Invictus 17139-045-NRecalls & reported problems

Alphatec Spine, Inc. · 4.5mm Tap, Cannulated, Navigation · Class 2

All models of Invictus (96)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00190376319209), received between May 6, 2021 and Dec 3, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Separation3
  • Fracture1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 3, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE TAP BROKE OFF INTO THE LEFT L3 PEDICLE DURING A CASE. THE TIP WAS RETRIEVED. A SCREW WAS NOT ABLE TO BE PLACED IN THE PEDICLE DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23709464 ↗
  • InjuryOct 22, 2025Break; Material Separation; Break; Material Separation

    THE TAP REPORTEDLY BROKE AT THE THREADS WHEN TAPPING INTO HARD BONE. THE PIECE WAS RETRIEVED FROM THE PATIENT. PERCUTANEOUS SCREW PLACEMENT WAS NOT POSSIBLE DUE TO THE PATIENT'S BONE CONDITION AND THEREFORE, THE CASE WAS ABORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23358978 ↗
  • InjuryMay 15, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE TAP BROKE OFF INSIDE OF THE PEDICLE DURING A SPINAL PROCEDURE. THE TIP WAS UNABLE TO BE RETRIEVED FROM THE PATIENT AND REMAINS IN THE PEDICLE. THIS IS REPORT 1 OF 3.

    Patient problems: Foreign Body In Patient

    MAUDE report 22029286 ↗
  • MalfunctionJul 26, 2023Fracture; Fracture

    IT WAS REPORTED THE TIP OF A 4.5MM TAP, CANNULATED, NAVIGATION IS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17397532 ↗
  • InjuryMay 6, 2021Material Fragmentation; Material Fragmentation

    THE DOCTOR ATTEMPTED TO BACK OUT THE TAP AND HEARD A SNAP. THE DOCTOR LOOKED DOWN AND REALIZED THE TAP HAD SNAPPED 20MM ABOVE THE THREADS. THIS CAUSED A DELAY IN SURGERY OF GREATER THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11784834 ↗

Other Orthopedic stereotaxic instrument devices