REAL INTELLIGENCE ROB10013Recalls & reported problems
Smith & Nephew, Inc. · REAL INTELLIGENCE ROBOTIC DRILL · Class 2
All models of REAL INTELLIGENCE (32)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Smith & Nephew Real Intelligence robotic drill has 531 adverse event reports, with no deaths and 211 injuries. Many reports describe the drill disconnecting, stopping mid-surgery or showing error messages during knee surgery, with surgeons often switching to a manual procedure, and a few patients had cuts. One recall is open, begun in November 2023, because surgeons may be unable to clear drill disconnection error messages. Reports fell from 79 in the prior 12 months to 37 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 527
- Deaths reported
- 0
- Injuries reported
- 211
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 527 adverse event reports that list this exact device (DI 00885556757321), received between Aug 17, 2020 and Apr 5, 2026 — classified as 211 injuries, 316 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (258 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Nov 13, 2023Class IIOpen, Classified
Possibility that the user is unable to resolve drill disconnection error messages.
Z-0708-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Connection Problem258
- Physical Resistance/Sticking118
- Intermittent Continuity82
- Unintended Application Program Shut Down61
- Unintended Movement35
- Break32
- Retraction Problem32
- Mechanics Altered30
Patient problems
- No Clinical Signs, Symptoms or Conditions495
- Laceration(s)22
- No Known Impact Or Consequence To Patient11
- Unspecified Tissue Injury6
- Injury4
- Failure of Implant2
- No Patient Involvement2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 25 most recent of 527 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 5, 2026Break; Overheating of Device; Unintended Movement; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, ONE (1) REAL INTELLIGENCE ROBOTIC DRILL BEGAN TO OVERHEAT AND THEN STOPPED WORKING. THE PROCEDURE WAS CONVERTED TO A MANUAL PROCEDURE BECAUSE A BACKUP WAS NOT AVAILABLE. THE PROCEDURE WAS COMPLETED WITHOUT CONSEQUENCES FOR THE PATIENT. AFTER THE SURGERY THE DRILL DIAGNOSTIC COULD NOT BE PERFORMED, BECAUSE THE BUR WAS NOT ENGAGING ONTO THE HANDPIECE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE (MANUAL PROCEDURE). NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24787072 ↗MalfunctionMar 31, 2026Material Deformation; Mechanics Altered
IT WAS REPORTED THAT, DURING CORI ASSISTED SURGERY, WHILE THE BONE CUTTING STEP, IT WAS OBSERVED THAT THE BUR HEAD WAS UNABLE TO CUT THE BONE EFFECTIVELY. AN ATTEMPT WAS MADE TO REPLACE IT WITH THE ROBOTIC DRILL EXTRA. DURING THIS PROCESS, IT WAS FOUND THAT THE ROBOTIC DRILL ATTACHMENT, BECAME STUCK. THEREFORE, THE INSTRUMENT WAS TAKEN TO THE ADMIN SYSTEM AND THE LONG ATTACHMENT ASSEMBLING FUNCTION WAS USED TO UNLOCK THE COLLET. AS THE SURGEON COULD NOT WAIT FOR FURTHER TROUBLESHOOTING, IT WAS DECIDED TO CONVERT TO A CONVENTIONAL SURGICAL TECHNIQUE TO CONTINUE THE PROCEDURE. AFTER THE INSTRUMENT WAS SUCCESSFULLY REMOVED, IT WAS OBSERVED THAT THE DISTAL END OF THE ROBOTIC DRILL WAS DAMAGED. THIS WAS CASE 5. THE CASES 1¿4 PROGRESSED NORMALLY WITHOUT ANY ISSUES. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24750515 ↗MalfunctionMar 30, 2026Material Integrity Problem; Mechanical Jam; Noise, Audible; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A TKA, THE REAL INTELLIGENCE ROBOTIC DRILL GOT STUCK AND WOULD NO LONGER WORK. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, WITH A MANUAL PROCEDURE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24724133 ↗MalfunctionMar 25, 2026Break; Moisture or Humidity Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING A REAL INTELLIGENCE CORI ASSISTED UKA SURGERY, THE REAL INTELLIGENCE ROBOTIC DRILL DID NOT DRILL. THE DEVICE WAS TESTED AFTER SURGERY, IN DIAGNOSTIC MODE AND THE MOTOR DID NOT RUN CONTINUOUSLY BUT REPEATEDLY STOPPED ROTATING. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A MANUAL PROCEDURE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690440 ↗MalfunctionMar 23, 2026Mechanical Jam
