ONE-LINK 7N8378Recalls & reported problems
Baxter International Inc. · Non-DEHP Standard Bore Catheter Extension Set, Needle-free IV Connector with Neutral Fluid Displacement, Male Luer Lock Adapter with Retractable Collar · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Baxter One-Link catheter extension set is a needle-free IV connector used to extend and flush intravenous lines. The 36 reports mostly describe the tubing splitting, cutting or rupturing so fluid leaks out, and lines that stop flowing or come apart at the connection. There are no deaths and no injuries. No recalls name this device. Reports rose from 3 in the earlier 12 months to 14 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 00085412091792), received between Jan 16, 2018 and Sep 25, 2026 — classified as 35 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (14 reports).
2 other recalls cover devices cleared under the same FDA submission (K132734); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak14
- Material Rupture8
- Material Split, Cut or Torn8
- Disconnection5
- Detachment of Device or Device Component4
- Crack3
- No Flow2
- Unsealed Device Packaging2
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- No Consequences Or Impact To Patient5
- No Patient Involvement4
- Hemorrhage/Blood Loss/Bleeding1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Material Rupture
It was reported that five (5) Non-DEHP Standard Bore Catheter Extension Sets fractured during infusion of intravenous contrast for computed tomography (CT) scans. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50900318 ↗MalfunctionSep 10, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50722379 ↗MalfunctionMay 15, 2026Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT AT LEAST TWO ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SETS SPLIT OPEN AND RESULTED IN LEAK DURING IV CONTRAST INJECTION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 25192607 ↗MalfunctionMay 15, 2026Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TUBING OF A ONE LINK NON DEHP STANDARD BORE CATHETER EXTENSION SET SPLIT OPEN BEFORE THE BONDED CONNECTOR AND LEAKED. THIS OCCURRED WHILE FLUSHING A NEWLY PLACED INTRAVENOUS (IV). THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25186766 ↗MalfunctionMay 14, 2026Material Split, Cut or Torn
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF NON-DEHP STANDARD BORE CATHETER EXTENSION SETS "SPLIT OPEN", RESULTING IN LEAKS. THE EVENTS OCCURRED DURING INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25176284 ↗MalfunctionJan 2, 2026Break; Fluid/Blood Leak
IT WAS REPORTED THAT AN ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET WAS SEVERED AND BROKE AT THE CONNECTION SITE. THE LINE WAS PRE PRIMED WITH RINGERS LACTATE SOLUTION BEFORE CONNECTION TO PATIENT. THE SET WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23950820 ↗MalfunctionDec 31, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT THE TUBING OF TWO (02) ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SETS RUPTURED. THIS OCCURRED DURING COMPUTED TOMOGRAPHY WITH INTRAVENOUS (IV) CONTRAST ADMINISTRATION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23934204 ↗MalfunctionDec 18, 2025Material Rupture; Material Split, Cut or Torn; Material Rupture; Material Split, Cut or Torn
IT WAS REPORTED THAT THE TUBING OF AN UNSPECIFIED QUANTITY OF ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SETS SPLIT WHILE USING A POWER INJECTOR. THE FIRST EXTENSION SET BLEW OUT DURING ADMINISTRATION OF INTRAVENOUS CONTRAST IN CT. A NEW EXTENSION SET WAS SET UP; HOWEVER, WHEN CONTRAST WAS ADMINISTERED, THE EXTENSION SET RUPTURED. THE INTRAVENOUS CONTRAST LEAKED AND THE QUANTITY OF LEAKING SETS WAS UNKNOWN. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23843064 ↗MalfunctionDec 3, 2025Material Rupture
DURING CT WITH IV CONTRAST ADMINISTRATION (NOT A HIGH-PRESSURE ADMINISTRATION) 2 DIFFERENT IV EXTENSION SET TUBES RUPTURED. BOTH EXTENSIONS WERE FROM THE SAME LOT NUMBER. THE STAFF ARE REMOVING THIS LOT NUMBER FROM THE STOCK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23704580 ↗MalfunctionDec 1, 2025Fluid/Blood Leak; Incomplete or Inadequate Connection; Fluid/Blood Leak; Incomplete or Inadequate Connection
IT WAS REPORTED THAT A ONE-LINK STANDARD BORE CATHETER EXTENSION SET DID NOT PROPERLY ENGAGE WITH THE PERIPHERAL VENOUS CATHETER AND LEAKED FROM AN UNSPECIFIED LOCATION DURING AN ATTEMPT TO ESTABLISH PATENCY. THIS WAS OBSERVED DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23687337 ↗MalfunctionNov 18, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT A ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET, WHICH WAS CONNECTED TO A INTRAVENOUS CATHETER, WAS CRACKED RESULTING IN A FLUID LEAK. DURING ADMINISTRATION, IT WAS OBSERVED THAT CONTRAST/SALINE WAS LEAKING FROM THE CRACK IN THE SET. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23588026 ↗MalfunctionNov 18, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT A ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET, WHICH WAS CONNECTED TO A INTRAVENOUS CATHETER, WAS CRACKED RESULTING IN A FLUID LEAK. DURING ADMINISTRATION, IT WAS OBSERVED THAT CONTRAST/SALINE WAS LEAKING FROM THE CRACK IN THE SET. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23587489 ↗MalfunctionNov 18, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT A ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET, WHICH WAS CONNECTED TO A INTRAVENOUS CATHETER, WAS CRACKED RESULTING IN A FLUID LEAK. DURING ADMINISTRATION, IT WAS OBSERVED THAT CONTRAST/SALINE WAS LEAKING FROM THE CRACK IN THE SET. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23587484 ↗MalfunctionAug 8, 2025Unsealed Device Packaging; Material Split, Cut or Torn; Unsealed Device Packaging; Material Split, Cut or Torn
