Peri-Guard PG0814Recalls & reported problems
Synovis Life Technologies, Inc. · PERI-GUARD is prepared from bovine pericardium cross-linked with glutaraldehyde, and treated with 1 molar sodium hydroxide for a minimum of 60 minutes at 20-25 degrees C. PERI-GUARD is terminally sterilized using gamma irradiation and packaged within a double-pouch system. The contents of the unopened, undamaged outer pouch are sterile. · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K221029, K223052); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Feb 28, 2024Open, Classified
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
Z-1379-2024 · Root cause: Labeling Change Control · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K221029, K223052. One submission often covers many models, so check the product and lot details in each record.
- Z-1380-2024 · Feb 28, 2024 · Open, ClassifiedBaxter Supple Peri-Guard Repair Patch (new product codes): a) SPG0404, b) SPG0406, c) SPG0608, d) SPG0814, e) SPG1016
- Z-0665-2024 · Dec 6, 2023 · Open, ClassifiedPeri-Guard Repair Patch, Product Codes: a) PC1016NBIO: 10X16CMUS, b) PC1225NBIO: 12X25CMUS, c) PC0404NBIO: 4X4CMUS, d) PC0608NBIO: 6X8CMUS, e) PC0814NBIO: 8X14CMUS
- Z-0668-2024 · Dec 6, 2023 · Open, ClassifiedPERI-GUARD SUPPLE, Product Codes: a) SPG1016: 10X16 CM US, TS; b) SPG0404: 4 X 4 CM US, TS; c) SPG0608: 6 X 8 CM US, TS; d) SPG0814: 8 X 14 CM US, TS; e) PC1016SNBIO: 10X
- Z-0664-2024 · Dec 6, 2023 · Open, ClassifiedPeri-Guard Repair Patch, Product Codes: a) PG1016 : 10 X 16 CMUS, TS b) PG1225: 12 X 25 CMUS, TS c) PG0404: 4 X 4 CMUS, TS d) PG0608: 6 X 8CMUS, TS e) PG0814: 8 X 14 CMUS, TS
Other Mesh, surgical devices
- MEDPOR TITAN 81032Stryker Leibinger GmbH & Co. KG · Mesh, surgical1 recall1 report
- Peri-Guard PC1016NSynovis Life Technologies, Inc. · Mesh, surgical1 recall1 report
- Supple Peri-Guard SPG0404Synovis Life Technologies, Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-403-001Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-300-006Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-300-019Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-001-009Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
- SurgiMend® 606-004-105Tei Biosciences Inc. · Mesh, surgical1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.