Roscoe Medical O2C5LRecalls & reported problems
Drive Devilbiss Healthcare · TRUAIRE-5 O2 STATIONARY CONCENTRATOR, 5 LITER · Class 2
All models of Roscoe Medical (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00092237623152), received between Sep 19, 2024 and Nov 15, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 9, 2024Class IOpen, Classified
Oxygen concentrator may melt or catch on fire during use.
Z-0795-2025 · Root cause: Component change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Fire1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 15, 2024Fire; Fire
O2 CONCENTRATOR CAUGHT ON FIRE. THEY PUT THE FIRE OUT AND THERE WERE NO INJURIES. PROPERTY MANAGEMENT FOR THE COMPANY STATED UNIT CAUGHT FIRE IN A RESIDENT'S ROOM IN OUR SKILLED NURSING CARE. THE CONTROL AREA OF THE DEVICE IS WHERE THE FIRE STARTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20694910 ↗MalfunctionSep 28, 2024No Apparent Adverse Event; No Apparent Adverse Event
A CONCENTRATOR CAUGHT ON FIRE IN A PATIENT'S BATHROOM. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED WITH THE INFORMATION CURRENTLY AVAILABLE. THE MANUFACTURER CONTINUES TO REQUEST THE RETURN OF THE DEVICE FOR FURTHER ANALYSIS. IT SHOULD BE NOTED THAT THE MANUFACTURER HAS CONDUCTED INDEPENDENT ELECTRICAL TESTING ON A DEVICE FROM A PROXIMAL LOT AND MONTH, (B)(6) (LOT# JA2311, MANUFACTURE DATE: NOVEMBER 24, 2023), AND DETERMINED THE DEVICE TO MEET ELECTRICAL STANDARDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20328761 ↗MalfunctionSep 19, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
WE HAD A CONCENTRATOR THAT CAUGHT ON FIRE IN A PATIENT'S HOME ON (B)(6) 2024. THE PATIENT HAD NO INJURIES AND DID NOT REQUIRE MEDICAL ATTENTION. THE PATIENT WAS IN A DIFFERENT ROOM FROM THE CONCENTRATOR. THE CONCENTRATOR WAS LOCATED IN THEIR BATHROOM, BECAUSE THEY DID NOT LIKE THE NOISE FROM THE CONCENTRATOR. THE DAUGHTER USED A FIRE EXTINGUISHER TO PUT THE OUT THE FIRE FROM THE CONCENTRATOR. THE FIRE DEPARTMENT WAS CALLED OUT BY THE DAUGHTER AFTER THE FIRE WAS EXTINGUISHED. PER THE DAUGHTER, THE CONCENTRATOR WAS TAKEN BY THE INSURANCE COMPANY. SHE WILL ASK IF SHE CAN GET IT BACK AND LET ME KNOW. THE FAMILY CALLED ME BACK AND THE FIRE INVESTIGATOR HAS THE CONCENTRATOR. THEY JUST SPOKE WITH HIM, AND THEY WILL NOT BE RELEASING THE CONCENTRATOR TO ANYONE. HE SAID THEY WOULD BE HAPPY TO DO A DUAL INVESTIGATION WITH THE MANUFACTURER/COMPANY. THIS EVENT IS NOT CONSIDERED A MANDATORY REPORTABLE EVENT AS THERE WAS NO REPORT OF DEATH OR SERIOUS INJURY. THE PRESENCE OR ABSENCE OF A MALFUNCTION CANNOT BE DETERMINED WITH THE INFORMATION CURRENTLY AVAILABLE. HOWEVER, THE MANUFACTURER HAS DECIDED TO MAKE THE FDA AWARE OF THE INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20256380 ↗MalfunctionSep 19, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
AT APPROXIMATELY 11:47PM ON JUNE 25, 2024, AT (B)(6), A FIRE ALARM SOUNDED ON THE 2 SOUTH UNIT IN ROOM 221. THE STAFF RESPONDED AND SAW HEAT AND SMOKE COMING FROM THE OXYGEN CONCENTRATOR, TRUAIRE-5, IN THE ROOM. THE STAFF EVACUATED THE PATIENTS (2) IN ROOM 221 AND THE PATIENTS IN THE ROOMS IN CLOSE VICINITY TO THE DINING ROOM BEYOND THE SMOKE BARRIER DOORS. THE PATIENTS FROM ROOM 221 WERE ASSESSED AFTER THEY WERE EVACUATED AND THERE WERE NO ADVERSE EFFECTS RELATED TO SMOKE INHALATION. AT THE SAME TIME ANOTHER STAFF MEMBER EXTINGUISHED THE FIRE USING THE FIRE EXTINGUISHER. THE FIRE DEPARTMENT RESPONDED AND ARRIVED AT APPROXIMATELY 11:54PM TO CONTAIN AND DISSIPATE THE SMOKE. THERE WAS NO NOTED DAMAGE TO THE STRUCTURE OF THE ROOM, THE TRUAIRE-5 OXYGEN CONCENTRATOR WAS REMOVED FROM THE ROOM, AND THE PATIENTS WERE ABLE TO RETURN TO THEIR ROOMS WITHIN 24 HOURS. THE PATIENT USING THE TRUAIRE-5 AS WELL AS THE OTHER PATIENTS THAT WERE EVACUATED WERE EVALUATED BY A PULMONARY NP THE NEXT DAY, (B)(6) 2024 AND IT WAS NOTED THAT THE PATIENT IN ROOM 221 RESPIRATORY STATUS WAS UNCHANGED AND HER OXYGEN DEMAND WAS STABLE, AND THE OTHER PATIENTS WERE NOTED TO HAVE RESPIRATORY/MENTAL STATUS AT BASELIN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20256039 ↗
Other Generator, Oxygen, Portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page