Why it was recalled
A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
FDA root cause category: Labeling Change Control
What the company is doing
Baxter issued a "SAFETY ALERT" to its consignees on 02/28/2024 via USPS first class mail. the notice explained the problem, potential hazard involved with off-label use of the product and the actions to be taken. The firm requested that the notice be forwarded all affected facilities or departments with institutions. Distributors are directed to notify their customers. For questions, contact Baxter Healthcare Center for Service at 888-229-0001, between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday.
Recall details
- Recall number
- Z-1379-2024
- Status
- Open, Classified
- Initiated by the company
- Feb 28, 2024
- Posted by the FDA
- Mar 26, 2024
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 2413 units
- Product code
- FTM
- FDA submissions
- K223052
- Distribution
- US Distribution to states of: CT, VA, TX, PA and IL; and OUS (foreign) countries of: Austria, Finland, Germany, Greece, Ireland, Italy, Poland, Spain, Sweden, Switzerland, and United Kingdom.
Product description
Baxter Peri-Guard Repair Patch (new product codes): a) PG0404, b) PG0608, c) PG0814, d) PG1016, e) PG1225
Code information (models, lots, serial numbers)
a) PG0404, UDI/DI 00085412818405; b) PG0608, UDI/DI 00085412818443; c) PG0814, UDI/DI 00085412818597; d) PG1016, UDI/DI 00085412818610; e) PG1225, UDI/DI 00085412818658; ALL LOTS WITHIN EXPIRY
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- Peri-Guard PG0404Synovis Life Technologies, Inc. · Mesh, surgical, collagen, diaphragmatic hernia1 recall0 reports
- Peri-Guard PG0608Synovis Life Technologies, Inc. · Mesh, surgical, collagen, thoracic, chest wall reconstruction1 recall0 reports
- Peri-Guard PG0814Synovis Life Technologies, Inc. · Mesh, surgical1 recall0 reports
- Peri-Guard PG1016Synovis Life Technologies, Inc. · Mesh, surgical, collagen, diaphragmatic hernia1 recall0 reports
- Peri-Guard PG1225Synovis Life Technologies, Inc. · Mesh, surgical, collagen, diaphragmatic hernia1 recall0 reports
Product lines
Other devices under the same FDA submission
5 devices cleared under K223052 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Supple Peri-Guard SPG0608Synovis Life Technologies, Inc. · Mesh, surgical1 recall0 reports
- Supple Peri-Guard SPG0814Synovis Life Technologies, Inc. · Intracardiac patch or pledget, biologically derived1 recall0 reports
- Supple Peri-Guard SPG1016Synovis Life Technologies, Inc. · Intracardiac patch or pledget, biologically derived1 recall0 reports
- Supple Peri-Guard SPG0404Synovis Life Technologies, Inc. · Mesh, surgical1 recall0 reports
- Supple Peri-Guard SPG0406Synovis Life Technologies, Inc. · Intracardiac patch or pledget, biologically derived1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 94152).
- Z-1377-2024 · Open, ClassifiedBaxter Peri-Guard Repair Patch (legacy product codes): a) PC0404NBIO, b) PC0608NBIO, c) PC0814NBIO, d) PC1016NBIO
- Z-1378-2024 · Open, ClassifiedBaxter Supple Peri-Guard Repair Patch (legacy product codes): a) PC0404SNBIO, b) PC0608NSBIO, c) PC0814SNBIO, d) PC1016SNBIO
- Z-1380-2024 · Open, ClassifiedBaxter Supple Peri-Guard Repair Patch (new product codes): a) SPG0404, b) SPG0406, c) SPG0608, d) SPG0814, e) SPG1016
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.