ONE-LINK 7N8310Recalls & reported problems
Baxter International Inc. · Non-DEHP Microbore Catheter Extension Set, Needle-free IV Connector with Neutral Fluid Displacement, Power Injectable, Male Luer Lock Adapter · Class 2
All models of ONE-LINK (12)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00085412476728), received between Jan 19, 2018 and Feb 13, 2026 — classified as 18 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (7 reports).
2 other recalls cover devices cleared under the same FDA submission (K132734); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak7
- Detachment of Device or Device Component5
- Disconnection5
- Backflow1
- Detachment Of Device Component1
- Device Damaged Prior to Use1
- Device-Device Incompatibility1
- Infusion or Flow Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient7
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A ONE-LINK NON-DEHP MICROBORE CATHETER EXTENSION SET LEAKED FROM AN UNSPECIFIED LOCATION, "CAUSING THE LIPID INFUSION TO NOT INFUSE AND THE IV TO BACK UP WITH BLOOD IN THE CANNULA". THE ISSUE OCCURRED DURING PATIENT INFUSION. THE TUBING WAS CHANGED TO RESOLVE THE ISSUE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24339406 ↗MalfunctionNov 20, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TUBING OF A NON-DEHP MICROBORE CATHETER EXTENSION SET CAME APART. THE ISSUE WAS FURTHER DESCRIBED THAT "THE EXTENSION PIECE HAD SPLIT AT THE EXTENSION WHERE IT CONNECTS TO PLASTIC." THE END CAPS WERE CHANGED AND "HEPARIN LOCKED." THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23610525 ↗MalfunctionNov 18, 2025Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT TUBING OF A ONE-LINK NON-DEHP MICROBORE CATHETER EXTENSION SET FELL APART FROM THE PATIENT'S PORT. THE ISSUE WAS FURTHER DESCRIBED AS, "THE PATIENT'S LINE DISCONNECTED FROM THE ONE-LINK CONNECTOR (CAP) ON THE PORT NEEDLE OF THE IV EXTENSION TUBING". THIS OCCURRED DURING PATIENT INFUSION WITH HEPARIN. THE SET WAS REPLACED WITH A NEW LINE AND FLUID ADMINISTRATION WAS RESTARTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23589190 ↗MalfunctionSep 5, 2025Backflow; Fluid/Blood Leak; Detachment of Device or Device Component; Backflow; Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT AN UNSPECIFIED AMOUNT OF ONE-LINK MICROBORE CATHETER EXTENSION SETS LEAKED FROM THE SCREW BY THE NEEDLE-FREE IV (INTRAVENOUS) CONNECTOR. THE LEAK IS FROM THE CONNECTION THAT IS UNBOUNDED. FLUID ALSO BACKS UP INTO THE EXTENSION TUBING AT TIMES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22983534 ↗MalfunctionJun 1, 2022Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED A ONE-LINK NON-DEHP MICROBORE CATHETER EXTENSION SET LEAKED. FURTHERMORE, A HOLE WAS OBSERVED "THE MIDDLE OF THE TUBING" DURING AN UNSPECIFIED ANTIBIOTIC INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14570876 ↗MalfunctionSep 2, 2021Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT THE UNSPECIFIED QUANTITY OF ONE-LINK NON-DEHP MICROBORE CATHETER EXTENSION SETS HAD DISCONNECTION ISSUES WITH CONTRAST LEAKING. THE DISCONNECTION WAS COMING FROM THE PART WHERE IT SCREWED INTO THE INTRAVENOUS CATHETER. THE CONTRAST WOULD LEAK OUT MID INJECTION, AS A RESULT THE PATIENT¿S DID NOT RECEIVE ENOUGH CONTRAST AND THE CT RESULTS WOULD BE UNDIAGNOSTIC; OR THE PATIENT WOULD HAVE TO RECEIVE EXTRA RADIATION TO COMPENSATE FOR THE LACK OF ENOUGH CONTRAST. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12416550 ↗MalfunctionSep 2, 2021Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED AN UNSPECIFIED QUANTITY OF ONE-LINK NON-DEHP MICROBORE CATHETER EXTENSION SETS HAD HEMOLYZING ISSUES. IT WAS FURTHER REPORTED THAT THE PATIENT¿S BLOOD WOULD HEMOLYZE IMMEDIATELY AFTER STARTING THE INTRAVENOUS LINES AND DRAWING THE BLOOD INTO THE TUBING. THIS ISSUE OCCURRED DURING USE AT THE EMERGENCY ROOM. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12416518 ↗MalfunctionMay 19, 2021No Flow; No Flow
