CLEARLINK 2R8858Recalls & reported problems
Baxter Healthcare Corporation · Paclitaxel Set, Polyethylene Lined Tubing, Non-DEHP Pump Segment, 0.2 Micron Downstream Filter, Luer Activated Valve, Male Luer Lock Adapter, 107" 2.7m approx., 10 drops/mL approx. · Class 2
All models of CLEARLINK (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Baxter ClearLink IV set has 129 adverse event reports on record, with no deaths and no injuries. The reports mostly describe fluid leaking from the set, and some mention breaks, air in the line, or no flow. Most say no clinical signs were reported. There is 1 open recall, started in August 2025, because the IV sets may leak.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 129
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI 00085412565774), received between Jul 2, 2019 and Mar 12, 2026 — classified as 129 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (109 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 29, 2025Class IIOpen, Classified
IV sets may leak.
Z-0203-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak109
- Break9
- Air/Gas in Device5
- No Flow5
- Detachment of Device or Device Component3
- Crack2
- Device Alarm System2
- Disconnection2
Patient problems
- No Clinical Signs, Symptoms or Conditions127
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 12, 2026No Flow
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET DID NOT RUN ONCE SPIKED WITH TAXOL. THE ISSUE WAS NOTICED DURING PATIENT USE. THE DEVICE WAS PRIMED WITH SALINE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24578952 ↗MalfunctionMar 11, 2026Fluid/Blood Leak; Air/Gas in Device
IT WAS REPORTED THAT A CLEARLINK SYSTEM, PE-LINED SOLUTION SET LEAKED FROM AN UNSPECIFIED LOCATION. THE FLUID AND AIR WERE FREELY FLOWING/EXCHANGING IN TUBING EVEN WITH THE ROLLER CLAMP LOCKED AND A NON BAXTER FEMALE ADAPTOR ATTACHED AT THE END. THIS OCCURRED AFTER PRIMING WITH FAM TRASTUZUMAB DERUXTECAN. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24569355 ↗MalfunctionMar 11, 2026Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM, PE-LINED SOLUTION SET LEAKED FROM AN UNSPECIFIED LOCATION. THIS DEVICE CONTAINED DOCETAXEL. THIS OCCURRED AFTER PRIMING WHEN THE CHEMOTHERAPY WAS ADDED TO THE BAG. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24569245 ↗MalfunctionFeb 25, 2026No Flow
IT WAS REPORTED THAT AN UNSPECIFIED AMOUNT OF CLEARLINK SYSTEM PE-LINED SOLUTION SETS HAD DIFFICULTY RUNNING IVIG (INTRAVENOUS IMMUNOGLOBULIN) AND ALBUMIN. A PUMP WAS USED IN CONJUNCTION WITH THE SET. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24448802 ↗MalfunctionDec 31, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AT LEAST ONE CLEARLINK, PACLITAXEL SET LEAKED FROM THE BOTTOM OF THE DRIP CHAMBER. THIS WAS OBSERVED AT THE END OF THE TAXOL INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23934177 ↗MalfunctionDec 5, 2025Illegible Information; Illegible Information
IT WAS REPORTED THAT THE WRITING ON THE PRIMARY PACKAGE LABEL OF UNSPECIFIED UNITS OF CLEARLINK SYSTEM, PE-LINED SOLUTION SETS WAS SMEARING. THIS OCCURRED WHILE THE PRODUCTS WERE WIPED WITH ISOPROPYL ALCOHOL. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23725785 ↗MalfunctionOct 21, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK PACLITAXEL SET WAS LEAKING FROM BETWEEN THE FILTER AND THE SEGMENT OF THE TUBING LOADED INTO THE PUMP. THE LEAK WAS OBSERVED WHILE UNSPECIFIED MEDICATION WAS INFUSING. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23347781 ↗MalfunctionSep 18, 2025No Flow; No Flow
IT WAS REPORTED THAT A CLEARLINK SYSTEM PE-LINED SOLUTION SET HAD NO FLOW THROUGH THE TUBING. THIS WAS OBSERVED WHEN THE TUBING WAS USED ON THE IV (INTRAVENOUS) PUMP PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23089963 ↗MalfunctionAug 6, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK SYSTEM PE-LINED SOLUTION SET LEAKED FROM THE WHITE BAND. THIS WAS OBSERVED WHILE THE SET WAS HANGING. THE DEVICE CONTAINED UNSPECIFIED CHEMO. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22728832 ↗MalfunctionJul 31, 2025Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK, PACLITAXEL SET HAD A HOLE. THIS OCCURRED DURING PRIMING THE SET. