Skip to content

Invictus 15004-105-100Recalls & reported problems

Alphatec Spine, Inc. · Polyaxial Screw, 10.5 mm x 100 mm · Class 2 · Implant

All models of Invictus (96)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00190376136998), received between Mar 16, 2020 and Mar 16, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Noise, Audible1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 16, 2020Break; Noise, Audible

    UPON INSERTING SCREWS INTO BONE, THE SCREW DRIVER MADE A CRACKING NOISE AS THE SCREW ADVANCED INTO HARD BONE. THE WINGED TIP OF THE SCREW DRIVER SHEARED OFF AND THE DRIVER FELL APART. THIS CAUSED THE SCREW SHANK TO STRIP AND THE BUSHING TO BREAK, SUBSEQUENTLY RESULTING IN CUTTING OFF THE HEAD OF THE SCREW TO EXTRACT THE SHANK SEPARATELY. THE SECOND DRIVER BROKE WHILE INSERTING THE OTHER S2AI SCREW. THE SURGEON ATTEMPT TO HELICOPTER THE SCREW DOWN THE REST OF THE WAY WITH THE COUNTER TORQUE AND THE SCREW HEAD SHEARED OFF. THE DETACHED TULIP/SCREW HEAD WAS REMOVED BUT THE THREADED SHANK REMAINS ENTRAPPED WITHIN S2AI (SACRAL ALAR ILIAC). THE EVENT CAUSED OVER A 30 MINUTE DELAY IN THE CASE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9835213 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page