Skip to content

CLEARLINK 2H8603Recalls & reported problems

Baxter International Inc. · Non-DEHP Extension Set, 1.2 Micron Filter, Luer Activated Valve, Male Luer Lock Adapter · Class 2

All models of CLEARLINK (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Baxter ClearLink non-DEHP extension set is a filtered tubing set used to administer IV fluids and medications such as total parenteral nutrition. The 35 reports mostly describe the set leaking, failing to flow or prime, coming apart at the filter, disconnecting from patients, or letting air into the tubing. No deaths or injuries are reported. One recall, started in August 2025 and still open, cites leaking IV sets.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
35
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 00085412046433), received between Dec 28, 2017 and Feb 25, 2026 — classified as 35 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (11 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 29, 2025Open, Classified

    IV sets may leak.

    Z-0200-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak11
  • Failure to Prime6
  • No Flow6
  • Break4
  • Air/Gas in Device3
  • Disconnection3
  • Partial Blockage3
  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions23
  • No Patient Involvement11
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2026Crack; Fluid/Blood Leak; Obstruction of Flow

    IT WAS REPORTED THAT THE FILTER OF A CLEARLINK SYSTEM NON-DEHP EXTENSION SET EXPLODED AND RESULTED IN LEAKAGE. THE ISSUE WAS DESCRIBED AS "THE FILTER WAS JAMMED AND WOULD NOT ALLOW MEDICATION THROUGH THE FILTER. A SALINE FLUSH WAS THEN USED TO TRY TO FLUSH THROUGH THE FILTER AND THE FILTER EXPLODED." THIS OCCURRED DURING PRIMING FOR AN INFLIXIMAB INFUSION; HOWEVER, IT WAS UNSPECIFIED WHETHER THE PATIENT WAS CONNECTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24444830 ↗
  • MalfunctionNov 12, 2025Break; Break

    IT WAS REPORTED THAT THE "CLEAR PART OF THE FILTER" OF A CLEARLINK SYSTEM NON-DEHP EXTENSION SET BROKE OFF. THE ISSUE OCCURRED WHILE PRIMING THE SET WITH TOTAL PARENTERAL NUTRITION (TPN). THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23532815 ↗
  • MalfunctionAug 12, 2025Break; Break

    IT WAS REPORTED THAT THE FILTER OF A CLEARLINK NON-DEHP EXTENSION SET FELL APART. WHILE ATTEMPTING TO PRIME, PRIOR TO ADMINISTRATION, IT WAS OBSERVED THAT THE CLEAR OUTER HOUSING FELL OFF THE FILTER. TO RESOLVE THIS ISSUE THE FILTER WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22786093 ↗
  • MalfunctionJul 29, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET LEAKED AT THE FILTER. THIS WAS OBSERVED DURING PATIENT INFUSION. THE DEVICE CONTAINED TPN (TOTAL PARENTERAL NUTRITION). THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22648626 ↗
  • MalfunctionJun 18, 2025No Flow; No Flow

    IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET DID NOT FLOW TOTAL PARENTERAL NUTRITION THROUGH THE FILTER. THE DEVICE WAS REMOVED AND FLUSHED WITH NORMAL SALINE, BUT THE NURSE WAS UNABLE TO FLUSH. THIS OCCURRED DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22286487 ↗
  • MalfunctionMay 29, 2025Disconnection; Fluid/Blood Leak; Air/Gas in Device; Disconnection; Fluid/Blood Leak; Air/Gas in Device

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF NON-DEHP EXTENSION SETS DISCONNECTED FROM AT LEAST ONE (1) PATIENT. THE EVENT WAS FURTHER DESCRIBED AS "THE FILTER HAS EVACUATED SPONTANEOUSLY". THIS OCCURRED WHILE THE SETS WERE BEING USED WITH A NON-BAXTER PRIMARY PUMP SET AND A NON-BAXTER INFUSION PUMP. THIS EVENT RESULTED IN FLUID/MEDICATIONS SPILLING ONTO THE FLOOR AND AIR IN THE INTRAVENOUS (IV) TUBING. THIS OCCURRED DURING PATIENT INFUSION WITH PROPOFOL AND/OR HYDROMORPHONE. THE MEDICATIONS WERE DISCONNECTED FROM THE PATIENT AFTER THE INFUSION WAS PAUSED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22109793 ↗
  • MalfunctionSep 9, 2024Air/Gas in Device; Air/Gas in Device

    IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET BACK FILLED THE LINE WITH AIR. THE EVENT OCCURRED WHEN THE DEVICE WAS CONNECTED TO THE PATIENT AND THE PUMP WAS STARTED. THE AIR FILTER WAS SUCCESSFULLY PRIMED WITH TPN (TOTAL PARENTERAL NUTRITION). AFTER THE AIR IN THE TUBING LINE WAS OBSERVED THE INFUSION WAS STOPPED AND THE TUBING WAS CHANGED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20171047 ↗
  • MalfunctionJul 23, 2024Filling Problem; Air/Gas in Device; Filling Problem; Air/Gas in Device

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM EXTENSION SETS UNDERINFUSED DURING PATIENT INFUSION. AIR WAS GATHERING IN THE FILTER DURING PRIMING. ONCE ALL THE AIR WAS ELIMINATED, THE PATIENTS WERE NOT GETTING THE FULL DOSE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19812399 ↗
  • MalfunctionMay 10, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak

    IT WAS REPORTED THE FILTERS IN THREE (3) CLEARLINK SYSTEM NON-DEHP EXTENSION SETS APPEARED TO BE CRACKED WHICH SUBSEQUENTLY LEAKED. THE EVENTS OCCURRED DURING UNKNOWN PROCESS STEPS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19291082 ↗
  • MalfunctionNov 28, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET WAS LEAKING TOTAL PARENTERAL NUTRITION FROM THE FILTER. THE LEAK WAS OBSERVED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18220101 ↗
  • MalfunctionMar 22, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP EXTENSION SET WAS LEAKING TOTAL PARENTERAL NUTRITION (TPN) FROM THE SIDE PORT. THIS WAS OBSERVED DURING PATIENT INFUSION USING AN UNSPECIFIED PUMP. WHEN THE LEAK WAS FOUND, A GREEN CAP (NON-BAXTER) WAS PLACED ON THE SIDE PORT; WHEN THE CAP WAS REMOVED, TPN SPRAYED OUT OF THE SIDE PORT. TO RESOLVE THE EVENT, THE TPN HAD TO BE RE-SPIKED WITH NEW TUBING. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16590332 ↗
  • MalfunctionMar 17, 2023Backflow; Backflow

    IT WAS REPORTED THAT THERE WAS BACKFLOW OF BLOOD INTO THE FILTER FROM THE NON-BAXTER CATHETER WHEN USING A CLEARLINK SYSTEM NON-DEHP EXTENSION SET; NOTHING WAS CLAMPED AND THE UNSPECIFIED PUMP WAS PROGRAMMED AND RUNNING. THIS WAS IDENTIFIED DURING PATIENT INFUSION OF LIPIDS. THE SET WAS REPLACED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16565334 ↗
  • MalfunctionMar 16, 2023No Flow; No Flow

    IT WAS REPORTED THAT THERE WAS NO FLUID FLOW THROUGH THE FILTER OF TWO (2) CLEARLINK SYSTEM NON-DEHP EXTENSION SETS. WHILE PRIMING THE SETS, THERE WAS NO FLUID FLOW THROUGH THE FILTER. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16554498 ↗
  • MalfunctionDec 22, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE FILTER OF A CLEARLINK SYSTEM NON-DEHP EXTENSION SET HAD FILTER "BUSTED". THIS EVENT OCCURRED DURING AN UNSPECIFIED PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16035008 ↗
  • MalfunctionDec 19, 2022Break; Break

    IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP EXTENSION SET HAD FILTER "BUSTED DURING INFUSION". THIS EVENT OCCURRED DURING AN UNSPECIFIED PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16009552 ↗
  • MalfunctionJun 20, 2022Disconnection; Disconnection

    IT WAS REPORTED THAT THE TUBING SEPARATED FROM THE FILTER OF A CLEARLINK SYSTEM NON-DEHP EXTENSION SET. THIS ISSUE WAS IDENTIFIED DURING THE START OF AN INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14745060 ↗
  • MalfunctionJun 20, 2022Disconnection; Disconnection

    IT WAS REPORTED THAT THE TUBING OF A CLEARLINK SYSTEM NON-DEHP EXTENSION SET SEPARATED FROM THE FILTER. THIS WAS OBSERVED DURING A PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14743966 ↗
  • MalfunctionMay 20, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF A CLEARLINK SYSTEM NON-DEHP EXTENSION SETS WERE LEAKING FROM THE SIDE OF THE PORT. THE ISSUE WAS IDENTIFIED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14462663 ↗
  • MalfunctionDec 14, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP EXTENSION SET LEAKED. THE EVENT WAS FURTHER DESCRIBED AS "TPN (TOTAL PARENTERAL NUTRITION) FLUID WAS SEEPING OUT OF ONE OF THE INDENTS (HOLES) IN THE BACK OF THE TPN FILTER CARTRIDGE". IT WAS STATED THE PATIENT AND THEIR BED LINENS WERE DAMP. THIS ISSUE WAS IDENTIFIED DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13002533 ↗
  • MalfunctionOct 12, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP EXTENSION SET LEAKED. THE EVENT WAS FURTHER DESCRIBED AS A LEAK AROUND THE ¿FILTER AREA¿ ON THE SET. THIS ISSUE WAS IDENTIFIED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12616186 ↗
  • MalfunctionOct 6, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE FILTER OF A CLEARLINK SYSTEM NON-DEHP EXTENSION SET LEAKED DURING PATIENT INFUSION WITH REMICADE. IT WAS FURTHER REPORTED APPROXIMATELY 20-30 ML LEAKED ONTO THE PATIENT'S LINENS AND CHAIR. THE SET WAS REPLACED AND THE INFUSION WAS COMPLETED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12583117 ↗
  • MalfunctionSep 20, 2021Partial Blockage; Partial Blockage

    IT WAS REPORTED THAT THERE WAS NO FLOW OF MEDICATION THROUGH A CLEARLINK SYSTEM NON-DEHP EXTENSION SET. IT WAS FURTHER REPORTED THAT THE FILTER BECAME CLOGGED DURING LIPID ADMINISTRATION WHICH CAUSES A DELAY OF THE ADMINISTRATION. TO RESOLVE THIS ISSUE, THE SET WAS REPLACED, AND THERAPY WAS CONTINUED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12502325 ↗
  • MalfunctionSep 2, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP EXTENSION SET LEAKED CLOSE TO ONE OF THE SEALED AREAS. THE ISSUE WAS IDENTIFIED DURING PRIMING WHILE USING A 20% UNSPECIFIED INTRALIPID. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12416157 ↗
  • MalfunctionSep 18, 2020No Flow; No Flow

    IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP EXTENSION SET WOULD NOT FLOW. THE EVENT WAS FURTHER DESCRIBED AS THE MEDICATION WOULD NOT FLOW PAST THE FILTER CHAMBER. THIS ISSUES WAS IDENTIFIED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Patient Involvement

    MAUDE report 10552813 ↗
  • MalfunctionFeb 6, 2020Failure to Prime; Failure to Prime

    IT WAS REPORTED THAT TWO (2) CLEARLINK SYSTEM NON-DEHP EXTENSION SETS WOULD NOT PRIME. THE EVENT WAS FURTHER DESCRIBED AS ¿THE MEDICATION FILLS THE FILTER PARTIALLY AND BECOMES BLOCKED; THE MEDICATION WILL NOT COMPLETELY FILL FILTER OR FLOW PAST FILTER¿. THIS ISSUE WAS IDENTIFIED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Patient Involvement

    MAUDE report 9679296 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page