Skip to content

INTERLINK 2N3379Recalls & reported problems

Baxter Healthcare Corporation · Injection Site · Class 2

All models of INTERLINK (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00085412004945), received between Jan 6, 2025 and Sep 16, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Crack” (4 reports).

26 other recalls cover devices cleared under the same FDA submission (K925126); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 5 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack4
  • Fluid/Blood Leak2
  • Break1
  • Incomplete or Inadequate Connection1
  • No Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2025Break; Break

    IT WAS REPORTED THAT AN INTERLINK SYSTEM INJECTION SITE HAD STRESS LINES. THIS WAS OBSERVED DURING VISUAL INSPECTION PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23066668 ↗
  • MalfunctionSep 15, 2025Crack; Crack

    IT WAS REPORTED THAT AN INTERLINK SYSTEM INJECTION SITE HAD A CRACK ON THE UPPER PART OF THE PLASTIC CASING. THIS WAS OBSERVED OUT OF THE BOX PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23058793 ↗
  • MalfunctionSep 5, 2025Crack; Crack

    IT WAS REPORTED THAT AN INTERLINK SYSTEM INJECTION SITE HAD STRESS LINES. THE ISSUE WAS FOUND BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22986525 ↗
  • MalfunctionSep 5, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (02) INTERLINK SYSTEM INJECTION SITES WERE FOUND TO HAVE CRACKS ON THE UPPER PART OF THE PLASTIC CASING OF THE SEPTUM, AROUND THE PIERCING. THE CRACKS RESULTED IN LEAK. THE ISSUE WAS FOUND DURING SET-UP; HOWEVER, MEDICATION WAS NOT INVOLVED AT THIS STAGE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22983862 ↗
  • MalfunctionSep 5, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THAT FOUR (4) INTERLINK SYSTEM INJECTION SITES WERE FOUND TO HAVE CRACKS ON THE UPPER PART OF THE PLASTIC CASING OF THE SEPTUM AROUND THE PIERCING, WHICH RESULTED IN A LEAK. THE ISSUE WAS FOUND DURING SET-UP; HOWEVER, MEDICATION WAS NOT INVOLVED AT THIS STAGE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22983706 ↗
  • MalfunctionJan 6, 2025No Flow; Incomplete or Inadequate Connection; No Flow; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT THE FLUID DID NOT FLOW THROUGH AN INTERLINK SYSTEM INJECTION SITE. THE ISSUE WAS FURTHER DESCRIBED AS, THE NEEDLE WAS UNABLE TO PIERCE THE INJECTION SITE WHICH LED TO NO FLOW. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21082280 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K925126. One submission often covers many models, so check the product and lot details in each record.

  • Z-1269-2013 · Mar 28, 2013 · Terminated
    A) Product Code 2N3371: Non-DEHP Y-Type Catheter Extension Set, 5.7", Vol. 0.8 mL, 2 Injection Sites, Male Luer Lock Adapter, Single use only Product Usage: This device is ind
  • Z-0471-2012 · Dec 6, 2011 · Terminated
    Baxter Interlink System Extension Set, 20" (51 cm), Vol. 2.9 mL, Injection Site, Male Luer Lock Adapter; an Rx sterile, nonpyrogenic fluid pathway for the administration of sterile
  • Z-2837-2011 · Jul 6, 2011 · Terminated
    Baxter Interlink System Non-DEHP I.V. Catheter Extension Set, 6.6(17 cm), Volume 0.8 mL, Injection Site, Male Luer Lock Adapter; an Rx sterile, single use nonpyrogenic fluid pathw
  • Z-0629-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Non-DEHP Standard Bore 3-Lead Catheter Extension Set, product code 2N3335; An Rx sterile, nonpyrogenic fluid pathway with 2 Interlink injection sites and a
  • Z-0644-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Minivolume Extension Set, product code 1C8643; An Rx sterile, nonpyrogenic fluid pathway, 7'', 0.3 mL Vol, with Male Luer Slip T-Connector; Baxter Healthcar
  • Z-0652-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Minivolume Extension Set, product code 1C8740; An Rx sterile, nonpyrogenic fluid pathway, 16.0'', 1.0 mL Vol., with injection site, 0.22 Micron Downstream F
  • Z-0630-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System T-Connector Extension Set, product code 2N3343; An Rx sterile, nonpyrogenic fluid pathway, 7.0'', 1.0 mL Vol., with 1 Interlink T-connector injection sites
  • Z-0638-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Minivolume Extension Set, product code 1C8378; An Rx sterile, nonpyrogenic fluid pathway, 9.0'', 0.6 mL Vol., with injection site and male Luer Lock adapter
  • Z-0640-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Minivolume Extension Set, product code 1C8412; An Rx sterile, nonpyrogenic fluid pathway, 16.0'', 1.0 mL Vol., with injection site, 0.22 Micron Downstream F
  • Z-0634-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Non-DEHP I.V. Catheter Extension Set, product code 2N3384; An Rx sterile, nonpyrogenic fluid pathway, 9.0'', 0.5 mL Vol., with injection site and male Luer
  • Z-0648-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Y-Type Minivolume Extension Set, product code 1C8645; An Rx sterile, nonpyrogenic fluid pathway, 9.0'', 0.8 mL Vol., with 2 injection sites and male Luer Lo
  • Z-0641-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Minivolume Extension Set, product code 1C8462; An Rx sterile, nonpyrogenic fluid pathway, 16.0'', 4.2 mL Vol., with 2 injection sites, 0.22 Micron Downstrea
  • Z-0637-06 · Dec 15, 2005 · Terminated
    Baxter Interlink System Minivolume Extension Set, product code 1C8290; An Rx sterile, nonpyrogenic fluid pathway, 11.0'', 0.86 mL Vol., with 2 injection sites and male Luer Lock ad
  • Z-1424-05 · Aug 3, 2005 · Terminated
    Baxter Interlink System Non-DEHP Y-Type Catheter Extension Set, product code 2N3371; An Rx sterile, nonpyrogenic fluid pathway, 5.7'', 0.8 mL Vol., with 2 injection sites and a mal
  • Z-1102-05 · Jul 6, 2005 · Terminated
    Baxter Interlink System Y-Type Minivolume Extension Set, product code 1C8645; 9'', 0.8 mL Vol., with 2 injection sites and a male Luer Lock adapter; Baxter Healthcare Corporation,

…and 11 more.

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →