CLEARLINK/BURETROL 2C8862Recalls & reported problems
Baxter International Inc. · BURETROL Solution Set, 150 mL Valveless Burette, 4 Luer Activated Valves, Male Luer Lock Adapter with Retractable Collar · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00085412073354), received between Dec 20, 2018 and Oct 24, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Crack” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K123874); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Crack2
- Disconnection2
- Filling Problem2
- Fluid/Blood Leak2
- Air/Gas in Device1
- Free or Unrestricted Flow1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient2
- No Patient Involvement2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2024Crack; Crack
IT WAS REPORTED THAT A CLEARLINK SYSTEM BURETROL SOLUTION SET CHAMBER CRACKED. DURING AN UNSPECIFIED INFUSION ON A PEDIATRIC PATIENT, THE UNKNOWN INFUSION PUMP WAS BEEPING, AND A STAFF MEMBER HEARD A ¿LOUD POP¿ ORIGINATING ¿FROM THE CHAMBER OF THE TUBING¿. THE BURETROL CHAMBER WAS OBSERVED ¿CRACKED AND SHATTERED¿. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20530034 ↗MalfunctionOct 14, 2024Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM BURETROL SOLUTION SETS HAD AIR IN THE LINE; FURTHER DESCRIBED AS, "BUBBLES BACK FLOWING". THIS OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20446735 ↗MalfunctionMar 31, 2021Disconnection; Disconnection
IT WAS REPORTED THAT THE DRIP CHAMBER CAME APART FROM THE BURETTE CHAMBER OF A CLEARLINK SYSTEM BURETROL SOLUTION SET. THIS WAS IDENTIFIED WHEN PRIMING WITH 0.9% NORMAL SALINE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11595804 ↗MalfunctionMar 16, 2021Disconnection; Disconnection
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK SYSTEM BURETROL SOLUTION SET DISCONNECTED BELOW THE DRIP CHAMBER. THIS WAS IDENTIFIED OUT OF THE PACKAGE, BEFORE SPIKING INTRAVENOUS (IV) FLUIDS. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11496855 ↗MalfunctionAug 9, 2019Fluid/Blood Leak; Free or Unrestricted Flow; Fluid/Blood Leak; Free or Unrestricted Flow
IT WAS REPORTED THAT A CLEARLINK BURETROL SOLUTION SET ¿FILLED DESPITE THE VALVE BEING CLAMPED. IT THEN STARTED LEAKING OUT THE TOP OF THE BURETROL CHAMBER WHILE IT WAS HANGING ON AN IV (INTRAVENOUS) POLE. THE IV WAS NEVER HOOKED TO A PATIENT.¿ THIS WAS IDENTIFIED WHILE PRIMING THE BURETROL TUBING WITH NORMAL SALINE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8880148 ↗MalfunctionJul 1, 2019Filling Problem; Filling Problem
IT WAS REPORTED THAT A CLEARLINK BURETROL SOLUTION SET OVER-INFUSED. IT WAS FURTHER REPORTED THAT THE BURETROL TUBING WAS FILLED WITH SALINE TO THE TOP DESPITE THE VENT BEING CLAMPED. THIS WAS IDENTIFIED DURING AN UNSPECIFIED STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8749095 ↗MalfunctionJul 1, 2019Filling Problem; Fluid/Blood Leak; Filling Problem; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK BURETROL SOLUTION SET FILLED DESPITE THE VALVE BEING CLAMPED. IT WAS FURTHER REPORTED THAT LEAKING WAS OBSERVED OUT OF THE TOP OF THE BURETROL CHAMBER WHILE HANGING THE SET ON AN INTRAVENOUS (IV) POLE. THIS WAS IDENTIFIED DURING PRIMING WITH NORMAL SALINE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 8749047 ↗MalfunctionDec 20, 2018Crack; Material Deformation; Crack; Material Deformation
IT WAS REPORTED THAT THE BURETROL COMPONENT OF A CLEARLINK BURETROL SOLUTION SET COLLAPSED DURING INFUSION. THE REPORTER STATED THAT THE BURETROL WAS ¿AUTOMATICALLY SUCKING IN¿ AND COLLAPSING. MULTIPLE ATTEMPTS WERE MADE BY THE CUSTOMER TO RECTIFY THE ISSUE, SUCH AS ENSURING THE CLAMPS WERE RELEASED; HOWEVER, THE BURETROL CONTINUED COLLAPSING. AFTER 16 HOURS OF INFUSION, THE BURETROL COMPONENT CRACKED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 8184862 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123874. One submission often covers many models, so check the product and lot details in each record.
- Z-2614-2014 · Jul 14, 2014 · Terminated***Baxter***BURETROL Solution Set***105" (2.7 m) 150 mL Burette***Drip Chamber Filter Valve***3 Luer Activated Valves***Male Luer Lock Adapter***with Retractable Collar***Fluid pat
- Z-2613-2014 · Jul 14, 2014 · Terminated***Baxter***Buretrol Solution Set***105" (2.7 m)***150 mL Burette***Drip Chamber Filter Valve***3 Injection Sites Male Luer Lock Adapter with***Retractable Collar***Fluid path is S
- Z-2615-2014 · Jul 14, 2014 · Terminated***Baxter***Non-DEHP***BURETROL Solution Set 105" (2.7 m), 150 mL Burette***Drip Chamber Filter Valve 3 Luer Activated Valves***Male Luer Lock Adapter***with Retractable Collar***F
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page