INTERLINK 2C7558Recalls & reported problems
Baxter Healthcare Corporation · Paclitaxel Set, Polyethylene Lined Tubing, Non-DEHP Pump Segment, 0.2 Micron Filter Injection Site, Male Luer Lock Adapter · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This is Baxter's Interlink paclitaxel administration set, a filtered tubing set used to deliver chemotherapy infusions. The 28 reports mostly describe fluid or saline leaking from the tubing, filter, or drip chamber, along with some trouble filling the set, air getting into the line, and filters coming apart. No deaths and no injuries were reported; in most cases the problem was noticed before or during use with no effect on the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00085412036380), received between Jan 15, 2018 and Sep 24, 2026 — classified as 30 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (24 reports).
5 other recalls cover devices cleared under the same FDA submission (K160007, K981792); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak24
- Filling Problem4
- Disconnection2
- Air/Gas in Device1
- Break1
- Failure to Prime1
Patient problems
- No Clinical Signs, Symptoms or Conditions26
- No Consequences Or Impact To Patient3
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50886558 ↗MalfunctionAug 19, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50326655 ↗MalfunctionMay 15, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A PACLITAXEL SET LEAKED NORMAL SALINE SOLUTION "NEAR THE DRIP CHAMBER". THE EVENT WAS IDENTIFIED DURING THE FINAL PHYSICAL LINE CHECK PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25189162 ↗MalfunctionAug 14, 2024Filling Problem; Filling Problem
IT WAS REPORTED THAT THERE WERE UNDERINFUSION OF SOLUTION DURING USE OF AN UNSPECIFIED QUANTITY OF PACLITAXEL SETS; FURTHER DESCRIBED AS "OVERFILL LEFT IN IV BAGS". THIS WAS OBSERVED DURING PATIENT INFUSION ON A PUMP THAT WAS SCHEDULED FOR OVER 4 HOURS. APPROXIMATELY 40ML OF OVERFILL VOLUME REMAINED IN THE INTRAVENOUS BAGS THAT WERE SPIKED TO NON-BAXTER DEVICES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19984785 ↗MalfunctionApr 2, 2024Filling Problem; Filling Problem
IT WAS REPORTED THAT AN UNDERINFUSION OF MEDICATION TO THE PATIENT OCCURRED WHILE USING A PACLITAXEL SET. THE EXPECTED THERAPY TIME WAS 3 HOURS; HOWEVER, THE BAG WAS NOT EMPTY AFTER 3 HOURS. THE INFUSION LASTED AN EXTRA 30 ¿ 40 MINUTES LONGER THAN EXPECTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19023427 ↗MalfunctionMar 29, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PACLITAXEL SET WAS LEAKING FROM THE FILTER. THE LEAK WAS DISCOVERED AT THE END OF PATIENT TREATMENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19006812 ↗MalfunctionMar 12, 2024Filling Problem; Air/Gas in Device; Filling Problem; Air/Gas in Device
IT WAS REPORTED THAT UNDER INFUSIONS OCCURRED AND THERE WAS AIR IN THE TUBING DURING THE USE OF EIGHT (8) PACLITAXEL SETS. DURING INFUSIONS, THERE WERE 8 INSTANCES WHERE ADDITIONAL THERAPY TIME WAS REQUIRED, RANGING FROM 30 TO 60 MINUTES, WITH DIFFERENT MEDICATIONS. ADDITIONALLY, THE TUBING WAS MORE SUSCEPTIBLE TO LARGE AIR BUBBLES DURING THE PRIMING PROCESS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18886539 ↗MalfunctionMar 12, 2024Break; Break
IT WAS REPORTED THAT THE FILTER OF A PACLITAXEL SET BROKE APART FROM THE CHAMBER. THIS EVENT OCCURRED DURING THE PRIMING PROCESS. IT WAS NOT SPECIFIED IF THERE WAS PATIENT INVOLVEMENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18886252 ↗MalfunctionMar 11, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF PACLITAXEL SETS LEAKED AFTER CONNECTING AND PRIMING TO INTRAVENOUS BAGS. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18876153 ↗MalfunctionJan 30, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PACLITAXEL SET TPN (TOTAL PARENTERAL NUTRITION) TUBING LEAKED FROM AN UNSPECIFIED LOCATION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18606601 ↗MalfunctionJun 13, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE PACLITAXEL SET LEAKED. THE ISSUE OCCURRED DURING AN UNKNOWN PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17118821 ↗MalfunctionJun 12, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PACLITAXEL SET LEAKED AT THE CONNECTION BETWEEN THE TUBING AND THE DRIP CHAMBER. IT WAS NOT SPECIFIED WHEN IN THE PROCESS STEP THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17107032 ↗MalfunctionMay 25, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF PACLITAXEL SETS LEAKED AT THE CONNECTION BETWEEN THE TUBING AND THE DRIP CHAMBER. THIS WAS IDENTIFIED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17007339 ↗MalfunctionMay 25, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF PACLITAXEL SETS LEAKED AT THE CONNECTION BETWEEN THE TUBING AND THE DRIP CHAMBER. THIS WAS IDENTIFIED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17007338 ↗MalfunctionMay 16, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT TWO (2) PACLITAXEL SETS WERE LEAKING WHERE THE TUBING CONNECTS TO THE DRIP CHAMBER. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16941243 ↗MalfunctionMay 11, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PACLITAXEL SET LEAKED IN AN UNKNOWN LOCATION. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16911538 ↗MalfunctionApr 25, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF PACLITAXEL SETS LEAKED FROM THE AREA JUST BELOW THE FILL DRIP CHAMBER. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807309 ↗MalfunctionApr 5, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF PACLITAXEL SETS LEAKED. THE LEAK WAS DISCOVERED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16683611 ↗MalfunctionMar 14, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE UNSPECIFIED QUANTITY OF PACLITAXEL SETS WERE LEAKING BELOW THE DRIP CHAMBER. THE LOCATION OF THE LEAK WAS FURTHER DESCRIBED AS IN THE AREA WHERE THE TUBING GOES INTO THE DRIP CHAMBER. THIS WAS OBSERVED DURING PRIMING PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16537546 ↗MalfunctionMar 6, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PACLITAXEL SET LEAKED FROM A "COUPLE DIFFERENT SITES IN THE TUBING". THE ISSUE WAS IDENTIFIED BEFORE PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16487062 ↗MalfunctionFeb 10, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF PACLITAXEL SETS LEAKED FROM THE DRIP CHAMBER. THE ISSUE WAS IDENTIFIED DURING SET UP IN THE ONCOLOGY UNIT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16357563 ↗MalfunctionFeb 8, 2023Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY [AT LEAST ONE(1)] OF PACLITAXEL SETS LEAKED FROM THE DRIP CHAMBER. THE ISSUE WAS IDENTIFIED DURING SETUP AND PREPARATION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16328805 ↗MalfunctionNov 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PACLITAXEL SET LEAKED FROM THE CHAMBER. THE ISSUE WAS IDENTIFIED DURING AN UNSPECIFIED PROCESS STEP PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15759528 ↗MalfunctionJun 8, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PACLITAXEL SET LEAKED FROM WHERE THE DRIP CHAMBER CONNECTS TO THE TUBING. THE DEVICE WAS USED WITH PACLITAXEL DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14633728 ↗MalfunctionJun 8, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED A PACLITAXEL SET WAS LEAKING FROM WHERE THE DRIP CHAMBER CONNECTS TO THE TUBING. THE LEAK WAS OBSERVED DURING PATIENT INFUSION OF CHEMOTHERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14633009 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160007, K981792. One submission often covers many models, so check the product and lot details in each record.
- Z-0619-2007 · Feb 16, 2007 · TerminatedBaxter Interlink System Vented Paclitaxel Set; a sterile fluid pathway with polyethylene lined tubing, non-DEHP pump segment, 0.22 micron downstream high pressure filter, injection
- Z-0618-2007 · Feb 16, 2007 · TerminatedBaxter Clearlink System Vented Paclitaxel Set; a sterile fluid pathway with polyethylene lined tubing, non-DEHP 12 inch pump segment, 0.22 micron downstream high pressure filter, L
- Z-0620-2007 · Feb 16, 2007 · TerminatedBaxter Vented Paclitaxel Set; a sterile fluid pathway with polyethylene lined tubing, non-DEHP pump segment, 0.22 micron downstream high pressure filter, injection site, male Luer
- Z-0193-05 · Apr 28, 2004 · TerminatedBaxter Clearlink Intravenous (IV) Delivery System; the system consists of solution sets, extension sets and blood sets that have a luer activated valve for IV access, which allows
- Z-1158-03 · Jul 16, 2003 · TerminatedClearlink System Vented Paclitaxel Set, 10 drops/mL, 111'', Polyethylene Lined Tubing, Non-DEHP Pump Segment, 0.22 Micron Extended Life Filter, Luer Activated Valve, Male Luer Lock
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page