CLEARLINK/CONTINU-FLO/DUO-VENT 2C8541Recalls & reported problems
Baxter Healthcare Corporation · CONTINU-FLO Solution Set with DUO-VENT Spike, 3 Luer Activated Valves, Male Luer Lock Adapter with Retractable Collar · Class 2
All models of CLEARLINK/CONTINU-FLO/DUO-VENT (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Baxter Continu-Flo IV administration set has 804 adverse event reports, 15 describing injuries and no deaths. Many describe fluid leaks, parts detaching or coming apart, and disconnections, and most mention no effect on the patient. There are 2 open recalls: one for sets that may leak, and one for connectors that need more force than expected to fully connect. Reports fell from 282 in the previous 12 months to 107 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 797
- Deaths reported
- 0
- Injuries reported
- 15
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 797 adverse event reports that list this exact device (DI 00085412071107), received between Sep 22, 2017 and Sep 28, 2026 — classified as 15 injuries, 782 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (445 reports).
2 FDA recall records name this device's identifier, 2 still open.
10 other recalls cover devices cleared under the same FDA submission (K203609, K961225); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 29, 2025Class IIOpen, Classified
IV sets may leak.
Z-0194-2026 · Root cause: Process control · Recall details · FDA record ↗
- Dec 12, 2019Class IIOpen, Classified
Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.
Z-0793-2020 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak445
- Detachment of Device or Device Component121
- Disconnection97
- Break49
- Material Puncture/Hole47
- Material Split, Cut or Torn37
- Backflow32
- No Flow21
Patient problems
- No Clinical Signs, Symptoms or Conditions638
- No Consequences Or Impact To Patient86
- No Patient Involvement44
- Low Blood Pressure/ Hypotension7
- No Known Impact Or Consequence To Patient5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- Hemorrhage/Blood Loss/Bleeding4
- Air Embolism3
Recent adverse event reports
The 25 most recent of 797 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026No Flow
It was reported that an Extension Set did not allow fluid to flow by gravity after connection to the patient. Resistance was encountered when attempting to flush fluid through the tubing and catheter. The intravenous catheter was restarted, resulting in more than one intravenous stick. When the tubing was replaced, flow returned to normal. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50918915 ↗MalfunctionSep 28, 2026No Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50918644 ↗MalfunctionSep 22, 2026Detachment of Device or Device Component
It was reported that a CLEARLINK System CONTINU-FLO Solution Set leaked at the connection between the spike and drip chamber while chemotherapy remained in the tubing. This occurred during an unspecified process step while the set was connected to the patient. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50850032 ↗MalfunctionSep 15, 2026Air/Gas in Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50780037 ↗MalfunctionSep 14, 2026Fitting Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50756617 ↗MalfunctionSep 12, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50744863 ↗MalfunctionSep 9, 2026Break; Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50694295 ↗MalfunctionAug 21, 2026Fluid/Blood Leak; Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50380332 ↗MalfunctionAug 17, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50287214 ↗MalfunctionAug 6, 2026Fluid/Blood Leak; Detachment of Device or Device Component
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50165713 ↗InjuryJul 31, 2026Detachment of Device or Device Component
It was reported that a CONTINU-FLO Solution Set with DUO-VENT Spike separated at an unspecified location during a Levophed infusion. The issue was identified when the patient¿s blood pressure monitoring indicated a decrease in blood pressure. As a result, a new infusion setup was initiated. No additional information is available.