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, ONE (1) REAL INTELLIGENCE ROBOTIC DRILL WOULD NOT UNLOCK TO ATTACH THE BURR. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE (SWITCH TO MANUAL PROCEDURE). NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24669057 ↗MalfunctionMar 7, 2026Break; Connection Problem; Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT, DURING A REAL INTELLIGENCE CORI ASSISTED UKR SURGERY, THE REAL INTELLIGENCE ROBOTIC DRILL STOPPED DURING MILLING. THE ERROR "TIME-OUT IN SYSTEM, THE HANDPIECE HAS BEEN DEACTIVATED" OCCURRED. THE HANDPIECE WAS RECONNECTED BUT THE ERROR OCCURRED AGAIN. AFTER A SECOND ATTEMPT IT WAS RECOVERED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A MANUAL PROCEDURE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24541064 ↗MalfunctionMar 6, 2026Connection Problem; Unintended Movement
IT WAS REPORTED THAT, DURING A REAL INTELLIGENCE CORI ASSISTED UKR SURGERY, THE REAL INTELLIGENCE ROBOTIC DRILL HAD A COMMUNICATION ERROR, IT COULD NOT LOCK THE BUR WHEN INSERTED. THERE WAS ALSO AN ERROR WHEN SELECTING A PATIENT UNABLE TO START THE PROCEDURE. THE PATCH (HANDPIECE_SETTING_CORRUPTION FIX.SH) HAS BEEN UPDATED, THE ROBOT DRILL DIAGNOSTICS HAVE BEEN PASSED, AND SO FAR THE UNIT HAS BEEN WORKING NORMALLY. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A MANUAL PROCEDURE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24536145 ↗MalfunctionMar 5, 2026Loose or Intermittent Connection; Connection Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING AN UKR SURGERY, WHEN BURRING, THE REAL INTELLIGENCE ROBOTIC DRILL KEPT STOPPING AND SLOWING DOWN. ALSO, AN ERROR MESSAGE WAS POPPING UP NONSTOP, SAYING THAT THE DRILL WAS DISCONNECTED. WHEN ATTEMPTED TO SWAP OUT THE DRILLS, GOT AN INTERNAL ERROR MESSAGE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE (MANUAL INSTRUMENTATION). NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24520138 ↗MalfunctionMar 3, 2026Break; Failure to Calibrate; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA, AT CALIBRATION STEP, THE BURR OF ONE (1) REAL INTELLIGENCE ROBOTIC DRILL WAS NOT SPINNING. ADDITIONALLY, IT FAILED THE SET MAXIMUM SPEED TEST. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE TO MANUAL PROCEDURE, AS NO BACK UP WAS AVAILABLE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24499663 ↗MalfunctionFeb 10, 2026Failure to Calibrate
IT WAS REPORTED THAT DURING A TKA, THE REAL INTELLIGENCE ROBOTIC DRILL HAD AN INSTALLATION PROBLEM, WHILE THE PATIENT WAS ALREADY UNDER ANESTHESIA. IT COULDN'T BE UNLOCKED EVEN THOUGH THE MOTOR WAS MAKING NOISES. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24310585 ↗MalfunctionFeb 10, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING A TKA, WHILE MAKING THE DISTAL CUT, THE BURR OF A REAL INTELLIGENCE ROBOTIC DRILL KEPT SLIPPING OUT AND HAD TO BE RE-INSERTED. THIS OCCURRED THREE TIMES, AND EVEN AFTER THE TWO CLICKS WERE CONFIRMED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24310536 ↗MalfunctionFeb 9, 2026Retraction Problem; Connection Problem
IT WAS REPORTED THAT, DURING A REAL INTELLIGENCE CORI ASSISTED TKA SURGERY, AT THE POINT OF STARTING BONE REMOVAL A REAL INTELLIGENCE ROBOTIC DRILL ERROR MESSAGE APPEARED ON SCREEN. DURING DIAGNOSTIC RETRACTION TIME DID NOT PASS THE TEST. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, CHANGING TO A MANUAL PROCEDURE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24299645 ↗MalfunctionFeb 5, 2026Mechanical Jam
IT WAS REPORTED THAT DURING A REAL INTELLIGENCE CORI ASSISTED TKA SURGERY, IT WAS NOTICED THAT THE INTERNAL CARRIAGE OF A REAL INTELLIGENCE ROBOTIC DRILL WAS COMPLETELY BLOCKED, AND IT WAS IMPOSSIBLE TO ENGAGE THE BUR. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, CHANGING TO A MANUAL PROCEDURE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24268496 ↗MalfunctionFeb 2, 2026Moisture Damage; Retraction Problem; Connection Problem
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, ONE (1) REAL INTELLIGENCE ROBOTIC DRILL HAD AN ERROR "SYSTEM TIME OUT, ROBOTIC DEVICE IS DEACTIVATED" AND IT WAS UNABLE TO CONTINUE WITH MILLING PROCESS. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE (MANUAL PROCEDURE). NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24241382 ↗MalfunctionFeb 2, 2026Connection Problem
IT WAS REPORTED THAT DURING A TKA, THE REAL INTELLIGENCE ROBOTIC DRILL WOULDN'T CONNECT AND UNLOCK. THE PROCEDURE WAS RESUMED, AFTER A 5-MINUTE DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24233161 ↗MalfunctionJan 14, 2026Unstable
IT WAS REPORTED THAT, DURING CORI ASSISTED TKA SURGERY, THE BURR WOULD NOT SEAT PROPERLY INSIDE THE REAL INTELLIGENCE ROBOTIC DRILL. ON THE CUT SCREEN, THE BURR RESECTION WAS BURRING ~1 MORE THAN PLANNED. SWAPPED OUT THE LONG ATTACHMENT, BUT THE BURR STILL DID NOT SEAT PROPERLY. THEN THEY SWAPPED OUT THE HANDPIECE, AND THE PROBLEM WAS RESOLVED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Laceration(s)
MAUDE report 24075515 ↗MalfunctionJan 13, 2026Positioning Failure
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKP SURGERY, ONE (1) REAL INTELLIGENCE ROBOTIC DRILL DID NOT ACCEPT THE BURR AND KEPT PUSHING IT OUT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE (MANUAL PROCEDURE). NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24053696 ↗MalfunctionJan 7, 2026Degraded; Unstable; Detachment of Device or Device Component; Mechanics Altered
IT WAS REPORTED THAT DURING A TKA WITH THE PATIENT ALREADY UNDER ANESTHESIA, THE REAL INTELLIGENCE ROBOTIC DRILL MANAGE TO CALIBRATE BUT FAILED DURING BURRING, EVEN AFTER TROUBLESHOOTING. IT COULDN'T EVEN GRAB THE BUR CORRECTLY AND IT SUNK INTO THE DRILL. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23995515 ↗MalfunctionDec 29, 2025Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, A SILVER TRACKER ARRAY WOULD GET STUCK ON ONE (1) REAL INTELLIGENCE ROBOTIC DRILL AND NOT ROTATE, THEN THE ARRAY WAS DIFFICULT TO REMOVE. A NEW TRACKER WAS TRIED WITH THE SAME REPORTED DRILL AND ENCOUNTERED THE SAME PROBLEM. THE PROCEDURE WAS PERFORMED, AFTER A SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23919158 ↗MalfunctionDec 26, 2025Positioning Failure; Positioning Failure
IT WAS REPORTED THAT, DURING CORI ASSISTED TKA SURGERY, THE BURR WOULD NOT CLICK IN THE REAL INTELLIGENCE ROBOTIC DRILL. AN ERROR WAS RECEIVED THAT BURR WAS NOT FULLY SEATED, THEN, IT WAS NOT OPERATED WITH THIS HANDPIECE. AFTER THE OR, CHECKED THE HANDPIECE, AND IT COULD NOT RUN ANY TESTS. ALSO, THE RI ROBOTIC DRILL ATTACHMENT COULD NOT BE TAKEN OFF ANYMORE AND WAS STUCK ON THE HANDPIECE. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23900684 ↗MalfunctionDec 22, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT, DURING A CORI ASSISTED SURGERY, THE SCREEN SHOWED IT WAS UNABLE TO THE CONTINUE WITH THE NEXT STEP. DISCONNECTED AND RECONNECTED THE CABLE, BUT IT STILL UNABLE. THEN REPLACED THE REAL INTELLIGENCE ROBOTIC DRILL´S CABLE, WITH NO RESULT. ALSO, THE RI ROBOTIC DRILL ATTACHMENT WAS MALFUNCTIONING. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23871354 ↗MalfunctionDec 22, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA, ONE (1) REAL INTELLIGENCE ROBOTIC DRILL FAILED TO WORK ON TEST SPIN, A NEW (1) REAL INTELLIGENCE ROBOTIC DRILL WAS TRIED AND DID NOT WORK EITHER. HAD TO RESTART THE DRILLS TO WORK. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE TO MANUAL PROCEDURE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23862884 ↗MalfunctionDec 16, 2025Retraction Problem; Retraction Problem
