CLEARLINK 2H8486Recalls & reported problems
Baxter Healthcare Corporation · Non-DEHP Solution Set, 1.2 Micron Downstream Filter, Luer Activated Valve, Male Luer Lock Adapter with Retractable Collar · Class 2
All models of CLEARLINK (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Baxter Clearlink solution set is a non-DEHP intravenous administration set with a downstream filter used to deliver IV medication and nutrition. The 28 reports mostly describe tubing that will not prime, leaks of fluid or medication from ports, filters that block flow, air bubbles in the line, and tubing disconnecting from the cap. None of the reports describe a death or an injury. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 29
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00085412074764), received between Feb 4, 2019 and Sep 11, 2026 — classified as 29 malfunctions.
The most frequently reported specific device problem is “Failure to Prime” (9 reports).
1 other recall cover devices cleared under the same FDA submission (K952074); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Prime9
- Fluid/Blood Leak9
- Air/Gas in Device6
- No Flow6
- Backflow1
- Crack1
- Detachment of Device or Device Component1
- Filling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Patient Involvement9
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50736287 ↗MalfunctionApr 24, 2026Backflow; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK NON-DEHP SOLUTION SET LEAKED MEDICATION FROM THE LUER LOCK IV PORT ABOVE THE FILTER. THIS OCCURRED DURING INFUSION OF INFLIXIMAB. THE INFUSION WAS STARTED AT A RATE OF 40ML/HR. AFTER 30 MINUTES, THE INFUSION RATE WAS SET TO 80ML/HR. 25 MINUTES LATER, THE HEALTHCARE PROFESSIONAL CHECKED ON THE PATIENT AND IDENTIFIED THAT THE MEDICATION HAD LEAKED ON THE FLOOR. IV AND IV LINES ALL APPEAR TO BE INTACT. THE MEDICATION BAG WAS EMPTY AND MEDICATION LINE APPEARS TO HAVE BACKFLOW OF BLOOD. THE HEALTHCARE PROFESSIONAL INSPECTED THE LINES FURTHER AND NOTED ONE OF THE LUER LOCK IV PORTS ABOVE THE FILTER TO BE LEAKING FLUIDS. THE LINE WAS FLUSHED WITH SALINE. PATIENT RECEIVED APPROXIMATELY 1/4-1/3 OF MEDICATION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24988216 ↗MalfunctionJan 15, 2026Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT A CLEARLINK SYSTEM SOLUTION SET LEAKED FROM A 1 INCH HOLE IN THE TUBING. THIS WAS OBSERVED WHEN THE NURSE TRIED TO PRIME THE TUBING WITH TPN (TOTAL PARENTERAL NUTRITION). THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24087084 ↗MalfunctionJan 13, 2025Filling Problem; Filling Problem
IT WAS REPORTED THAT AN OVERINFUSION OF MEDICATION OCCURRED WHILE USING AN UNSPECIFIED QUANTITY OF CLEARLINK, NON-DEHP SOLUTION SETS. THE INFUSIONS ARE SET FOR 24 HOURS AND APPROXIMATELY 2L, BUT THE PUMPS ARE RUNNING OUT OF FLUID ABOUT 30 MINUTES EARLY AND DELIVERING 50-60 ML MORE THAN PRESCRIBED. THESE OVERINFUSIONS OCCURRED WITH SIGMA SPECTURM PUMPS DURING TPN (TOTAL PARENTERAL NUTRITION) TREATMENTS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21138693 ↗MalfunctionOct 2, 2024No Flow; No Flow
IT WAS REPORTED THAT SOLUTION WOULD NOT FLOW THROUGH THE FILTER OF A CLEARLINK SYSTEM NON-DEHP SOLUTION SET. THIS ISSUE WAS FURTHER DESCRIBED AS, ¿TPN TUBING WOULD NOT FLUSH THROUGH FILTER ¿ TPN FLUSHED BACK INTO BAG ¿ NORMAL SALINE USED TO CLEAR TUBING AND ALSO WOULD NOT FLUSH PAST FILTER¿. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20354805 ↗MalfunctionJul 8, 2024Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK SYSTEM NON-DEHP SOLUTION SET SEPARATED FROM THE LUER LOCK WHICH RESULTED IN A FLUID LEAK. THIS EVENT WAS FURTHER DESCRIBED AS, ¿THE TUBING USED TO INFUSE TPN HAD DISCONNECTED FROM THE CAP AND THE CAP REMAINED ATTACHED TO THE PICC LUMEN¿. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19693309 ↗MalfunctionMay 7, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP SOLUTION SET WAS LEAKING AT THE FILTER SITE. THE LEAK WAS OBSERVED WHILE RUNNING LIPIDS DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19260128 ↗MalfunctionDec 27, 2022No Flow; Air/Gas in Device; No Flow; Air/Gas in Device