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22680870 ↗MalfunctionMay 15, 2025Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET LEAKED AND SEPARATED AT THE BOTTOM OF MICRON FILTER WHERE IT CONNECTS TO REMAINING LINE. THIS OCCURRED DURING USE. THE DEVICE CONTAINED 20 MG DEXAMETHASONE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22030831 ↗MalfunctionJan 9, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT THE FILTER OF TWO (2) CLEARLINK PACLITAXEL SETS LEAKED AT THE FILTER. THERE WAS A CRACK/BREAK IN THE FILTER COMPONENT. THE DEVICES CONTAINED ETOPOSIDE SOLUTION. THIS OCCURRED WHILE INFUSING ETOPOSIDE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21114177 ↗MalfunctionJan 8, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT THE FILTER OF A CLEARLINK PACLITAXEL SET WAS ALTERED FROM THE DRUG (ETOPOSIDE) USED. THE FILTER OF THE DEVICE WAS DAMAGED BY THE CHEMO. THIS WAS OBSERVED DURING FINAL INSPECTION AFTER LINE INSTALLATION ON THE MINI BAG PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21106703 ↗MalfunctionNov 12, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK PACLITAXEL SET LEAKED FROM THE TUBING. THIS OCCURRED WHILE TITRATING INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20667839 ↗MalfunctionNov 12, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET LEAKED RIGHT UNDER THE DRIP CHAMBER, "WHERE THE LINE CONNECTS". THIS OCCURRED WHILE PRIMING THE TUBING WITH SALINE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20667291 ↗MalfunctionNov 12, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET LEAKED UNDERNEATH THE DRIP CHAMBER. THIS OCCURRED WHILE PRIMING THE TUBING WITH SALINE. A NEW SET WAS USED TO CONTINUE PREPARATION WITH NO FURTHER ISSUES. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20665636 ↗MalfunctionNov 1, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET WAS LEAKING FROM RIGHT UNDER THE DRIP CHAMBER WHERE THE TUBING CONNECTS. THE LEAK WAS DISCOVERED WHILE PRIMING THE SET WITH SALINE SOLUTION. TO RESOLVE THIS ISSUE, A NEW SET WAS PRIMED WITH NO ISSUES NOTED. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20592203 ↗MalfunctionAug 15, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT THE TUBING OF A CLEARLINK, PACLITAXEL SET WAS DAMAGED; THE SET WAS CUT. THIS WAS OBSERVED DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19995741 ↗MalfunctionJul 23, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET LEAKED FROM THE FILTER. DURING A PACLITAXEL INFUSION THE NURSE WAS PERFORMING BLOOD RETURN CHECKS AND NOTED APPROXIMATELY 30 ML OF CLEAR LIQUID UNDER CHEMOTHERAPY BAG. THE INFUSION WAS STOPPED WITH 149 ML/98.6 MG REMAINING OF 400 MG PACLITAXEL BAG. THE PATIENT CONFIRMED THEY HAD NOT EXITED THE BED NOR HAD ANYONE WALKED THROUGH THE SPILL AREA. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19810938 ↗MalfunctionJul 22, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK PACLITAXEL SET LEAKED FROM THE FILTER. THE TUBING WAS NOTED TO BE CONTINUALLY/SLOWLY DRIPPING CHEMOTHERAPY AT THE FILTER SITE. THIS OCCURRED DURING PATIENT INFUSION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19799954 ↗MalfunctionJun 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET LEAKED AT THE BASE OF THE DRIP CHAMBER, AT THE JOINT WHERE IT MEETS THE TUBING. THIS WAS NOTICED DURING PRIMING IN THE HOOD. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19514718 ↗MalfunctionMay 21, 2024Filling Problem; Filling Problem
IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF UNDERINFUSIONS OCCURRED DURING THE USE OF CLEARLINK, PACLITAXEL SETS DURING INFUSION. IT WAS OBSERVED THAT THERE WERE MULTIPLE INCIDENTS OF TAXOL INFUSING TOO SLOW AND RUNNING 30 MINUTES TO 1 HOUR LONGER THAN EXPECTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19367116 ↗MalfunctionMay 14, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET LEAKED AT THE FILTER DURING PATIENT INFUSION. THE DEVICE WAS FILLED WITH EPCOH (ETOPOSIDE, PREDNISONE, VINCRISTINE, CYCLOPHOSPHAMIDE, AND DOXORUBICIN) SOLUTION. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19310565 ↗MalfunctionMay 2, 2024Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A CLEARLINK PACLITAXEL SET LEAKED FROM THE DRIP CHAMBER. THIS WAS OBSERVED DURING PATIENT INFUSION WITH A CHEMOTHERAPY DRUG. THE TREATMENT WAS STOPPED TO RESOLVE THE EVENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19234569 ↗MalfunctionMar 14, 2024Break; Fluid/Blood Leak; Protective Measures Problem; Break; Fluid/Blood Leak; Protective Measures Problem
IT WAS REPORTED THAT A CLEARLINK, PACLITAXEL SET LEAKED FROM "ONE OF THE SPLICES" (PUMP SEGMENT/BUSHING); THERE WAS A SMALL OPENING IN THE SPLICED AREA. THIS WAS IDENTIFIED DURING PATIENT INFUSION WITH TAXOL ON AN UNSPECIFIED PUMP. THE ISSUE WAS FURTHER DESCRIBED AS, "THE TAXOL STARTING SPRAYING OUT OF THE SPLICED AREA ABOVE THE CLAMP" AND THE PUMPED ALARMED AN AIR IN LINE ALARM. A NEW SET WAS USED TO CONTINUE TREATMENT AND A SPILL KIT USED TO CLEAN UP THE LEAK. THE NURSE WAS EXPOSED TO THE MEDICATION ON THEIR HANDS; HOWEVER, THE NURSE WASHED THEIR HANDS WITH NO FURTHER ISSUES NOTED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18904637 ↗
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page