NA 2N3350Recalls & reported problems

Baxter International Inc. · Non-DEHP MICRO-VOLUME Extension Set, 0.2 Micron Downstream Filter, Male Luer Lock Adapter · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Baxter Micro-Volume extension set is a short tubing line with a built-in 0.2 micron filter used for drug infusions. The 32 reports mostly describe fluid or blood leaking from the line — often at the air vent or through a cracked filter — plus lines that would not flush, tubing that cracked, and filter alarms on Baxter pumps. No deaths and no injuries were reported, and reports rose from 4 to 9 over the past 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00085412004846), received between Mar 23, 2018 and Aug 31, 2026 — classified as 32 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 3 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak25
  • Crack6
  • No Flow3
  • Failure to Prime2
  • Protective Measures Problem2
  • Backflow1
  • Break1
  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 31, 2026Fluid/Blood Leak

    It was reported that an unspecified quantity of Micro-Volume Extension Sets intermittently leaked through the air vent filter. The reporter stated that it was a continuous leakage.The process step at which the issue occurred was not specified. There was no report of patient injury or medical intervention associated with these events. No additional information is available at this time.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50528289 ↗
  • MalfunctionJun 5, 2026Device-Device Incompatibility; No Flow; Protective Measures Problem

    IT WAS REPORTED THAT UNSPECIFIED UNITS OF NON DEHP MICRO VOLUME EXTENSION SETS CAUSED DOWNSTREAM OCCLUSION ALARMS ON THE BAXTER PUMP. THE TUBING WAS INSPECTED FOR UNCLAMPED CLAMPS AND KINKS; NO ISSUES WERE IDENTIFIED. THE PUMP WAS RESTARTED; HOWEVER, STILL OBSERVED DOWNSTREAM OCCLUSION ALARMS. IT WAS SUBSEQUENTLY DETERMINED THAT THE 0.2 MICRON FILTER WAS NOT COMPATIBLE WITH THE PUMP. A NEW BAG OF PACLITAXEL WAS PREPARED WITH A DIFFERENT 0.2 MICRON FILTER, AFTER WHICH THE PUMP FUNCTIONED WITHOUT OCCLUSION ALARMS. THE PREVIOUS DAY, SIMILAR OCCLUSION ALARMS OCCURRED DURING PACLITAXEL ADMINISTRATION WHEN THE SAME FILTER WAS CONNECTED TO THE SHORT SET. THE FILTER WAS REMOVED AND REPOSITIONED ON THE FLUSH LINE CLOSER TO THE PATIENT AND THE INFUSION CONTINUED WITHOUT ISSUES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25410845 ↗
  • MalfunctionMar 4, 2026Failure to Prime; No Flow

    IT WAS REPORTED THAT AN UNSPECIFIED AMOUNT OF MICRO-VOLUME EXTENSION SETS WERE UNABLE TO BE FLUSHED. THIS WAS OBSERVED SETUP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24512148 ↗
  • MalfunctionMar 4, 2026Fluid/Blood Leak; Failure to Prime

    IT WAS REPORTED THAT AN UNSPECIFIED AMOUNT OF MICRO-VOLUME EXTENSION SETS LEAKED FROM AN UNSPECIFIED LOCATION. THIS WAS OBSERVED DURING SETUP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24506317 ↗
  • MalfunctionJan 8, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT A MICRO-VOLUME EXTENSION SET LEAKED AT THE FILTER. THIS WAS OBSERVED DURING SETUP. THE DEVICE CONTAINED BENADRYL AND NORMAL SALINE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24016591 ↗
  • MalfunctionJan 5, 2026Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT A MICRO-VOLUME EXTENSION SET LEAKED FROM A CRACKED FILTER. THIS WAS OBSERVED WHILE RUNNING ANTIBIOTIC. THE LINE WAS CHANGED MIDWAY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23968894 ↗
  • MalfunctionNov 17, 2025Break; Break

    IT WAS REPORTED THAT THE MEDIPORT NEEDLE OF A MICRO-VOLUME EXTENSION SET BROKE DURING PATIENT INFUSION. THE DEVICE CONTAINED IVF AND MESNA. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23576825 ↗
  • MalfunctionNov 13, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED FROM THE AIR VENT. THE ISSUE OCCURRED AFTER PRIMING. THE DEVICE WAS ATTACHED TO A PERIPHERALLY INSERTED CENTRAL CATHETER (PICC). THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23546248 ↗
  • MalfunctionNov 12, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED FROM THE AIR-VENT. THE ISSUE OCCURRED AFTER PRIMING. THE DEVICE WAS ATTACHED TO A PERIPHERALLY INSERTED CENTRAL CATHETER (PICC). THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23538903 ↗
  • MalfunctionOct 21, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT A MICRO-VOLUME EXTENSION SET LEAKED FROM A CRACK IN THE FILTER. THIS WAS OBSERVED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23348095 ↗
  • MalfunctionJul 11, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF NON-DEHP MICRO-VOLUME EXTENSION SETS LEAKED THROUGH THE AIR VENT HOLE. THE ISSUE OCCURRED WHEN THE LINE WAS PRIMED USING A PUMP OR MANUALLY WHEN THEY ARE CHANGING MEDICATIONS(IN PARTICULAR, SEDATION/ANALGESIA) OR FLUSHES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22480809 ↗
  • MalfunctionMay 19, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (2) MICRO-VOLUME EXTENSION SETS LEAKED AT THE FILTER PRIOR TO THERAPY. A SYRINGE WAS USED TO PRIME THE TUBING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22047194 ↗
  • MalfunctionJan 17, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED AT THE FILTER DURING USE. THE FILTER WAS LEAKING EVEN WHEN MANUALLY FLUSHED. THE DEVICE CONTAINED FAMOTIDINE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21178508 ↗
  • MalfunctionSep 13, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF MICRO-VOLUME EXTENSION SETS LEAKED FROM THE FILTER. THE LEAK OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20214505 ↗
  • MalfunctionSep 10, 2024Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT WERE FLOW ISSUES THROUGH AN UNSPECIFIED QUANTITY OF MICRO-VOLUME EXTENSION SETS. THERE WERE OCCLUSIONS WHICH RESULTED IN DIFFICULTIES WITH FLUSHING LIPIDS THROUGH. THIS OCCURRED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20185528 ↗
  • MalfunctionSep 10, 2024Backflow; Backflow

    IT WAS REPORTED THAT A MICRO-VOLUME EXTENSION SET HAD LIPIDS BACKING UP IN THE TUBING. THE BACKFLOW WAS CAUSING THE OCCLUSION LIMIT TO GO OFF ON THE PUMP. THIS OCCURRED DURING INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20180310 ↗
  • MalfunctionJan 29, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED FROM THE FILTER. THIS OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18597249 ↗
  • MalfunctionJan 29, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED FROM THE FILTER. THE SET WAS CONNECTED TO AN UPPER EXTREMITY PEDIATRIC PICC (PERIPHERALLY INSERTED CENTRAL CATHETER) LINE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18596799 ↗
  • MalfunctionJan 29, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET WAS CRACKED AND LEAKED AT THE CIRCULAR FILTER OF THE LINE CONNECTED TO THE INTERNAL JUGULAR CATHETER PORT. NO OTHER MEDICATIONS WERE RUNNING THROUGH THE PORT DUE TO ACCESS AVAILABILITY FOR DRIPS AND ALPROSTADIL AND TPN (TOTAL PARENTERAL NUTRITION) LIPIDS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18596170 ↗
  • MalfunctionDec 21, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A MICRO-VOLUME EXTENSION SET LEAKED AT THE FILTER. THIS OCCURRED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18372747 ↗
  • MalfunctionDec 18, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED AT THE FILTER. THIS OCCURRED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18345589 ↗
  • MalfunctionOct 11, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THREE (3) NON-DEHP MICRO-VOLUME EXTENSION SETS LEAKED. THIS ISSUE WAS IDENTIFIED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17916027 ↗
  • MalfunctionOct 5, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED AROUND THE FILTERED AREA. THE ISSUE WAS IDENTIFIED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17875966 ↗
  • MalfunctionAug 23, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE FILTER OF A NON-DEHP MICRO-VOLUME EXTENSION SET LEAKED. IT WAS NOT SPECIFIED WHEN IN THE PROCESS STEP THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17605927 ↗
  • MalfunctionJul 31, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT THE FILTER OF THE NON-DEHP MICRO-VOLUME EXTENSION SET WAS CRACKED ON THE BONDED SIDE WHERE THE IV TUBING WAS CONNECTED AND LEAKED OUT OF THE AIR VENT. THE ISSUE OCCURRED AFTER PRIMING. THE SET WAS USED WITH A MEDFUSION SYRING PUMP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17425700 ↗

FDA clearance and classification

Cleared through Special 510(k) K113227 on Dec 8, 2011 (decision: Substantially Equivalent). Submitted by Baxter Healthcare Corp as “Non-Dehp Micro-Volume Extension Set With 0.22 Micron Filter”. The FDA review took 37 days. FDA record ↗

The FDA regulates product code FPB (filter, infusion line) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 2N3350

Has the NA 2N3350 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00085412004846), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 2N3350?

The FDA's MAUDE database lists 32 adverse event reports for this device, received between Mar 23, 2018 and Aug 31, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (25), “Crack” (6) and “No Flow” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 2N3350?

None of the reports are classified as deaths, 0 as injuries and 32 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 2N3350 FDA approved?

It was cleared through Special 510(k) K113227 on Dec 8, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NA 2N3350, and is it still sold?

The FDA's device database lists Baxter International Inc. as the company responsible for it (brand NA, model 2N3350). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.