IT WAS REPORTED THAT WHEN THE AIR AT THE FILTER OF A CLEARLINK SYSTEM NON-DEHP SOLUTION SET IS EMPTIED, IT NO LONGER FLOWS AT ALL. THE ISSUE WAS IDENTIFIED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16051981 ↗MalfunctionDec 2, 2022No Flow; Air/Gas in Device; No Flow; Air/Gas in Device
IT WAS REPORTED THAT AIR BUBBLES WERE OBSERVED IN A CLEARLINK SYSTEM NON-DEHP SOLUTION SET. IT WAS FURTHER REPORTED THAT THE CLAMP WAS NOT ALLOWING SOLUTION TO FLOW. THE ISSUE WAS IDENTIFIED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15905564 ↗MalfunctionDec 2, 2022No Flow; Air/Gas in Device; No Flow; Air/Gas in Device
IT WAS REPORTED THAT AIR BUBBLES WERE LEAKING INTO A CLEARLINK SYSTEM NON-DEHP SOLUTION SET. ADDITIONALLY, IT WAS REPORTED THAT CLAMP WAS NOT ALLOWING SOLUTION TO FLOW. THIS ISSUE WAS IDENTIFIED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15904469 ↗MalfunctionDec 1, 2022No Flow; Air/Gas in Device; No Flow; Air/Gas in Device
IT WAS REPORTED THAT AN AIR BUBBLES WERE OBSERVED IN A CLEARLINK SYSTEM NON-DEHP SOLUTION SET. IT WAS FURTHER REPORTED THAT THE CLAMP WAS NOT ALLOWING SOLUTION TO FLOW. THE ISSUE WAS IDENTIFIED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15895767 ↗MalfunctionSep 13, 2021Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP SOLUTION SET FILLED WITH AIR. IT WAS STATED THAT THE SET WAS PRIMED FOR USE WITH ENTERAL NUTRITION. WHEN THE NURSE INVERTED THE FILTER TO ATTACH THE LINE TO THE PATIENT'S CENTRAL LINE, THE LINE BEGAN TO FILL WITH AIR FROM THE MALE LUER ADAPTER BACK TOWARD THE FILTER. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12461480 ↗MalfunctionSep 10, 2021Air/Gas in Device; Air/Gas in Device
IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP SOLUTION SET HAD AIR IN LINE. IT WAS STATED THAT THE SET WAS PRIMED FOR USE WITH ENTERAL NUTRITION. WHEN THE NURSE INVERTED THE FILTER TO ATTACH THE LINE TO THE PATIENT'S CENTRAL LINE, THE LINE BEGAN TO FILL WITH AIR FROM THE MALE LUER ADAPTER BACK TOWARD THE FILTER. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12452670 ↗MalfunctionAug 19, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE FILTER OF A CLEARLINK SYSTEM NON-DEHP SOLUTION SET WAS LEAKING. THIS WAS OBSERVED DURING PRIMING WITH LIPIDS. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12344470 ↗MalfunctionAug 6, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP SOLUTION SET LEAKED. IT WAS FURTHER REPORTED THAT UPON STARTING A LIPID ADMINISTRATION, LIPIDS WERE COMING OUT OF THE PORT LOCATED RIGHT ABOVE THE FILTER. THE ISSUE WAS IDENTIFIED DURING USE. AS A RESULT, A NEW LIPID SOLUTION WAS MADE BY THE PHARMACY, AND A NEW SETUP WAS USED TO DELIVER THE PATIENT INFUSION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12291530 ↗MalfunctionJun 28, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM NON-DEHP SOLUTION SET LEAKED. THE LEAK WAS FURTHER DESCRIBED AS "THE BAG WAS SPIKED AND THE TUBING WAS PRIMED, LIPID SOLUTION WAS NOTED TO BE LEAKING FROM THE PORT LOCATED ABOVE THE FILTER." THIS OCCURRED DURING PRIMING OF THE DEVICE, PRIOR TO PATIENT USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12078766 ↗MalfunctionDec 2, 2020Crack; Crack
IT WAS REPORTED A CLEARLINK SYSTEM NON-DEHP SOLUTION SET BROKE. IT WAS FURTHER REPORTED THE LINE BROKE OFF WHEN THE USER UNCLAMPED THE LINE AFTER THE FILTER ON THE END OF THE LINE. THE PLASTIC APPEARED CRACKED. THIS ISSUE WAS IDENTIFIED DURING PRIMING; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE..