IT WAS REPORTED THAT, DURING A BILATERAL REAL INTELLIGENCE CORI TKA SURGERY, A REAL INTELLIGENCE ROBOTIC DRILL REGISTERED PROPERLY. HOWEVER, WHEN THEY ATTEMPTED TO BURR THE DISTAL FEMUR CUT, THE BURR WOULD NOT RETRACT AS FAR AS NORMAL (APPROX. 50% AS NORMAL). AFTER THE CASES, THE HANDPIECE PASSED ALL TESTS. PROCEDURE WAS PERFORMED, USING TIBIAL CUTTING BLOCK TO DO THE DISTAL CUT. IT IS UNKNOWN IF THERE WAS ANY DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23816104 ↗InjuryDec 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, DURING A REAL INTELLIGENCE CORI ASSISTED UKR SURGERY, WHEN CUTTING THE FEMUR AND WENT TO THE TIBIA AN FINALIZE THE FEMUR IN THE POSTERIOR CONDYLE, THE BURR RESECTED MORE BONE THAT WAS SUPPOSED TO CUT. STOPPED RESECTION AND MADE AN EXERT WITH CEMENT TO COMPENSATE THE EXTRA RESECTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER ISSUES REPORTED.
Patient problems: Laceration(s)
MAUDE report 23794853 ↗MalfunctionDec 9, 2025Degraded; Failure to Calibrate; Moisture or Humidity Problem; Degraded; Failure to Calibrate; Moisture or Humidity Problem
IT WAS REPORTED THAT DURING AN UKR, THE REAL INTELLIGENCE ROBOTIC DRILL WOULDN'T CALIBRATE. A NOTIFICATION APPEARING SAYING THE DEVICE WAS DEACTIVATED. THE PROCEDURE WAS RESUMED, AFTER A DELAY GREATER THAN 30 MIN, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23764188 ↗
FDA clearance and classification
Cleared through Special 510(k) K201022 on Jun 12, 2020 (decision: Substantially Equivalent). Submitted by Blue Belt Technologies, Inc. as “Real Intelligence Cori (Cori)”. The FDA review took 53 days. FDA record ↗
The FDA regulates product code OLO (orthopedic stereotaxic instrument) under 21 CFR 882.4560 as a Class II device. 21 CFR 882.4560 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the REAL INTELLIGENCE ROB10013
Has the REAL INTELLIGENCE ROB10013 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0708-2024 (Class II), began on Nov 13, 2023: Possibility that the user is unable to resolve drill disconnection error messages. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the REAL INTELLIGENCE ROB10013?
The FDA's MAUDE database lists 527 adverse event reports for this device, received between Aug 17, 2020 and Apr 5, 2026. In those reports, the most common specific device problems are “Connection Problem” (258), “Physical Resistance/Sticking” (118) and “Intermittent Continuity” (82); the most common patient problems are “Laceration(s)” (22), “Unspecified Tissue Injury” (6) and “Injury” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the REAL INTELLIGENCE ROB10013?
None of the reports are classified as deaths, 211 as injuries and 316 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the REAL INTELLIGENCE ROB10013 FDA approved?
It was cleared through Special 510(k) K201022 on Jun 12, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the REAL INTELLIGENCE ROB10013, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand REAL INTELLIGENCE, model ROB10013). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall923 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls598 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument602 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument573 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument560 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
- EXCELSIUS GPS 6143.1001Globus Medical, Inc. · Orthopedic stereotaxic instrument1 death447 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page