Skip to content

NA 2C6226Recalls & reported problems

Baxter International Inc. · Extension Set, Male Luer Lock Adapter · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00085412003900), received between May 20, 2024 and Aug 14, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Remove Packaging Material” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Remove Packaging Material3
  • Detachment of Device or Device Component1
  • Disconnection1
  • Failure to Infuse1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2026Difficult to Open or Remove Packaging Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50262995 ↗
  • MalfunctionAug 14, 2026Difficult to Open or Remove Packaging Material

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50262885 ↗
  • InjuryApr 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED AN EXTENSION SET SEPARATED DURING SURGERY. DURING THE INDUCTION OF ANESTHESIA, A PATIENT WAS RECEIVING PROPOFOL AND ROCURONIUM WHEN THE EXTENSION SET SEPARATED; FURTHER REPORTED AS "THE TUBING APPEARED TO FALL OUT OF THE MALE PORTION OF THE EXTENSION TUBING". THE SURGICAL TEAM WAS UNABLE TO ADMINISTER MEDICATIONS THROUGH THE EXTENSION SET DUE TO THIS MALFUNCTION; FURTHERMORE, DUE TO THE LOCATION OF THE SEPARATION, IT WAS NOT POSSIBLE TO CONNECT A NEW INTRAVENOUS LINE TO THE EXTENSION SET. IT WAS ALSO REPORTED THAT THE ADAPTER WAS TOO TIGHT TO DISENGAGE FROM THE ANGIOCATH THAT IT WAS ATTACHED TO. AS A RESULT, A NEW INTRAVENOUS LINE WAS QUICKLY ESTABLISHED WHILE THE PATIENT¿S AIRWAY WAS MAINTAINED VIA BAG AND MASK. THE SURGERY WAS COMPLETED WITHOUT FURTHER ISSUE. NO FURTHER INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21762132 ↗
  • MalfunctionDec 17, 2024Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT THE STERILE PACKAGE OF AN EXTENSION SET WAS DIFFICULT TO OPEN. THIS ISSUE WAS DESCRIBED AS, ¿WHEN THEY TRYING TO OPEN, THE PACKAGE BEGINS RIPPING. APPARENTLY, THE SEAL IS TOO STRONG, SHREDDING THE PACKAGE.¿ THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20952452 ↗
  • MalfunctionMay 20, 2024Disconnection; Fluid/Blood Leak; Failure to Infuse; Disconnection; Fluid/Blood Leak; Failure to Infuse

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF EXTENSION SETS LEAKED. THE EVENT OCCURS AFTER THE USER ATTACHED THE EXTENSION TUBING TO THE PATIENT¿S PERIPHERAL IV (CATHETER HUB). THE IV TUBING SEPARATES FROM THE CATHETER HUB AND SUBSEQUENTLY LEAKS AT THE SEPARATION OF THE CONNECTORS. THIS LEAK WOULD THEN PREVENT THE FULL VOLUME OF MEDICATION FROM BEING INFUSED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19352208 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates