CLEARLINK 2C8634Recalls & reported problems

Baxter Healthcare Corporation · Extension Set, Luer Activated Valve, Male Luer Lock Adapter with Retractable Collar · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00085412049021), received between Apr 23, 2020 and Apr 15, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).

1 FDA recall record name this device's identifier, 1 still open.

7 other recalls cover devices cleared under the same FDA submission (K112893); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 29, 2025Open, Classified

    IV sets may leak.

    Z-0199-2026 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak7
  • Disconnection3
  • Detachment of Device or Device Component2
  • Incomplete or Inadequate Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 15, 2026Fluid/Blood Leak; Detachment of Device or Device Component

    IT WAS REPORTED THAT A STRAIGHT-TYPE EXTENSION SET "CAME APART" DURING INFUSION OF UNSPECIFIED FLUIDS. THE PATIENT WAS USING THE RESTROOM PRIOR TO SURGERY. WHEN STAFF RETURNED TO ESCORT THE PATIENT BACK TO THE ROOM, THE TUBING WAS FOUND TO HAVE COME APART, WITH "MINIMAL" BLOOD OBSERVED ON THE BATHROOM FLOOR. STAFF PROMPTLY CLAMPED THE TUBING NEAR THE PATIENT, FLUSHED THE IV LINE, AND REPLACED THE TUBING AND IV BAG. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24888470 ↗
  • MalfunctionSep 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET SPLIT (SEPARATED) IN TWO PIECES. THIS WAS OBSERVED DURING AN APHERESIS PROCEDURE. THE PATIENT WAS ACCESSED WITH A 16G NON-BAXTER HIGH-FLOW NEEDLE. AT THE END OF THE PROCEDURE, THE NURSE PLACED A NON-BAXTER CONNECTOR AND FLUSHED THE LINE WITHOUT ISSUES. WHEN THE NURSE RETURNED TO THE ROOM APPROXIMATELY 10 MINUTES LATER, THE EXTENSION SET SPLIT IN TWO PIECES, LEAVING THE PORT OPEN TO AIR. THE NURSE IMMEDIATELY INSERTED THE ONE END BACK INTO THE OTHER AND ASPIRATED ANY AIR THAT MAY HAVE BEEN IN THE LINE. THE PORT NEEDLE WAS THEN REMOVED PER PROTOCOL. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20260969 ↗
  • MalfunctionMay 2, 2024Fluid/Blood Leak; Incomplete or Inadequate Connection; Fluid/Blood Leak; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK LUER ACTIVATED VALVE EXTENSION SETS LEAKED FROM THE LUER LOCK HUB THAT CONNECTS TO THE IV CATHETER SYSTEM. THE LUER LOCK IS NOT COMPLETELY SEATED AND WILL NOT MOVE PAST THE FIRST THREAD. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19233559 ↗
  • MalfunctionFeb 6, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK EXTENSION SETS LEAKED; FURTHER DESCRIBED AS "LEAKING AT THE LUER LOCK HUB THAT CONNECTS TO THE IV CATHETER SYSTEM." THIS WAS OBSERVED DURING USE OF THE PRODUCT; HOWEVER, PATIENT INVOLVEMENT IS NOT SPECIFIED. THE LEAK WAS NOT CORRECTED BY RESEATING THE CONNECTION. A NEW EXTENSION SET WOULD BE PLACED TO RESOLVE THE EVENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18655944 ↗
  • MalfunctionNov 28, 2022Disconnection; Disconnection

    IT WAS REPORTED THAT THE LUER LOCK COLLAR OF TWO (2) CLEARLINK SYSTEM EXTENSION SETS WERE SNAPPED OFF CAUSING A PIECE OF THE SET TO BECAME LODGED INSIDE THE LUMEN OF THE CENTRAL LINE. THIS ISSUE WAS IDENTIFIED DURING INFUSION. THE CENTRAL LINE HAD TO BE REPLACED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15870976 ↗
  • MalfunctionOct 12, 2021Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak

    IT WAS REPORTED THAT THE TUBING OF A CLEARLINK SYSTEM EXTENSION SET SEPARATED FROM THE CONNECTOR WHICH RESULTED IN A LEAK. THE ISSUE WAS IDENTIFIED DURING USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12614900 ↗
  • MalfunctionApr 26, 2021Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak

    IT WAS REPORTED THAT THE LUER OF CLEARLINK SYSTEM EXTENSION SET SEPARATED FROM THE TUBING WHICH RESULTED IN A LEAK. THIS WAS OBSERVED DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11728574 ↗
  • MalfunctionApr 23, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET WAS LEAKING FROM THE CONNECTOR WHEN ATTACHED TO THE INTRAVENOUS (IV) HUB. IT WAS FURTHER REPORTED THAT THE LEAK CONTINUED AS THE NURSE TRIED TO TIGHTEN THE CONNECTION AT THE HUB. THIS WAS NOTED DURING PATIENT USE. THE SET WAS REMOVED TO CONTINUE THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9990646 ↗
  • MalfunctionApr 23, 2020Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A CLEARLINK SYSTEM EXTENSION SET WAS LEAKING FROM THE CONNECTOR WHEN ATTACHED TO THE INTRAVENOUS (IV) HUB. IT WAS FURTHER REPORTED THAT THE LEAK CONTINUED AS THE NURSE TRIED TO TIGHTEN THE CONNECTION AT THE HUB. THIS WAS NOTED DURING PATIENT USE. THE SET WAS REMOVED TO CONTINUE THERAPY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9990644 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112893. One submission often covers many models, so check the product and lot details in each record.

  • Z-0198-2026 · Aug 29, 2025 · Open, Classified
    CLEARLINK SYSTEM EXTENSION SET, Straight-Type Extension Set, Standard Bore, CLEARLINK Luer Activated Valve, Retractable Collar, 14-inch (35 centimeters), Product Code 2C8632
  • Z-0197-2026 · Aug 29, 2025 · Open, Classified
    CLEARLINK SYSTEM EXTENSION SET, Straight-Type Extension Set, Standard Bore, CLEARLINK Luer Activated Valve, Retractable Collar, 44-inch (111 centimeters), Product code 2C8610
  • Z-0191-2026 · Aug 29, 2025 · Open, Classified
    CLEARLINK SYSTEM Solution Set, Non-Vented, two CLEARLINK Luer Activated Valves, 10 drops/mL, 100-inch (2.5 meters), Product Code 2C8425
  • Z-0190-2026 · Aug 29, 2025 · Open, Classified
    CLEARLINK System Solution Set, DUO-VENT, CLEARLINK Luer Activated Valve, 10 drops/mL, 92-inch (2.3 meters), Product code 2C8419
  • Z-1653-2022 · Aug 9, 2022 · Terminated
    Baxter CLEARLINK System, Non-DEHP Solution Set with DUO-VENT Spike, Product Code 2R8403. An in-line injection site.
  • Z-0099-2014 · Sep 4, 2013 · Terminated
    Baxter CLEARLINK System Non-DEHP Three Lead Extension Set. 6.0" (15 cm), 1.0 mL, Luer Activated Valve, male Luer Lock Adapter, Sterile, nonpyrogenic, Baxter Healthcare Corporation,
  • Z-1271-2013 · Mar 28, 2013 · Terminated
    A) Product Code 2N8371: Non-DEHP Y-Type Catheter Extension Set Kit, 6.0", Vol. 1.0 mL, 2 Luer Activated Valves, Male Luer Lock Adapter, Single use only Product Usage: The IV exte

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.