Invictus 17950-NRecalls & reported problems
Alphatec Spine, Inc. · MIS T27 Polyaxial Screwdriver, Locking, Navigation · Class 2
All models of Invictus (95)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Alphatec Spine Invictus 17950-N screwdriver is a hand tool used to tighten and adjust screws during minimally invasive spine surgery, including navigated procedures. Most of the 36 reports describe the tool breaking during a case, often with the tip shearing off, and in some instances pieces separating or remaining in the patient. There are 2 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00190376475202), received between Nov 20, 2024 and Sep 24, 2026 — classified as 2 injuries, 31 malfunctions.
The most frequently reported specific device problem is “Break” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break33
- Material Separation23
Patient problems
- No Clinical Signs, Symptoms or Conditions27
- Foreign Body In Patient3
- Insufficient Information3
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Break; Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50886589 ↗MalfunctionSep 24, 2026Break; Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50886419 ↗MalfunctionMay 6, 2026Break
THE TIP OF THE SCREWDRIVER WAS BROKEN OFF WHILE ATTEMPTING TO ADJUST A SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25094360 ↗MalfunctionApr 29, 2026Break
THE TIP OF THE SCREWDRIVER BROKE OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25028608 ↗MalfunctionMar 19, 2026Break
IT WAS REPORTED THAT THE DRIVER BROKE DURING A CASE.
Patient problems: Insufficient Information
MAUDE report 24642401 ↗MalfunctionMar 11, 2026Break; Material Separation
IT WAS REPORTED THAT THE TIP WAS DISCOVERED SHEARED OFF AFTER A CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24572657 ↗MalfunctionMar 5, 2026Break; Material Separation
IT WAS REPORTED THAT THE DRIVER ARRIVED WITH A BROKEN TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24524917 ↗MalfunctionFeb 2, 2026Break; Material Separation
IT WAS REPORTED THAT THE DRIVER HAS A BROKEN TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24238971 ↗MalfunctionJan 29, 2026Break
IT WAS REPORTED THAT THE DRIVER IS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24210395 ↗MalfunctionJan 29, 2026Break
IT WAS REPORTED THAT THE DRIVER IS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24210223 ↗MalfunctionJan 26, 2026Break; Material Separation
IT WAS REPORTED THAT THE TIP OF HTE DRIVER BROKE OFF WHEN INSERTING A SCREW. THE TIP WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24178497 ↗MalfunctionDec 30, 2025Break; Break
IT WAS REPORTED THAT THE DRIVER HAS A BROKEN TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23930985 ↗MalfunctionDec 30, 2025Break; Break
IT WS REPORTED THAT THE DRIVER BROKE DURING SCREW REMOVAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23929952 ↗MalfunctionDec 3, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23707577 ↗MalfunctionNov 21, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING SCREW INSERTION. THE TIP WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23621306 ↗MalfunctionNov 18, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE DRIVER HAS A BROKEN TIP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23585706 ↗MalfunctionNov 4, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23465799 ↗MalfunctionOct 22, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23358474 ↗InjuryOct 16, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF WHEN PLACING A SCREW INTO A PATIENT. THE TIP OF THE DRIVER REMAINS IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 23311560 ↗MalfunctionOct 16, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF WHEN PLACING A SCREW INTO A PATIENT. THE TIP WAS RETRIEVED.
Patient problems: Foreign Body In Patient
MAUDE report 23311513 ↗MalfunctionSep 17, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE DRIVER IS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23076322 ↗MalfunctionAug 25, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS RETRIEVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22884677 ↗MalfunctionAug 20, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS NOT LEFT IN THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22851324 ↗MalfunctionJul 7, 2025Break; Material Separation; Break; Material Separation
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THIS IS REPORT 1 OF 2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22421287 ↗MalfunctionJun 17, 2025Break; Break
IT WAS REPORTED THAT THE DRIVER BROKE WHEN TRYING TO THREAD BETWEEN THE LATERAL SI JOINT SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22235373 ↗
FDA clearance and classification
Cleared through 510(k) K153603 on Apr 29, 2016 (decision: Substantially Equivalent). Submitted by Alphatec Spine, Inc. as “Alphatec Navigation Instruments”. The FDA review took 134 days. FDA record ↗
The FDA regulates product code OLO (orthopedic stereotaxic instrument) under 21 CFR 882.4560 as a Class II device. 21 CFR 882.4560 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Invictus 17950-N
Has the Invictus 17950-N been recalled?
No FDA recall record in our data names this exact device identifier (DI 00190376475202), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Invictus 17950-N?
The FDA's MAUDE database lists 33 adverse event reports for this device, received between Nov 20, 2024 and Sep 24, 2026. In those reports, the most common specific device problems are “Break” (33) and “Material Separation” (23); the most common patient problem is “Foreign Body In Patient” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Invictus 17950-N?
None of the reports are classified as deaths, 2 as injuries and 31 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Invictus 17950-N FDA approved?
It was cleared through 510(k) K153603 on Apr 29, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Invictus 17950-N, and is it still sold?
The FDA's device database lists Alphatec Spine, Inc. as the company responsible for it (brand Invictus, model 17950-N). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall923 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls598 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument602 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall527 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument573 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument560 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page