FreeStyle Precision H 80116Recalls & reported problems
Abbott Diabetes Care Inc · FreeStyle Precision H Blood Glucose Test Strips · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00093815801160), received between Feb 27, 2018 and Oct 16, 2018 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K070984); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message1
- No Device Output1
Patient problems
- Loss of consciousness2
- Diaphoresis1
- High Blood Pressure/ Hypertension1
- Hypoglycemia1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 16, 2018No Device Output; No Device Output
A HEALTHCARE PROVIDER REPORTED THAT ON (B)(6) 2018 AT 5PM A CLINIC¿S ADC BLOOD GLUCOSE METER WOULD TURN ON WITH BUTTON PRESS, BUT NOT WITH TEST STRIP INSERTION, THEREBY PREVENTING A READING FROM BEING OBTAINED. IT WAS FURTHER REPORTED THE CUSTOMER/PATIENT BECAME ¿HYPERTENSIVE, DIAPHORETIC¿, AND SUBSEQUENTLY ¿LOST CONSCIOUSNESS¿. PARAMEDICS WERE CALLED AND UPON ARRIVAL RECEIVED A READING OF 20 MG/DL ON THEIR METER AND TRANSPORTED HIM TO A HOSPITAL. CALLER DID NOT KNOW THE SPECIFIC TREATMENT RENDERED BY THE PARAMEDICS OR AT THE HOSPITAL. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: Diaphoresis; High Blood Pressure/ Hypertension; Loss of consciousness
MAUDE report 7971963 ↗InjuryFeb 27, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
A PATROL OFFICER REPORTED THAT ON (B)(6) 2018 AT 11:00 A.M., A MAN WAS FOUND UNCONSCIOUS. THE PATROL OFFICER ATTEMPTED TO USE THE ADC METER TO TEST THE MAN'S GLUCOSE AND WAS UNSUCCESSFUL DUE TO AN ER-3 MESSAGE APPEARING ON THE METER DISPLAY. PARAMEDICS WERE CALLED AND UPON THEIR ARRIVAL AT 11:15 A.M., A BLOOD GLUCOSE RESULT OF 38 MG/DL WAS OBTAINED ON THE PARAMEDIC METER AND THE MAN WAS TREATED WITH D-50 GLUCOSE VIA INTRAVENOUS INFUSION. ONCE HE REGAINED CONSCIOUSNESS, HE WAS PROVIDED SOLID FOOD AND TRANSPORTED TO A HOSPITAL AT 11:40 A.M. THERE WAS NO REPORT OF DEATH OR PERMANENT IMPAIRMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Hypoglycemia; Loss of consciousness
MAUDE report 7301716 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070984. One submission often covers many models, so check the product and lot details in each record.
- Z-1558-2020 · Nov 14, 2019 · TerminatedFreeStyle Precision H Blood Glucose and Ketone Monitoring System, UDI: 00093815711810
- Z-0998-2011 · Dec 22, 2010 · TerminatedAbbott Diabetes Care, Precision Point of Care Blood Glucose Test Strips;
- Z-0994-2011 · Dec 22, 2010 · TerminatedAbbott Diabetes Care Precision Xceed Pro Blood Glucose Test Strips
Other Glucose dehydrogenase, glucose devices
- OneTouch Verio IQ 022-275LifeScan Europe GmbH · Glucose dehydrogenase, glucose6,035 reports
- OneTouch Verio 022-659LifeScan Europe GmbH · Glucose dehydrogenase, glucose4,219 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE TEST STRIPS - 50 Test StripsTrividia Health, Inc. · Glucose Dehydrogenase, Glucose2 deaths2,548 reports
- FreeStyle Lite 70805Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose2,366 reports
- FreeStyle Freedom Lite 70920Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose1,037 reports
- FreeStyle Freedom Lite 70914Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose760 reports
- Accu-Chek Guide 08256292001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose476 reports
- ACCU-CHEK AVIVA PLUS 06908268001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose472 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page