvement

    MAUDE report 10535342 ↗
  • MalfunctionJun 7, 2019Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT AN EXTENSION SET WAS CRACKED WHICH RESULTED IN A LEAK. DURING FILLING, IT WAS DISCOVERED THAT THE FEMALE EXTENSION SET CONNECTED TO THE MALE STOPCOCK ASSEMBLY WAS CRACKED WHICH CAUSED A LEAK. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8680499 ↗
  • MalfunctionMay 29, 2019Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT PARTICULATE MATTER WAS OBSERVED DURING USE OF AN EXTENSION SET. THE PARTICULATE MATTER WAS NOTED IN A ¿CRYOBAG¿ (NON-BAXTER PRODUCT) ATTACHED TO THE SET. THIS WAS IDENTIFIED IN THE FINAL PRODUCT PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Patient Involvement

    MAUDE report 8652285 ↗
  • MalfunctionMay 2, 2019Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED ONE EXTENSION SET CRACKED, CAUSING A LEAK. THE CRACK WAS FURTHER DESCRIBED AS " THE FEMALE EXTENSION SET CONNECTED TO THE MALE STOPCOCK ASSEMBLY WAS DISCOVERED TO BE CRACKED". THIS WAS NOTED DURING FILLING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Patient Involvement

    MAUDE report 8575066 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →