INTERLINK 1C8740Recalls & reported problems

Baxter International Inc. · Non-DEHP Minivolume Extension Set, Injection Site, 0.2 Micron Downstream Filter, Male Luer Lock Adapter · Class 2

All models of INTERLINK (17)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00085412082028), received between Jan 25, 2019 and May 14, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak9
  • Detachment of Device or Device Component2
  • Material Deformation1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE LUER LOCK ADAPTER OF AN INTERLINK NON-DEHP MINIVOLUME EXTENSION SET SEPARATED FROM THE TUBING. IT WAS OBSERVED THAT THE DISTAL END OF THE ¿CARDIAC FILTER TUBING¿ REMAINED ATTACHED TO THE PATIENT¿S CVD RED LUMEN WHILE THE REMAINDER OF THE TUBING HAD BROKEN OFF. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25178676 ↗
  • MalfunctionApr 29, 2026Fluid/Blood Leak; Material Deformation

    IT WAS REPORTED THAT A NON-DEHP MINIVOLUME EXTENSION SET LEAKED FROM THE "RUBBER PORT ON THE MED LINE FILTER" (INTERLINK COMPONENT) DURING A MAGNESIUM INFUSION; THE RUBBER STOPPER WAS "HALF OUT". THE PATIENT STILL RECEIVED A MAJORITY OF THE INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25033825 ↗
  • MalfunctionFeb 9, 2026Fluid/Blood Leak; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE RUBBER HUB AT THE Y-CONNECTOR OF AN INTERLINK NON-DEHP MINIVOLUME EXTENSION SET WAS DISPLACED WHICH RESULTED IN A FLUID LEAK. THIS ISSUE WAS OBSERVED WHILE PRIMING THE SET PRIOR TO PATIENT CONNECTION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24299174 ↗
  • MalfunctionSep 15, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE NEEDLE ACCESS POINT OF A MICRO-VOLUME EXTENSION SET HAD A DEFECTIVE RUBBER PORTION AND LEAKED. THIS WAS OBSERVED DURING USE. A PEDI PICC (PERIPHERALLY INSERTED CENTRAL CATHETER) WAS USED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23057665 ↗
  • MalfunctionSep 9, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN INTERLINK SYSTEM MINIVOLUME EXTENSION SET LEAKED FROM THE AIR FILTER TUBING. THIS WAS OBSERVED PRIOR TO USE FOR A METHYLENE INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23009497 ↗
  • MalfunctionAug 4, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN INTERLINK NON-DEHP MINIVOLUME EXTENSION SET WAS LEAKING FROM THE FILTER. THE LEAK FROM THE FILTER WAS OBSERVED WHILE FLUSHING THE SET WITH NORMAL SALINE AFTER THE PATIENT HAD RECEIVED AN ELECTROLYTE REPLETION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22699038 ↗
  • MalfunctionOct 7, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (2) INTERLINK SYSTEM NON-DEHP MINIVOLUME EXTENSION SETS LEAKED FROM AN UNSPECIFIED LOCATION AFTER PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20385133 ↗
  • MalfunctionOct 7, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN INTERLINK SYSTEM NON-DEHP MINIVOLUME EXTENSION SET LEAKED AFTER PRIMING. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20384751 ↗
  • MalfunctionFeb 20, 2020Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF INTERLINK SYSTEM NON-DEHP MINIVOLUME EXTENSION SETS WERE LEAKING FROM A HOLE IN THE FILTER. THE LEAKS WERE OBSERVED A FEW HOURS LATER AFTER THE PATIENT WAS CONNECTED FOR THERAPY. THIS ISSUE WAS NOT OBSERVED DURING PRIMING. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9734936 ↗
  • MalfunctionJan 25, 2019Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENTS. IT WAS REPORTED THAT AN INTERLINK MICRO-VOLUME EXTENSION SET WAS LEAKING FROM AN UNSPECIFIED LOCATION. IT WAS NOT SPECIFIED IF A PATIENT WAS INVOLVED IN THE EVENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Information

    MAUDE report 8280044 ↗

FDA clearance and classification

Cleared through Special 510(k) K113227 on Dec 8, 2011 (decision: Substantially Equivalent). Submitted by Baxter Healthcare Corp as “Non-Dehp Micro-Volume Extension Set With 0.22 Micron Filter”. The FDA review took 37 days. FDA record ↗

The FDA regulates product code FPB (filter, infusion line) under 21 CFR 880.5440 as a Class II device. 21 CFR 880.5440 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the INTERLINK 1C8740

Has the INTERLINK 1C8740 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00085412082028), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the INTERLINK 1C8740?

The FDA's MAUDE database lists 10 adverse event reports for this device, received between Jan 25, 2019 and May 14, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (9), “Detachment of Device or Device Component” (2) and “Material Deformation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the INTERLINK 1C8740?

None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the INTERLINK 1C8740 FDA approved?

It was cleared through Special 510(k) K113227 on Dec 8, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the INTERLINK 1C8740, and is it still sold?

The FDA's device database lists Baxter International Inc. as the company responsible for it (brand INTERLINK, model 1C8740). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Filter, infusion line devices

All Filter, infusion line devices →