Patient problems: No Patient Involvement
MAUDE report 10931812 ↗MalfunctionOct 8, 2020Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT A LARGE AIR BUBBLE WAS OBSERVED COMING FROM THE DISTAL END FILTER OF AN UNSPECIFIED QUANTITY OF CLEARLINK SYSTEM NON-DEHP SOLUTION SETS. THE SET WAS USED WITH LIPID THAT WAS CONNECTED TO AN UNSPECIFIED BAXTER PUMP. THIS ISSUE WAS IDENTIFIED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10651267 ↗MalfunctionMay 1, 2020Failure to Prime; Failure to Prime
IT WAS REPORTED THAT A CLEARLINK SYSTEM SOLUTION SET WOULD NOT PRIME PAST THE FILTER. THE CUSTOMER WAS PRIMING LIPIDS THROUGH THE SET PRIOR TO PATIENT USE; HOWEVER, THE TUBING WOULD NOT PRIME ONCE THE LIPIDS REACHED THE FILTER. MULTIPLE UNSUCCESSFUL ATTEMPTS TO PUSH LIPIDS THROUGH THE SET WERE ATTEMPTED. TO RESOLVE THE ISSUE, THE SET WAS REPLACED, AND TREATMENT WAS ABLE TO CONTINUE WITH THE NEW SET. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 10018888 ↗MalfunctionApr 6, 2020Failure to Prime; Failure to Prime
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK NON-DEHP SOLUTION SETS WOULD NOT FLOW DURING PRIMING. THE REPORTER STATED THAT THE LIPIDS STOPPED FLOWING ONCE THEY REACHED THE FILTER. THE LIPIDS BEING USED WERE 50 GRAMS IN 250 MLS. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 9930922 ↗MalfunctionApr 6, 2020Failure to Prime; Failure to Prime
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF CLEARLINK NON-DEHP SOLUTION SETS WOULD NOT FLOW DURING PRIMING. THE REPORTER STATED THAT THE LIPIDS STOPPED FLOWING ONCE THEY REACHED THE FILTER. THE LIPIDS BEING USED WERE 50 GRAMS IN 250 MLS. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 9930921 ↗MalfunctionMar 25, 2020Failure to Prime; Failure to Prime
IT WAS REPORTED THAT TWO CLEARLINK NON-DEHP SOLUTION SETS WERE UNABLE TO BE PRIMED. THE EVENT OCCURRED DURING PRIMING WITH UNSPECIFIED LIPIDS, DESCRIBED AS "50 GRAMS IN 250 MLS". THE REPORTER STATED THAT THE CHECK VALVE WAS INVERTED AND TAPPED, THE BAG WAS NOT INITIALLY SQUEEZED BUT AFTER THE LIPID STOPPED FLOWING AT THE FILTER THE BAG WAS THEN SQUEEZED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 9878779 ↗MalfunctionMar 20, 2020No Flow; No Flow
IT WAS REPORTED A CLEARLINK SYSTEM NON-DEHP SOLUTION SET WOULD NOT FLOW. THE DEVICE WAS FILLED WITH 50G OF LIPIDS IN 250ML. IT WAS FURTHER REPORTED THE ¿LIPIDS WOULD NOT FLOW THROUGH THE FILTER.¿ ¿THE CHECK VALVE WAS INVERTED AND TAPPED, THE BAG WAS NOT INITIALLY SQUEEZED BUT AFTER THE LIPID WOULD STOP FLOWING AT THE FILTER THE BAG WOULD THEN BE SQUEEZED.¿ THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9862453 ↗MalfunctionMar 9, 2020Failure to Prime; Failure to Prime
IT WAS REPORTED THAT SEVEN (7) CLEARLINK SYSTEM NON-DEHP SOLUTION SETS FAILED TO PRIME. DURING PRIMING, IT WAS DISCOVERED THAT THE CUSTOMER COULD NOT PUSH LIPIDS THROUGH THE SETS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9807476 ↗MalfunctionMar 5, 2020Failure to Prime; Failure to Prime
IT WAS REPORTED A CLEARLINK SYSTEM NON-DEHP SOLUTION SET WOULD NOT PRIME. IT WAS FURTHER REPORTED THE USER "COULD NOT GET THE LIPID FLUID TO PRIME PAST THE FILTER". THIS ISSUE WAS IDENTIFIED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 9793468 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K952074. One submission often covers many models, so check the product and lot details in each record.
- Z-0200-2026 · Aug 29, 2025 · Open, ClassifiedCLEARLINK System Non_DEHP Extenstion Set, Straight-Type Extension Set, Standard Bore, 1.2 Micron Filter, CLEARLINK Luer Activated Valve, 19-inch (47 centimeters), Product Code 2H86
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page