Skip to content

ONE-LINK 7N8332KRecalls & reported problems

Baxter International Inc. · One-Link Non-DEHP 3 Lead Microbore Catheter Extension Set Needle-free IV Connector with Neutral Fluid Displacement Male Luer Lock Adapter · Class 2

All models of ONE-LINK (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00085412599625), received between Mar 10, 2023 and May 29, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “No Flow” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K132734); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Flow2
  • Air/Gas in Device1
  • Crack1
  • Detachment of Device or Device Component1
  • Fluid/Blood Leak1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 29, 2025Crack; Crack

    IT WAS REPORTED THAT THE HUB OF A ONE-LINK 3 LEAD MICROBORE CATHETER EXTENSION SET WAS CRACKED. THE TUBING HEAD WAS ALSO CRACKED. THIS WAS OBSERVED DURING SETUP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22110360 ↗
  • MalfunctionApr 25, 2025Obstruction of Flow; Air/Gas in Device; Obstruction of Flow; Air/Gas in Device

    IT WAS REPORTED THAT AN UNSPECIFIED QUANTITY OF ONE-LINK NON-DEHP 3 LEAD MICROBORE CATHETER EXTENSION SETS HAD AIR ENTERING THE LINES; FURTHER DESCRIBED AS "INTERRUPTED LINES". THIS WAS OBSERVED DURING PATIENT USE IN THE NEONATAL INTENSIVE CARE UNIT (NICU). ADDITIONALLY, THE EVENT WAS FURTHER DESCRIBED AS DEVICES NOT HAVING "NEGATIVE PRESSURE" WHICH RESULTED IN THE LINES BACKING UP; IF THE TO KEEP OPEN (TKO) PRESSURE WAS NOT HIGH ENOUGH, THERE WAS CLOTTING AS WELL. THE CUSTOMER WAS WORKING WITH STAFF TO ENSURE PROPER TIGHTENING OF THE CONNECTION TO THE INTRAVENOUS (IV) TUBING. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21913662 ↗
  • MalfunctionApr 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A ONE-LINK NON-DEHP 3 LEAD MICROBORE CATHETER EXTENSION SET HAD A SEPARATION BETWEEN THE TUBING AND THE MALE LUER LOCK ADAPTER. THIS OCCURRED WHILE PATIENT WAS CONNECTED TO 3 UNSPECIFIED CONTINUOUS INFUSIONS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21893849 ↗
  • MalfunctionMar 28, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT TWO (2) ONE-LINK 3 LEAD MICROBORE CATHETER EXTENSION SETS LEAKED FROM AN UNSPECIFIED LOCATION. THIS OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21720373 ↗
  • MalfunctionMar 13, 2023No Flow; No Flow

    IT WAS REPORTED THAT THE PORT (CONNECTOR) OF AN ONE-LINK NON-DEHP 3 LEAD MICROBORE CATHETER EXTENSION SET WAS OCCLUDED. THE EVENT OCCURRED DURING AN UNSPECIFIED PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16530356 ↗
  • MalfunctionMar 10, 2023No Flow; No Flow

    IT WAS REPORTED THAT THE PORT OF THE TRIFUSE OF TWO (2) ONE-LINK NON-DEHP 3 LEAD MICROBORE CATHETER EXTENSION SETS WERE OCCLUDED. IT WAS NOT SPECIFIED WHEN IN THE PROCESS STEP THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16522344 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K132734. One submission often covers many models, so check the product and lot details in each record.

  • Z-1102-2021 · Dec 21, 2020 · Terminated
    Baxter ONE-LINK Non-DEHP Y-Type Microbore Catheter Extension Set, Bonded Needle-free IV Connector, Neutral Fluid Displacement, 5.7 " (14 cm), REF 7N8370K - Product Usage: intended
  • Z-0132-2016 · May 21, 2015 · Terminated
    ONE-LINK NON-DEHP Y-TYPE MICROBORE CATHETER EXTENSION SET NEEDLE-FREE IV CONNECTOR WITH NEUTRAL FLUID DISPLACEMENT

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page