Patient problems: Low Blood Pressure/ Hypotension
MAUDE report 50094078 ↗MalfunctionJul 2, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A CONTINU-FLO SOLUTION SET HAD FLUIDS LEAKING AROUND THE SECTION OF TUBING THAT CONNECTS THE SPIKE/VENT TO THE CHAMBER. THIS OCCURRED DURING INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25693475 ↗MalfunctionJul 1, 2026Air/Gas in Device
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY CLEARLINK SYSTEM CONTINU-FLO SOLUTION SETS HAD AIR BUBBLES IN THE TUBING. THIS OCCURRED AFTER THE LINES WERE PRIMED AND THE AIR BUBBLES WERE REMOVED. THE FLUID FLOWED WITHOUT BUBBLES INITIALLY; HOWEVER, ONCE THE LINE WAS PLACED INTO THE PUMP, BUBBLES APPEARED FROM THE VALVES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25676824 ↗MalfunctionJun 30, 2026
FOREIGN OBJECT OBSERVED IN IV TUBING (ACTIVELY INFUSING). IV SITE CLAMPED; IV TUBING REMOVED AND IMMEDIATELY REPORTED PER ORGANIZATIONAL PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25664601 ↗MalfunctionJun 17, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A CONTINU-FLO SOLUTION SET LEAKED AT THE JUNCTION WHERE THE DRIP CHAMBER CONNECTS TO THE IV TUBING. THE ISSUE WAS IDENTIFIED IN THE RHEUMATOLOGY DEPARTMENT WHEN THE REGISTERED NURSE SPIKED THE BAG MEDICATION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25533680 ↗MalfunctionMay 29, 2026Fluid/Blood Leak; Material Split, Cut or Torn
IT WAS REPORTED THAT A CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET LEAKED FROM A "TEAR" IN THE TUBING. THIS OCCURRED AFTER PRIMING WITH PROPOFOL IN THE ANESTHESIA DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333767 ↗MalfunctionMay 27, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A CONTINU-FLO SOLUTION SET WITH DUO-VENT SPIKE LEAKED FROM THE Y-SITE PORT. THIS OCCURRED DURING INFUSION OF INSULIN VIA A NOVUM IQ LVP. THE CLINICIAN IDENTIFIED THE LEAKAGE WHEN THE PATIENT¿S BLOOD GLUCOSE LEVELS INCREASED AND THE BED SHEET WAS OBSERVED TO BE WET. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 25308650 ↗MalfunctionMay 26, 2026Material Puncture/Hole
IT WAS REPORTED THAT HOLES WERE IDENTIFIED IN TEN (10) CONTINU-FLO SOLUTION SETS WITH DUO-VENT SPIKE. THE EVENT WAS DISCOVERED PRIOR TO PATIENT USE AT THE OR DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25285034 ↗MalfunctionMay 19, 2026Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET SPONTANEOUSLY DETACHED AT THE MIDPOINT WHILE ACTIVELY INFUSING SALINE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 25219653 ↗MalfunctionMay 18, 2026
LABELING ON THE BACK OF THE BAG IS WORN OFF
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25199741 ↗MalfunctionMay 8, 2026Backflow
IT WAS REPORTED THAT A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET WAS NOT INFUSING PROPERLY AND THE PRIMARY BAG (HUNG LOWER) GOT FULLER THAN THE SECONDARY BAG (HUNG HIGHER). THE ISSUE OCCURRED DURING KCL (POTASSIUM CHLORIDE) INFUSION. THE EXPECTED INFUSION TIME WAS 2 HOURS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25125904 ↗MalfunctionMay 8, 2026Backflow
IT WAS REPORTED THAT A CLEARLINK SYSTEM CONTINU-FLO SOLUTION SET WAS NOT INFUSING PROPERLY AND THE PRIMARY BAG (HUNG LOWER) GOT FULLER THAN THE SECONDARY BAG (HUNG HIGHER). THE ISSUE OCCURRED DURING KCL (POTASSIUM CHLORIDE) INFUSION. THE EXPECTED INFUSION TIME WAS 2 HOURS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25125887 ↗InjuryMay 7, 2026Break
