FreeStyle Precision Neo 75175Recalls & reported problems
Abbott Diabetes Care Inc · FreeStyle Precision Neo Blood Glucose Monitoring System · Class 2
All models of FreeStyle Precision Neo (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Abbott FreeStyle Precision Neo blood glucose meter is an over-the-counter meter used to check blood sugar. Reports most often describe the meter failing to turn on, and a few mention low readings and low blood sugar episodes. The database holds 243 reports, including no deaths and 3 injuries, with 240 malfunctions. No recalls name this device, and reports fell over the last year (7) compared with the year before (35).
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 243
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 243 adverse event reports that list this exact device (DI 00093815751755), received between Dec 28, 2016 and Apr 9, 2026 — classified as 3 injuries, 240 malfunctions.
The most frequently reported specific device problem is “Failure to Power Up” (238 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Power Up238
- Incorrect, Inadequate or Imprecise Result or Readings2
- Low Readings2
- Device Displays Incorrect Message1
- Unable to Obtain Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions166
- No Known Impact Or Consequence To Patient74
- Hypoglycemia2
- Loss of consciousness2
- Convulsion/Seizure1
- Dizziness1
- Shaking/Tremors1
Recent adverse event reports
The 25 most recent of 243 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 9, 2026Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24832229 ↗MalfunctionApr 8, 2026Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24812333 ↗MalfunctionJan 30, 2026Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24222096 ↗MalfunctionNov 24, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23631005 ↗MalfunctionNov 20, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23604048 ↗MalfunctionNov 14, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23551766 ↗MalfunctionNov 12, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23531259 ↗MalfunctionOct 29, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23411700 ↗MalfunctionOct 17, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23324166 ↗MalfunctionSep 19, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23093303 ↗MalfunctionAug 26, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22893978 ↗MalfunctionAug 25, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22883445 ↗MalfunctionAug 18, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22828054 ↗MalfunctionAug 7, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22738797 ↗MalfunctionJul 28, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22628446 ↗MalfunctionJul 14, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22493950 ↗MalfunctionJul 8, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22437265 ↗MalfunctionJul 3, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING A LOW READING FROM THEIR ABBOTT DIABETES CARE DEVICE. CUSTOMER REPORTED A LOW READING OF 45 MG/DL WHICH WAS LOWER THAN A METER READING OF 168 MG/DL. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22393373 ↗MalfunctionJun 26, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22337117 ↗MalfunctionMay 27, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22086366 ↗MalfunctionMay 14, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22017934 ↗MalfunctionApr 30, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21934123 ↗MalfunctionApr 25, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21910711 ↗InjuryApr 21, 2025Low Readings; Low Readings
A LOW GLUCOSE READINGS ISSUE WITH THE ABBOTT DIABETES CARE (ADC) DEVICE WAS REPORTED. THE CUSTOMER RECEIVED AN UNSPECIFIED LOWER GLUCOSE DEVICE RESULTS COMPARED TO UNSPECIFIED GLUCOSE READING OBTAINED ON AN ADC METER. IT IS UNKNOWN WHETHER THE CUSTOMER NOTED A TREND ARROW ON THE DEVICE. AS A RESULT, THE CUSTOMER EXPERIENCED DIZZINESS AND DECIDED TO GO TO THE EMERGENCY ROOM (ER) WHERE THE CUSTOMER HAD A GLUCOSE READING AT 300 MG/DL. HOWEVER, IT WAS NOT CONFIRMED WHETHER THE CUSTOMER RECEIVED THIRD PARTY TREATMENT. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Dizziness
MAUDE report 21873465 ↗MalfunctionMar 12, 2025Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21584921 ↗
Other SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER devices
- OneTouch Verio Flex 023-271LifeScan Europe GmbH · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER3,703 reports
- FreeStyle Libre 2 71953Abbott Diabetes Care Inc · System, test, blood glucose, over the counter2 recalls1 death3,286 reports
- OneTouch Ultra2 021-105LifeScan U.S. LLC · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER1,731 reports
- Omnipod DASH PDM PDM-USA1-D001-MG-USA1Insulet Corporation · System, Test, Blood Glucose, Over The Counter1 recall10 deaths1,129 reports
- Omnipod UST400Insulet Corporation · System, Test, Blood Glucose, Over The Counter12 deaths816 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE METER - NOT FOR RETAIL SALETrividia Health, Inc. · System, Test, Blood Glucose, Over The Counter418 reports
- FreeStyle Freedom Lite 70923-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
- FreeStyle Lite 70829-70Abbott Diabetes Care Inc · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER378 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page