Skip to content

LineSider LS-N6SA7550Recalls & reported problems

Integrity Implants Inc. · 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated - 7.5mm x 50mm · Class 2 · Implant

All models of LineSider (71)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00818613027891), received between Apr 11, 2022 and Apr 11, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).

2 FDA recall records name this device's identifier.

2 other recalls cover devices cleared under the same FDA submission (K203367); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jan 27, 2022Class IITerminated

    Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

    Z-0419-2023 · Root cause: Device Design · Recall details · FDA record ↗

  2. Jan 27, 2022Class IITerminated

    Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

    Z-0418-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration or Expulsion of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 11, 2022Migration or Expulsion of Device; Migration or Expulsion of Device

    ON 03FEB22 ACCELUS WAS INFORMED OF A COMPLAINT ON A LINESIDER NON-MODULAR STANDARD SCREW. IT WAS ORIGINALLY REPORTED BY THE REP THAT AN INTRAOPERATIVE SCREW HEAD SEPARATION HAD OCCURRED DURING THE PROCEDURE. AT THE TIME, THIS WAS ALL THE INFORMATION THAT WAS AVAILABLE. NEW INFORMATION WAS RECEIVED ON 11MAR22. PER THE COMPLAINT ENTRY FORM RECEIVED ON 11MAR22, THE REPORTED WELD FAILURE OCCURRED DURING A REVISION PROCEDURE. THE INITIAL SURGERY WAS A 4 LEVEL PLIF AND FIXATION PROCEDURE THAT TOOK PLACE ON (B)(6) 2021. DURING A POSTOPERATIVE FOLLOW UP VISIT IT WAS NOTED THAT ONE OF CEPHALAD SCREWS WAS LOOSENING, AS A RESULT, THE SURGEON OPTED TO REVISE THE PROCEDURE. THE REVISION PROCEDURE TOOK PLACE ON (B)(6) 2022. IT WAS NOTED DURING THE REVISION PROCEDURE THAT ONE OF THE CAUDAL SCREWS HAD A SEPARATED TULIP HEAD. IT IS UNKNOWN IF THIS SEPARATION OCCURRED POSTOPERATIVELY OR DURING THE REVISION PROCEDURE. THE PROCEDURE WAS COMPLETED BY REMOVING AND REPLACING THE LOOSE CEPHALAD SCREW AND THE SEPARATED CAUDAL SCREW. THE REPORTED WELD FAILURE RESULTED IN A 10 MINUTE DELAY TO THE REVISION PROCEDURE. NO ADDITIONAL PATIENT HARM WAS REPORTED AS A RESULT OF THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14074144 ↗

FDA clearance and classification

Cleared through Special 510(k) K203367 on Dec 3, 2020 (decision: Substantially Equivalent). Submitted by Integrity Implants, Inc. as “LineSider™ Spinal System”. The FDA review took 17 days. FDA record ↗

The FDA regulates product code NKB (thoracolumbosacral pedicle screw system) under 21 CFR 888.3070 as a Class II device. 21 CFR 888.3070 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: Intended to provide immobilization and stabilization of spinal segments in thoracic, lumbar, and sacral spine as an adjunct to fusion.

Recalls under the same FDA submission

These recalls involve devices cleared under K203367. One submission often covers many models, so check the product and lot details in each record.

  • Z-0421-2023 · Aug 3, 2022 · Terminated
    LineSider Spinal System pedicle screws, 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Model Number LS-N6SG6545. Thoracolumbosacral pedicle screw system; Spinal i
  • Z-0420-2023 · Jan 27, 2022 · Terminated
    LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-

Common questions about the LineSider LS-N6SA7550

Has the LineSider LS-N6SA7550 been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 closed (terminated) by the FDA. The most recent, Z-0419-2023 (Class II), began on Jan 27, 2022: Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the LineSider LS-N6SA7550?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 11, 2022. That report lists device problem “Migration or Expulsion of Device” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LineSider LS-N6SA7550?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LineSider LS-N6SA7550 FDA approved?

It was cleared through Special 510(k) K203367 on Dec 3, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the LineSider LS-N6SA7550 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the LineSider LS-N6SA7550, and is it still sold?

The FDA's device database lists Integrity Implants Inc. as the company responsible for it (brand LineSider, model LS-N6SA7550). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Thoracolumbosacral Pedicle Screw System devices

All Thoracolumbosacral Pedicle Screw System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page