Skip to content

SimplyGo Mini SimplyGo Mini, USRecalls & reported problems

Respironics, Inc. · SimplyGo Mini Device, USA · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
1
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00606959032385), received between Dec 9, 2021 and Mar 11, 2026 — classified as 1 death, 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Failure to Power Up” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 7, 2025Open, Classified

    Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.

    Z-0481-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Failure to Power Up2
  • Adverse Event Without Identified Device or Use Problem1
  • Low Readings1
  • Overheating of Device1
  • Sparking1
  • Temperature Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Heart Failure/Congestive Heart Failure1
  • Insufficient Information1
  • Respiratory Failure1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2026Sparking

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING THAT A SIMPLYGO MINI DEVICE. THE PATIENT STATED HER DEVICE DOES NOT SEEM TO BE CHARGING, JUST TRIED PLUG THE DEVICE IN AND IT SPARKED. THERE WAS NO REPORT OF MEDICAL INTERVENTION. THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION AT THIS TIME. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE FILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24570979 ↗
  • InjuryOct 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER BECAME AWARE OF AN ALLEGATION RELATED TO A SIMPLYGO MINI DEVICE. THE MANUFACTURER RECEIVED INFORMATION ALLEGING CONGESTIVE HEART FAILURE. THERE WAS NO REPORT OF MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION CAN BE REQUESTED AT THIS TIME. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Heart Failure/Congestive Heart Failure

    MAUDE report 20539525 ↗
  • MalfunctionNov 8, 2023Overheating of Device; Failure to Power Up; Overheating of Device; Failure to Power Up

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING THE PATIENT'S DEVICE IS NOT TURNING ON OR OFF AND THE DEVICE IS GETTING HOT WHEN IT IS PLUGGED IN TO CHARGE. THERE IS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. NO MEDICAL INTERVENTION WAS REQUIRED BY THE PATIENT. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18094323 ↗
  • DeathAug 21, 2023Failure to Power Up; Failure to Power Up

    A CONSUMER REPORTED THE PORTABLE OXYGEN CONCENTRATOR WAS NO LONGER FUNCTIONING AFTER 6 WEEKS OF USE. AT THE TIME OF THE INITIAL REPORT, THE PATIENT HAD GONE WITHOUT USE OF THE DEVICE FOR TWO DAYS DUE TO THE DEVICE NOT FUNCTIONING. A MANAGER AT THE STORE WHERE THE DEVICE WAS PURCHASED ALLEGES THE DEVICE WAS DEFECTIVE AND THE MANAGER WOULD NOT EXCHANGE THE DEVICE. DURING AN ATTEMPT TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, AND AN ATTEMPT TO RETRIEVE THE DEVICE, AN ADULT CHILD OF THE PATIENT REPORTED THE PATIENT PASSED AWAY TWO MONTHS AGO. THERE WAS NO MEDICAL INTERVENTION SPECIFIED. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION. THE USER MANUAL STATES THE DEVICE IS NOT INTENDED FOR LIFE SUPPORT. WHERE THE PRESCRIBING HEALTH CARE PROFESSIONAL HAS DETERMINED THAT AN INTERRUPTION IN THE SUPPLY OF OXYGEN, FOR ANY REASON, MAY HAVE SERIOUS CONSEQUENCES TO THE USER, AN ALTERNATE SOURCE OF OXYGEN SHOULD BE AVAILABLE FOR IMMEDIATE USE. THE USER MANUAL ALSO STATES THE PHILIPS RESPIRONICS SIMPLYGO MINI PORTABLE OXYGEN CONCENTRATOR IS FOR PRESCRIPTION USE BY PATIENTS REQUIRING HIGH CONCENTRATIONS OF OXYGEN ON A SUPPLEMENTAL BASIS.

    Patient problems: Insufficient Information; Respiratory Failure

    MAUDE report 17586519 ↗
  • MalfunctionDec 9, 2021Low Readings; Temperature Problem; Low Readings; Temperature Problem

    THE MANUFACTURER RECEIVED INFORMATION ALLEGING A SIMPLYGO MINI OXYGEN CONCENTRATOR HAVE EVIDENCE OF THERMAL DAMAGE TO THE ENCLOSURE. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE HAS YET TO BE RETURNED FOR EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER HAS COMPLETED THE INVESTIGATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12970926 ↗

Other Generator, oxygen, portable devices

All Generator, oxygen, portable devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.