IT WAS REPORTED TEN (10) ONE-LINK NON-DEHP MICROBORE CATHETER EXTENSION SETS WOULD NOT FLOW. IT WAS FURTHER REPORTED, "WHEN THE NURSE CONNECTED THE IV CONNECTOR THE ARE NOT ABLE TO DRAW BLOOD AS THE TUBING HAS NO SUCTION TO DRAW UP THE BLOOD AND END UP USING A SYRINGE TO DRAW UP THE BLOOD". THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11850750 ↗MalfunctionOct 17, 2019Device-Device Incompatibility
16 CASES OF THE BAXTER ONE-LINK AND SMITHS MEDICAL JELCO IV CATHETER SEPARATING DURING POWER INJECTION REQUIRED FOR CT SCAN. THIS HAS LEAD TO THE PATIENTS BEING REQUIRED TO RECEIVE REINJECTION AND REIMAGING. APPEARS TO BE SECONDARY TO THE TWO DEVICES NOT BEING COMPATIBLE WITH ONE ANOTHER. MANUFACTURER RESPONSE FOR CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS, JELCO (PER SITE REPORTER). INSERVICES HAVE BEEN PROVIDED TO STAFF BY BAXTER AND SMITH'S MEDICAL REGARDING THE USE OF THESE TWO DEVICES TOGETHER WITH LESS THAN OPTIMAL RESULTS. MANUFACTURER RESPONSE FOR SET, ADMINISTRATION, INTRAVASCULAR, ONE-LINK (PER SITE REPORTER). BOTH BAXTER AND SMITH'S MEDICAL HAVE PROVIDED SEVERAL INSERVICES IN AN ATTEMPT TO ADDRESS ANY EDUCATION ISSUES WHICH MIGHT BE AT HAND. INSERVICES HAVE NOT BEEN AS EFFECTIVE AS PLANNED AND THESE TWO DEVICES HAVE CONTINUED TO PRODUCE ISSUES DURING POWER INJECTIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9202758 ↗MalfunctionJul 10, 2019Device Damaged Prior to Use
EXTENSION CATHETER FOUND TO BE IN TWO PIECES PRIOR TO PACKAGING BEING OPENED. PRODUCT WAS DAMAGED IN THE MIDDLE OF THE CATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8775585 ↗MalfunctionApr 20, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A ONE-LINK CATHETER EXTENSION SET DISCONNECTED. THE CATHETER AND TUBING CAME APART DURING IV CONTRAST DYE INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7446801 ↗MalfunctionApr 18, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED A ONE-LINK CATHETER EXTENSION SET DISCONNECTED. THE CATHETER AND TUBING CAME APART DURING IV CONTRAST DYE INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7441649 ↗MalfunctionApr 18, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED A ONE-LINK CATHETER EXTENSION SET DISCONNECTED. THE CATHETER AND TUBING CAME APART DURING IV CONTRAST DYE INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7441389 ↗MalfunctionApr 18, 2018Disconnection; Disconnection
IT WAS REPORTED THAT A ONE-LINK CATHETER EXTENSION SET EXPERIENCED A CONNECTION ISSUE WHICH RESULTED IN A LEAK. THE CATHETER AND TUBING BECAME DISCONNECTED DURING INTRAVENOUS INFILTRATION OF CONTRAST DYE TO THE PATIENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7439853 ↗MalfunctionApr 18, 2018Disconnection; Disconnection
IT WAS REPORTED THAT A ONE-LINK CATHETER EXTENSION SET EXPERIENCED A CONNECTION ISSUE WHICH RESULTED IN A LEAK. THE CATHETER AND TUBING BECAME DISCONNECTED DURING INTRAVENOUS INFILTRATION OF CONTRAST DYE TO THE PATIENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7439851 ↗MalfunctionApr 18, 2018Disconnection; Fluid/Blood Leak; Disconnection; Fluid/Blood Leak
IT WAS REPORTED THAT AN EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR AND TUBING CAME APART DURING INTRAVENOUS CONTRAST DYE INFILTRATION. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7439599 ↗MalfunctionMar 22, 2018Detachment Of Device Component
IV CATHETER AND TUBING CAME APART DURING IV CONTRAST DYE INFILTRATION (SMITHS MEDICAL IV CATHETER AND BAXTER ONE-LINK EXTENSION SET).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7361408 ↗MalfunctionJan 19, 2018Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A PATIENT¿S ONE-LINK EXTENSION SET LEAKED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7206394 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K132734. One submission often covers many models, so check the product and lot details in each record.
- Z-1102-2021 · Dec 21, 2020 · TerminatedBaxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K - Product Usage: intended
- Z-0132-2016 · May 21, 2015 · TerminatedONE-LINK NON-DEHP Y-TYPE MICROBORE CATHETER EXTENSION SET NEEDLE-FREE IV CONNECTOR WITH NEUTRAL FLUID DISPLACEMENT
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page