IT WAS REPORTED THAT THE TUBING OF A ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET HAD A UNIFORM CUT. THE ISSUE WAS FURTHER DESCRIBED AS; "THE TUBING WAS INCOMPLETE WITH A UNIFORM CUT". THE TUBING WAS OUTSIDE OF THE PRIMARY PACKAGING, DESPITE BEING HERMETICALLY SEALED. THIS WAS DISCOVERED DURING DISPENSING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22747930 ↗MalfunctionJul 15, 2025Backflow; Backflow
IT WAS REPORTED THAT A ONE-LINK STANDARD BORE CATHETER EXTENSION SET HAD BLOOD BACKFLOW. THIS WAS OBSERVED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22513865 ↗MalfunctionJan 7, 2025Material Rupture; No Flow; Material Rupture; No Flow
IT WAS REPORTED THAT THE TUBING OF TWO (2) ONE-LINK STANDARD BORE CATHETER EXTENSION SETS RUPTURED APPROXIMATELY AN INCH FROM THE LOCK FROM FLOW ISSUES. THIS WAS OBSERVED DURING A CT (COMPUTED TOMOGRAPHY) PE (PULMONARY EMBOLISM) STUDY. THE TUBING WAS USED WITH A CT, AND THERE WAS NO FLOW OF FLUID. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21097676 ↗MalfunctionFeb 14, 2024Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT AN ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET LEAKED. THE DEVICE DISCONNECTED FROM A CATHETER RESULTING IN SPILLAGE OF SALINE SOLUTION. THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18707677 ↗MalfunctionNov 20, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT THE TUBING OF A ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET RUPTURED DURING ADMINISTRATION OF INTRAVENOUS (IV) CONTRAST. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18179249 ↗MalfunctionNov 17, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT THE TUBING OF AN ONE-LINK STANDARD BORE CATHETER EXTENSION SET RUPTURED DURING THE ADMINISTRATION OF INTRAVENOUS (IV) CONTRAST. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18161055 ↗MalfunctionNov 17, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT THE TUBING OF AN ONE-LINK STANDARD BORE CATHETER EXTENSION SET RUPTURED DURING THE ADMINISTRATION OF INTRAVENOUS (IV) CONTRAST. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18160872 ↗MalfunctionDec 29, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET LEAKED. THIS WAS OBSERVED DURING PATIENT USE. THE LEAK WAS NOTED ON THE SIDE OF THE TUBING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13113981 ↗MalfunctionJun 24, 2021Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn
IT WAS REPORTED THE "J LOOP" SPLIT ON A ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET AND SUBSEQUENTLY LEAKED. THIS ISSUE OCCURRED DURING CONTRAST INJECTION OF 1.5ML/SEC USING AN UNSPECIFIED POWER INJECTOR WITH 300 PSI. IT WAS FURTHER REPORTED THAT THE INJECTION WAS NOT PRESSURE LIMITED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12058303 ↗MalfunctionJun 24, 2021Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn
IT WAS REPORTED THE "J LOOP" SPLIT ON A ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET AND SUBSEQUENTLY LEAKED. THIS ISSUE OCCURRED DURING CONTRAST INJECTION OF 1.5ML/SEC USING AN UNSPECIFIED POWER INJECTOR WITH 300 PSI. IT WAS FURTHER REPORTED THAT THE INJECTION WAS NOT PRESSURE LIMITED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12058294 ↗MalfunctionMay 28, 2021Fluid/Blood Leak; Material Split, Cut or Torn; Fluid/Blood Leak; Material Split, Cut or Torn
IT WAS REPORTED THAT THE TUBING OF A ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SET SPLIT OPEN AND LEAKED. IT WAS FURTHER REPORTED THAT THE DISTAL END OF TUBING BULGED AND SPLIT OPEN WHEN POWER INJECTOR WAS FLUSHING WITH SALINE. THE DEVICE HAD BEEN IN USE FOR FIVE (5) MINUTES WHEN THE ISSUE OCCURRED. THERE WAS NO DIRECT CONTACT OR EXPOSURE AND NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11902156 ↗MalfunctionApr 5, 2021Packaging Problem; Material Split, Cut or Torn; Packaging Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT TWO (2) ONE-LINK NON-DEHP STANDARD BORE CATHETER EXTENSION SETS HAD STERILE PACKAGING ISSUES. DURING SETUP/PREPARATION, IT WAS DISCOVERED THE TWO SETS WERE RECEIVED CONJOINED. THIS ISSUE WAS FURTHER DESCRIBED AS HALF OF ONE SET WAS PROTRUDING OUTSIDE OF ITS STERILE PACKAGING AND WAS WITHIN THE PACKAGING OF THE SECOND SET. THE TUBING OF THE SET THAT WAS BETWEEN THE TWO PACKAGES WAS ALSO CUT. THIS EVENT WAS IDENTIFIED DURING SETUP/PREPARATION PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11617016 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K132734. One submission often covers many models, so check the product and lot details in each record.
- Z-1102-2021 · Dec 21, 2020 · TerminatedBaxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K - Product Usage: intended
- Z-0132-2016 · May 21, 2015 · TerminatedONE-LINK NON-DEHP Y-TYPE MICROBORE CATHETER EXTENSION SET NEEDLE-FREE IV CONNECTOR WITH NEUTRAL FLUID DISPLACEMENT
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026