CLEARLINK 2C8671Recalls & reported problems
Baxter International Inc. · Extension Set, 0.2 Micron Filter, Luer Activated Valve, Male Luer Lock Adapter with Retractable Collar · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Baxter ClearLink 2C8671 extension set is a filtered tubing set with a Luer-activated valve used for IV infusions. The 38 reports mostly describe fluid or blood leaking from the tubing — often at the filter connection or a cracked Y-site — along with lines coming apart, flow that stops or runs freely, and blocked filters that would not refill. Three reports involve injuries; no deaths are described. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 39
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 39 adverse event reports that list this exact device (DI 00085412046334), received between Apr 16, 2018 and Sep 24, 2026 — classified as 3 injuries, 36 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (15 reports).
2 other recalls cover devices cleared under the same FDA submission (K153158); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak15
- Disconnection8
- Crack4
- Improper Flow or Infusion4
- Free or Unrestricted Flow3
- Improper or Incorrect Procedure or Method3
- No Flow3
- Air/Gas in Device2
Patient problems
- No Consequences Or Impact To Patient15
- No Clinical Signs, Symptoms or Conditions14
- No Patient Involvement4
- High Blood Pressure/ Hypertension2
- No Known Impact Or Consequence To Patient2
- Hypoglycemia1
- Irritability1
- Low Blood Pressure/ Hypotension1
Recent adverse event reports
The 25 most recent of 39 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Fluid/Blood Leak; Detachment of Device or Device Component
It was reported that a exhibited a bond separation at the Clearlink Y-site, resulting in leakage from the Total Parenteral Nutrition and lipid line and interruption of the Total Parenteral Nutrition infusion. This occurred during patient infusion. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50883069 ↗MalfunctionApr 23, 2026Fluid/Blood Leak
IT WAS REPORTED THAT AN UNSPECIFIED BAXTER ACCESS SET LEAKED FROM AN UNSPECIFIED LOCATION DURING A TOTAL PARENTERAL NUTRITION (TPN) INFUSION. THE PATIENT BECAME ACUTELY AGITATED DESPITE FENTANYL TREATMENT, PROMPTING THE REGISTERED NURSE TO ASSESS THE INTRAVENOUS (IV) LINE, AT WHICH TIME LEAKAGE OF THE TPN WAS IDENTIFIED. THE TUBING WAS DEEMED UNRELIABLE AND DISCONTINUED, AND THE PROVIDER ORDERED DEXTROSE CONTAINING IV FLUIDS IN PLACE OF THE TPN. BLOOD GLUCOSE LEVELS WERE MONITORED, AND THE PATIENT WAS SUBSEQUENTLY FOUND TO BE HYPOGLYCEMIC AND REQUIRED A 10% DEXTROSE (D10) BOLUS. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Hypoglycemia; Irritability
MAUDE report 24973814 ↗MalfunctionApr 10, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT AN IV EXTENSION SET CRACKED AT THE CLEARLINK Y-SITE LUER, RESULTING IN A LEAK. THE EVENT OCCURRED DURING DOXURUBICIN INFUSION IN THE ACUTE CARE INPATIENT UNIT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24844964 ↗MalfunctionMar 16, 2026Fluid/Blood Leak
IT WAS REPORTED THAT TWO(2) CLEARLINK SYSTEM EXTENSION SETS LEAKED FROM THE END OF THE TUBING WHERE THE MICRON FILTER WAS ATTACHED. THE ISSUE WAS NOTICED DURING PATIENT INFUSION. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24609103 ↗MalfunctionNov 13, 2024Complete Blockage; No Flow; Complete Blockage; No Flow
IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET HAD NO SOLUTION FLOW. THE FILTER CHAMBER SEEMED BLOCKED. ONCE THE FILTER CHAMBER WAS FILLED AND EMPTIED, THE FILTER CHAMBER WOULD NOT REFILL. THIS OCCURRED DURING SETUP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20674508 ↗MalfunctionOct 8, 2024Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM EXTENSION SETS HAD WHITE DISCOLORATION OR A BLEMISH INSIDE THE TUBING, NEAR THE FILTER. SOME OF THE WHITE DISCOLORATION (SUBSTANCE) COULD BE SCRAPED OFF THE TUBING. THIS WAS OBSERVED WHEN RETRIEVING THE SETS FROM THE CASES, PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20403759 ↗MalfunctionOct 27, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM EXTENSION SETS HAD FLOW ISSUES (NOT FURTHER SPECIFIED). THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18021565 ↗InjuryMar 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A DURING A PATIENT INFUSION OF A VASOPRESSOR WITH TWO CLEARLINK SYSTEM EXTENSION SETS, THE PATIENT'S BLOOD PRESSURE INCREASED. IT WAS FURTHER REPORTED THAT WHEN THE LINE WAS PRIMED THE FILTER FILLS BUT EMPTIES OUT ONCE THEY START RUNNING THE INFUSION, WHICH AFFECTED THE VASOPRESSOR TITRATIONS. WHEN THE FILTER WAS NOTED TO BE EMPTY, A NEW FILTER WAS PRIMED AND REPLACED HOWEVER, THE SAME ISSUE OCCURRED WITH THE NEW FILTER. THE BLOOD PRESSURE NORMALIZED AFTER THE FILTER WAS REMOVED FROM THE LINE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: High Blood Pressure/ Hypertension
MAUDE report 16471809 ↗MalfunctionFeb 16, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK SYSTEM EXTENSION SET APPEARED CRACKED AND SUBSEQUENTLY LEAKED. AFTER THE SET WAS PRIMED, THE NURSE "WENT TO PUT THE TUBING INTO THE PUMP" AND OBSERVED "THE FILTER FILLED WITH AIR WHEN CLAMPED". THE NURSED NOTED THE LINE WAS LEAKING. THE SET WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16386641 ↗InjuryMay 9, 2022Free or Unrestricted Flow; Free or Unrestricted Flow
IT WAS REPORTED THAT DURING PATIENT INFUSION OF LEVOPHED, PHENYLEPHRINE AND AMIODARONE WITH 3 UNITS OF CLEARLINK SYSTEM EXTENSION SETS, AN INADVERTENT BOLUS FROM THE EXTENSION FILTER WAS OBSERVED. THIS RESULTED IN THE PATIENT EXPERIENCING HYPERTENSION REPORTED AS ¿CAUSED THE BLOOD PRESSURE TO SKYROCKET¿. ACCORDING TO THE REPORTER, THE FILTER EMPTIED OUT ON ITS OWN. NO CONNECTION ISSUES WERE NOTED AND THE LINES WERE SETUP PROPERLY. AFTER THE "FILTERS WERE REMOVED, THE PATIENT'S BLOOD PRESSURE "EVENED OUT". NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: High Blood Pressure/ Hypertension
MAUDE report 14339574 ↗MalfunctionMar 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET LEAKED FROM THE INJECTION PORT. THE ISSUE WAS IDENTIFIED DURING A PATIENT INFUSION IN THE CANCER CENTER; FURTHER DESCRIBED AS ¿NOTED DURING THE NS (NORMAL SALINE) FLUSH PRIOR TO CHEMOTHERAPY ADMINISTRATION¿. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13715615 ↗MalfunctionMar 9, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET WAS LEAKING COMING FROM THE FILTER AREA. THIS WAS OBSERVED AN INFUSION OF IRINOTECAN TO A PATIENT IN THE CANCER CENTER. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13715413 ↗MalfunctionJan 4, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET LEAKED ONTO THE FLOOR FROM THE FILTER. THE ISSUE WAS IDENTIFIED DURING PATIENT INFUSION WITH KEYTRUDA. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13158754 ↗MalfunctionJun 2, 2021Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED A CLEARLINK SYSTEM EXTENSION SET APPEARED TO HAVE AIR IN THE LINE. IT WAS FURTHER REPORTED, "DESPITE PROPER PRIMING OF THE FILTER, THE FILTER WAS COMPLETELY FILLED WITH AIR SHORTLY AFTER INFUSION STARTED". THE SET WAS FILLED WITH VELETRI. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11920266 ↗MalfunctionJun 1, 2021Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED A CLEARLINK SYSTEM EXTENSION SET APPEARED TO HAVE AIR IN THE LINE. IT WAS FURTHER REPORTED, "DESPITE PROPER PRIMING OF THE FILTER, THE FILTER WAS COMPLETELY FILLED WITH AIR SHORTLY AFTER INFUSION STARTED". THE SET WAS FILLED WITH REMODULIN. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11914378 ↗MalfunctionMay 21, 2021Filling Problem; Filling Problem
IT WAS REPORTED THAT TWO (2) CLEARLINK SYSTEM EXTENSION SETS OVER-INFUSED DURING PATIENT USE. IT WAS FURTHER REPORTED THAT THE FILTER PORTION WAS EMPTYING OUT "WHILE INLINE AND CAUSING THE PATIENT TO GET BLOUSED WITH THE MEDICATION INFUSING". THE SET WAS USED WITH NOREPINEPHRINE 4MG/250ML. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11865430 ↗MalfunctionMay 19, 2021No Flow; No Flow
IT WAS REPORTED THAT "MEDICATION WAS NOT PASSING OR FLOWING THROUGH" A CLEARLINK SYSTEM EXTENSION SET. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11849620 ↗MalfunctionJan 6, 2021Backflow; Crack; Backflow; Crack
IT WAS REPORTED THAT THE FILTER OF A CLEARLINK SYSTEM EXTENSION SET WAS CRACKED. IT WAS FURTHER REPORTED THERE WAS DIFFICULTY WITH THE PERIPHERALLY INSERTED CENTRAL CATHETER (PICC) INFUSING; THE FLUID WAS BACKING UP. THIS WAS IDENTIFIED DURING USE OF THE DEVICE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11127106 ↗MalfunctionSep 9, 2020Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THE IV TUBING OF A CLEARLINK SYSTEM EXTENSION SET POPPED OFF AT THE SCREW-ON COLLAR WHILE CONNECTED TO A NEONATE. THE EVENT WAS FURTHER REPORTED AS "THE COLLAR DOES NOT ENGAGE, JUST KEEPS SPINNING AROUND AND NOT TIGHTENING" WHICH LED TO A "SMALL AMOUNT OF BLOOD LOSE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10507365 ↗MalfunctionSep 5, 2019Improper or Incorrect Procedure or Method; Improper Flow or Infusion; Improper or Incorrect Procedure or Method; Improper Flow or Infusion
IT WAS REPORTED A CLEARLINK SYSTEM EXTENSION SET COMPLETELY EMPTIED. HOWEVER, UPON FURTHER CLARIFICATION IT WAS DETERMINED THE USER WAS NOT "INVERTING AND TAPPING" THE FILTER WHEN PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 8967277 ↗MalfunctionSep 5, 2019Improper or Incorrect Procedure or Method; Improper Flow or Infusion; Improper or Incorrect Procedure or Method; Improper Flow or Infusion
IT WAS REPORTED A CLEARLINK SYSTEM EXTENSION SET COMPLETELY EMPTIED. HOWEVER, UPON FURTHER CLARIFICATION IT WAS DETERMINED THE USER WAS NOT "INVERTING AND TAPPING" THE FILTER WHEN PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 8967274 ↗MalfunctionSep 5, 2019Improper or Incorrect Procedure or Method; Improper Flow or Infusion; Improper or Incorrect Procedure or Method; Improper Flow or Infusion
IT WAS REPORTED A CLEARLINK SYSTEM EXTENSION SET COMPLETELY EMPTIED. HOWEVER, UPON FURTHER CLARIFICATION IT WAS DETERMINED THE USER WAS NOT "INVERTING AND TAPPING" THE FILTER WHEN PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 8967235 ↗MalfunctionSep 5, 2019Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED A CLEARLINK SYSTEM EXTENSION SET COMPLETELY EMPTIED. HOWEVER, UPON FURTHER CLARIFICATION IT WAS DETERMINED THE USER WAS NOT "INVERTING AND TAPPING" THE FILTER WHEN PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 8967174 ↗MalfunctionAug 29, 2019Free or Unrestricted Flow; Free or Unrestricted Flow
IT WAS REPORTED THAT THE FLOW WAS UNABLE TO BE CONTROLLED ON AN UNKNOWN QUANTITY OF CLEARLINK SOLUTION ADMINISTRATION SETS WHICH RESULTED IN FREE FLOW OF MEDICATION. THIS ISSUE WAS IDENTIFIED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THESE EVENTS. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8947682 ↗MalfunctionAug 29, 2019Free or Unrestricted Flow; Free or Unrestricted Flow
IT WAS REPORTED THAT THE FLOW WAS UNABLE TO BE CONTROLLED ON AN UNKNOWN QUANTITY OF CLEARLINK SOLUTION ADMINISTRATION SETS WHICH RESULTED IN FREE FLOW OF MEDICATION. THIS ISSUE WAS IDENTIFIED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THESE EVENTS. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8947653 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K153158. One submission often covers many models, so check the product and lot details in each record.
- Z-0196-2026 · Aug 29, 2025 · Open, ClassifiedCLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, 0.2 Micron Filter, three CLEARLINK Luer Activated Valves, Backcheck Valve, 10 drops/mL, 102-inch (2.6 meters), Product Code 2
- Z-0195-2026 · Aug 29, 2025 · Open, ClassifiedCLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, 0.2 Micron Filter, two CLEARLINK Luer Activated Valves, Backcheck Valve, 10 drops/mL, 104-inch (2.6 meters), Product Code 2C8
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page