Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00190376319223), received between Mar 31, 2022 and May 15, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Detachment of Device or Device Component1
  • Fracture1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 15, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE TAP BROKE OFF INSIDE OF THE PEDICLE DURING A SPINAL PROCEDURE. THE TIP WAS UNABLE TO BE RETRIEVED FROM THE PATIENT AND REMAINS IN THE PEDICLE. THIS IS REPORT 2 OF 3.

    Patient problems: Foreign Body In Patient

    MAUDE report 22029327 ↗
  • InjuryJun 28, 2024Break; Break

    IT WAS REPORTED THAT THE TIP OF THE TAP BROKE OFF INTO THE PATIENT'S BONE WHEN REVERSING DIRECTION TO REMOVE IT FROM THE PEDICLE. THE TIP OF THE TAP REMAINS IN THE PATIENT. THERE WAS A GREATER THAN 30 MINUTE DELAY IN SURGERY DUE TO THE EVENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 19634495 ↗
  • MalfunctionJan 10, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THE TIP OF A TAP BROKE OFF INSIDE THE PEDICLE DURING A SPINAL PROCEDURE. THE SURGEON WAS ABLE TO RETRIEVE THE TIP FROM THE BONE. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED AS RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18489736 ↗
  • MalfunctionMay 5, 2023Fracture; Fracture

    IT WAS REPORTED THE TIP OF A TAP BENT DURING SURGERY. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16882318 ↗
  • InjuryMar 31, 2022Break; Break

    THE SURGEON USED THE TAP, AND IT BROKE OFF IN THE VERTEBRAL BODY. IT WAS REPORTED THE PATIENT HAD VERY HARD BONE. THE SURGEON WAS ABLE TO RETRIEVE THE TAP TIP FROM THE PATIENT; HOWEVER, IT DID CAUSE A DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13958208 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →