Accu-Chek Guide 08256292001Recalls & reported problems

Roche Diagnostics Operations, Inc. · Accu-Chek Guide 50ct HN 36/cs · Class 2

All models of ACCU-CHEK GUIDE (14)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Accu-Chek Guide blood glucose test system is used to check blood sugar at home. Its 476 adverse event reports mostly describe readings that were incorrect, erratic, or much higher or lower than repeat tests taken minutes apart. There are no deaths and 10 injuries. Reports fell from 56 in the prior 12 months to 15 in the last 12. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
476
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 476 adverse event reports that list this exact device (DI 00365702719101), received between Mar 31, 2020 and Sep 28, 2026 — classified as 10 injuries, 466 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (372 reports).

2 other recalls cover devices cleared under the same FDA submission (K160944); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 11 reportsDec 2023: 3 reportsJan 2024: 6 reportsFeb 2024: 12 reportsMar 2024: 2 reportsApr 2024: 15 reportsMay 2024: 11 reportsJun 2024: 1 reportsJul 2024: 9 reportsAug 2024: 8 reportsSep 2024: 7 reportsOct 2024: 6 reportsNov 2024: 7 reportsDec 2024: 5 reportsJan 2025: 6 reportsFeb 2025: 7 reportsMar 2025: 2 reportsApr 2025: 10 reportsMay 2025: 3 reportsJun 2025: 7 reportsJul 2025: 3 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 3 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 3 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings372
  • Imprecise or Erratic Results55
  • High Test Results41
  • Low Test Results2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Markings/Labelling Problem1
  • Display Difficult to Read1
  • High Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions364
  • No Consequences Or Impact To Patient55
  • Hypoglycemia31
  • Hyperglycemia12
  • Blurred Vision5
  • Dizziness5
  • Fatigue5
  • Shaking/Tremors5

Recent adverse event reports

The 25 most recent of 476 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026High Test Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917613 ↗
  • MalfunctionSep 25, 2026High Test Results

    It was reported the patient received the following results within 15 minutes: 479 mg/dL and 126 mg/dL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897028 ↗
  • MalfunctionMar 19, 2026High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 84 MG/DL (SYSTEM 1) AND 163 MG/DL (SYSTEM 2). THE BLOOD GLUCOSE RESULTS WERE OBTAINED USING THE SAME VIAL OF TEST STRIPS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24641174 ↗
  • MalfunctionFeb 24, 2026Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 300 MG/DL, 215 MG/DL, 439 MG/DL AND 209 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24427540 ↗
  • MalfunctionFeb 12, 2026Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 354 MG/DL AND 119 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24321778 ↗
  • MalfunctionFeb 9, 2026High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 601 MG/DL AND 201 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24291919 ↗
  • MalfunctionJan 30, 2026Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: HI MG/DL AT 9:11AM (WHICH ON THE SYSTEM INDICATES A RESULT IN EXCESS OF 600 MG/DL) 258 MG/DL AT 9:12AM, 222 MG/DL AT 9:13AM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24219426 ↗
  • MalfunctionJan 29, 2026High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 550 MG/DL AND 150 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24209016 ↗
  • MalfunctionJan 20, 2026High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 320 MG/DL (USER'S METER) AND 109 MG/DL (PROFESSIONAL'S METER).

    Patient problems: Hypoglycemia

    MAUDE report 24113811 ↗
  • MalfunctionDec 9, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 400'S MG/DL AND 90'S MG/DL. AND THEN ON A SEPARATE DAY THE USER TESTED AGAIN AND RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 300'S MG/DL AND 90'S MG/DL. THE USER COULD NOT RECALL THE EXACT VALUES FOR EACH SET OF COMPARISONS. THE USER STATED THAT THE READINGS WERE OBTAINED USING THE SAME VIAL OF STRIPS, LOT. 102981, AND THAT THE TEST STRIPS WERE NOT EXPIRED AT THE TIME OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23758316 ↗
  • MalfunctionDec 1, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 524 MG/DL, 181 MG/DL AND 131 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23680063 ↗
  • MalfunctionDec 1, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 493 MG/DL AND 199 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23679753 ↗
  • MalfunctionNov 21, 2025High Test Results; High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 192 MG/DL AND 100 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23619292 ↗
  • InjuryNov 11, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THAT THE PATIENT RECEIVED THE FOLLOWING RESULTS OVER A PERIOD OF SEVERAL HOURS DURING THE INCIDENT DAY: 104 MG/DL, 120 MG/DL, 89 MG/DL, 125 MG/DL AND 130 MG/DL. AT AN UNKNOWN TIME ON THAT DAY, THE PATIENT WAS NOT FEELING WELL, EXPERIENCED SYMPTOMS LIKE DIZZINESS, AND PASSED OUT. THE PATIENT´S WIFE AND DAUGHTER CALLED THE PARAMEDICS WHO GAVE THE PATIENT AN UNKNOWN AMOUNT OF INSULIN AS TREATMENT. THE PATIENT WAS STABILIZED AND NOT TAKEN TO THE HOSPITAL.

    Patient problems: Dizziness; Loss of consciousness

    MAUDE report 23520106 ↗
  • MalfunctionNov 5, 2025Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem

    THE USER REPORTED THE USE BY DATE LISTED ON THE TEST STRIP VIAL WAS 2026-05-20. THE CORRECT EXPIRATION DATE FOR LOT. 104526 IS 2025-05-20.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23477604 ↗
  • MalfunctionOct 24, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 222 MG/DL @ 01:00PM ON (B)(6) 2025, 98 MG/DL @ 01:02PM ON (B)(6) 2025. 123 MG/DL @ 03:00PM ON (B)(6) 2025, 379 MG/DL @ 03:02PM ON (B)(6) 2025. 341 MG/DL @ 03:00PM ON (B)(6) 2025, 141 MG/DL @ 03:02PM ON (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23378415 ↗
  • MalfunctionOct 16, 2025High Test Results; High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 237 MG/DL, 213 MG/DL AND 514 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23307689 ↗
  • MalfunctionOct 15, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 200 MG/DL, 76 MG/DL AND 69 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23288467 ↗
  • MalfunctionSep 15, 2025High Test Results; High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: ON (B)(6) 2025 A READING OF 800 MG/DL AT 6:30 AM. ON (B)(6) 2025 A READING OF 140 MG/DL AT 6:34 AM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23051650 ↗
  • MalfunctionAug 25, 2025High Test Results; High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 534 MG/DL AND 123 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22882938 ↗
  • MalfunctionAug 18, 2025High Test Results; High Test Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 473 MG/DL AT 5:00 AM. 465 MG/DL AT 5:05 AM. 158 MG/DL AT 5:08 AM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22832187 ↗
  • MalfunctionJul 24, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 107 MG/DL, 400 MG/DL, AND 300 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22585774 ↗
  • MalfunctionJul 8, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: HI (= HIGHER THAN 600 MG/DL), 175 MG/DL AND 554 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22435322 ↗
  • MalfunctionJul 3, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 29 MG/DL AT 5:00AM . 288 MG/DL AT 5:01AM . 131 MG/DL AT 5:02AM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22399297 ↗
  • MalfunctionJun 23, 2025Imprecise or Erratic Results; Imprecise or Erratic Results

    IT WAS REPORTED THE PATIENT RECEIVED THE FOLLOWING RESULTS WITHIN 15 MINUTES: 362 MG/DL AND 167 MG/DL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22307113 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160944. One submission often covers many models, so check the product and lot details in each record.

  • Z-2174-2024 · May 13, 2024 · Open, Classified
    Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectivenes
  • Z-0747-2020 · Sep 30, 2019 · Terminated
    Blood Glucose Monitoring System, Model Number 07400870001 ACCU-CHEK GUIDE SAMPLE KIT MG/DL - Product Usage: The Accu-Chek Guide Blood Glucose Monitoring System is intended to quant

Other Glucose Dehydrogenase, Glucose devices

All Glucose Dehydrogenase, Glucose devices →