FreeStyle Precision Neo 71579Recalls & reported problems

Abbott Diabetes Care Inc · FreeStyle Precision Neo Blood Glucose Test Strips · Class 2

All models of FreeStyle Precision Neo (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00093815715795), received between Apr 27, 2017 and Dec 24, 2021 — classified as 2 injuries, 11 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings12
  • Device Displays Incorrect Message1
  • Unable to Obtain Readings1

Patient problems

  • No Known Impact Or Consequence To Patient7
  • No Clinical Signs, Symptoms or Conditions4
  • Hyperglycemia2
  • Loss of consciousness2
  • Convulsion/Seizure1
  • Dizziness1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 24, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13084646 ↗
  • InjuryJul 21, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED AN UNSPECIFIED INCORRECT READINGS ISSUE USING ADC FREESTYLE LIBRE 2 READER. ON (B)(6) 2021, AS A RESULT, THE CUSTOMER WAS NOT FEELING WELL AND EXPERIENCED DIZZINESS, PASSED OUT, HAD A SEIZURE, AND A LOSS OF CONSCIOUSNESS. THE CUSTOMER REGAINED CONSCIOUSNESS AND SELF-TREATED WITH UNSPECIFIED MEDICATION AND METFORMIN FOR TREATMENT. HCP CONTACT WAS MADE AND PROVIDED UNSPECIFIED MEDICAL TREATMENT. NO FURTHER INFORMATION WAS PROVIDED. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: Convulsion/Seizure; Dizziness; Hyperglycemia; Loss of consciousness

    MAUDE report 12202884 ↗
  • MalfunctionJun 22, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12038932 ↗
  • MalfunctionJun 22, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12038917 ↗
  • MalfunctionJun 16, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 44 MG/DL AND 302 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12009626 ↗
  • InjuryNov 10, 2020Unable to Obtain Readings; Device Displays Incorrect Message; Unable to Obtain Readings; Device Displays Incorrect Message

    CUSTOMER'S HUSBAND REPORTED BEING UNABLE TO TEST DUE TO RECEIVING AN ERROR 7 ON THE DISPLAY OF HIS WIFE'S ADC BLOOD GLUCOSE METER. CUSTOMER EXPERIENCED LOSS OF CONSCIOUSNESS DUE TO HYPERGLYCEMIA AND RECEIVED UNSPECIFIED THIRD PARTY TREATMENT. NO FURTHER INFORMATION WAS PROVIDED AS THE CALLER REFUSED TO CONTINUE TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: Hyperglycemia; Loss of consciousness

    MAUDE report 10816351 ↗
  • MalfunctionOct 30, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10761565 ↗
  • MalfunctionMay 14, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS FROM THEIR ADC FREESTYLE LIBRE BUILT IN METER. CUSTOMER REPORTED RECEIVING READINGS OF 60 MG/DL, 60 MG/DL, 115 MG/DL AND 207 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10057619 ↗
  • MalfunctionMar 20, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS FROM THEIR ADC FREESTYLE LIBRE BUILT-IN METER. CUSTOMER REPORTED RECEIVING READINGS OF 32 MG/DL AND 230 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8438047 ↗
  • MalfunctionFeb 4, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER WITHIN 10 MINUTES. RESULTS OF 101 MG/DL AND 19 MG/DL WERE PLOTTED ON THE PARKES ERROR GRID. THE RESULTS FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8301859 ↗
  • MalfunctionMay 18, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 28 MG/DL AND 123 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7527746 ↗
  • MalfunctionApr 4, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 464 MG/DL AND 110 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7396407 ↗
  • MalfunctionApr 27, 2017Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CALLER (CUSTOMER¿S MOTHER) REPORTED THE CUSTOMER RECEIVED ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. READINGS OF 475 MG/DL AND "LO" (LESS THAN 20 MG/DL) WERE REPORTEDLY RECEIVED WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6528413 ↗

Other SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER devices

All SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page