IT WAS REPORTED THAT THE DISTAL MALE LUER CONNECTOR (CLEARLINK) OF A CLEARLINK DUO VENT CONTINU FLO SOLUTION FRACTURED AND BECAME LODGED IN A Y SITE PORT OF ANOTHER PRIMARY TUBING SET. THE EVENT OCCURRED WHILE MULTIPLE VASOPRESSORS WERE BEING ADMINISTERED TO A CRITICALLY ILL PATIENT. AN EMERGENT REPLACEMENT OF BOTH INFUSION LINES WAS PERFORMED, AS THE PATIENT EXPERIENCED A DROP IN BLOOD PRESSURE DUE TO THE TUBING FAILURE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Low Blood Pressure/ Hypotension
MAUDE report 25117261 ↗MalfunctionApr 29, 2026Delivered as Unsterile Product
IT WAS REPORTED THAT A CONTINU-FLO SOLUTION SET WITH DUO-VENT SPIKE WAS MISSING THE BLUE CAP ON THE PATIENT END OF THE TUBING. THE ISSUE WAS IDENTIFIED WHEN THE NURSE WAS PREPARING TO PRIME THE SET, AND THE CAP COULD NOT BE LOCATED IN THE PACKAGING. A NEW TUBING SET WAS OBTAINED, PRIMED, AND USED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25029075 ↗MalfunctionApr 22, 2026Material Separation
IT WAS REPORTED THAT A CLEARLINK CONTINU-FLO SOLUTION SET WITH DUO-VENT SPIKE "SNAPPED IN HALF WHILE PRIMING IV FLUIDS". THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24956867 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K203609, K961225. One submission often covers many models, so check the product and lot details in each record.
- Z-0192-2026 · Aug 29, 2025 · Open, ClassifiedCLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, two CLEARLINK Luer Activated Valves, Backcheck Valve, 10 drops/mL, 112-inch (2.8 meters), Product Code 2C8519
- Z-0193-2026 · Aug 29, 2025 · Open, ClassifiedCLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, three CLEARLINK Luer Activated Valves, Backcheck Valve, Retractable Collar, 10 drops/mL, 120-inch (3.0 meters), Product Code
- Z-2594-2025 · Aug 8, 2025 · Open, ClassifiedBaster Continu-Flo Solution Set, Non-Vented, 2 Clearlink Luer Activated Valves, Backcheck Valve. 10 drops/mL. 112" (2.8 m), Product Code REF 2C8519; intravascular administration s
- Z-1011-2025 · Dec 20, 2024 · Open, ClassifiedBaxter Continu-Flo Solution Set with Duo-Vent Spike, REF UC8519
- Z-0794-2020 · Dec 12, 2019 · Open, ClassifiedONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8377, 7N8378, 7N8390, 7N8391, 7N8399 - Pro
- Z-0475-2012 · Dec 6, 2011 · TerminatedBaxter Interlink System Extension Set, 13" (32 cm), Check Valve, Injection Site, Male Luer Lock Adapter; an Rx sterile, nonpyrogenic fluid pathway for the administration of sterile
- Z-2920-2011 · Jul 21, 2011 · TerminatedBaxter Solution Set with Duo-Vent Spike; 109" (2.8 m), Male Luer Lock Adapter 10 drops/mL. Baxter Healthcare Corporation. A sterile, single use, nonpyrogenic fluid pathway for use
- Z-0653-06 · Dec 15, 2005 · TerminatedBaxter Interlink System Continu-Flo Solution Set with Standard Bore 4-Way Stopcock Manifold Extension Set and Minivolume Extension Set, product code 3C0078; An Rx sterile, nonpyrog
- Z-0845-05 · Feb 28, 2005 · TerminatedBaxter ClearLink Intravenous (IV) Solution Sets and Extension Sets that have a ClearLink luer activated valve for IV access, which allows the administration of medications/solutio
- Z-1152-03 · Jul 16, 2003 · TerminatedInterlink System Continu-Flo Solution Set with Microbore Tubing, 60 drops/mL, 128'', 3 injection sites, Male Luer Lock Adapter, catalog 2C6521; Baxter Healthcare Corporation, Deerf
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page