Alphatec Spine, Inc. · MIS T27 Polyaxial Screwdriver, Locking, Navigation · Class 2

Quick summary

This Alphatec Spine Invictus 17950-N screwdriver is a hand tool used to tighten and adjust screws during minimally invasive spine surgery, including navigated procedures. Most of the 36 reports describe the tool breaking during a case, often with the tip shearing off, and in some instances pieces separating or remaining in the patient. There are 2 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 00190376475202), received between Nov 20, 2024 and Sep 24, 2026 — classified as 2 injuries, 32 malfunctions.

The most frequently reported specific device problem is “Break” (34 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 4 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 3 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 3 reportsFeb 2026: 1 reportsMar 2026: 3 reportsApr 2026: 2 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break34
  • Material Separation24

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • Foreign Body In Patient3
  • Insufficient Information3

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Break; Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50886589 ↗
  • MalfunctionSep 24, 2026Break; Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50886419 ↗
  • MalfunctionMay 6, 2026Break

    THE TIP OF THE SCREWDRIVER WAS BROKEN OFF WHILE ATTEMPTING TO ADJUST A SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25094360 ↗
  • MalfunctionApr 29, 2026Break

    THE TIP OF THE SCREWDRIVER BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25028608 ↗
  • MalfunctionApr 9, 2026Break; Material Separation

    IT WAS REPORTED THAT THE DRIVER HAS A DAMAGED TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24831138 ↗
  • MalfunctionMar 19, 2026Break

    IT WAS REPORTED THAT THE DRIVER BROKE DURING A CASE.

    Patient problems: Insufficient Information

    MAUDE report 24642401 ↗
  • MalfunctionMar 11, 2026Break; Material Separation

    IT WAS REPORTED THAT THE TIP WAS DISCOVERED SHEARED OFF AFTER A CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24572657 ↗
  • MalfunctionMar 5, 2026Break; Material Separation

    IT WAS REPORTED THAT THE DRIVER ARRIVED WITH A BROKEN TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24524917 ↗
  • MalfunctionFeb 2, 2026Break; Material Separation

    IT WAS REPORTED THAT THE DRIVER HAS A BROKEN TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24238971 ↗
  • MalfunctionJan 29, 2026Break

    IT WAS REPORTED THAT THE DRIVER IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24210395 ↗
  • MalfunctionJan 29, 2026Break

    IT WAS REPORTED THAT THE DRIVER IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24210223 ↗
  • MalfunctionJan 26, 2026Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF HTE DRIVER BROKE OFF WHEN INSERTING A SCREW. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24178497 ↗
  • MalfunctionDec 30, 2025Break; Break

    IT WAS REPORTED THAT THE DRIVER HAS A BROKEN TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23930985 ↗
  • MalfunctionDec 30, 2025Break; Break

    IT WS REPORTED THAT THE DRIVER BROKE DURING SCREW REMOVAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23929952 ↗
  • MalfunctionDec 3, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23707577 ↗
  • MalfunctionNov 21, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING SCREW INSERTION. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23621306 ↗
  • MalfunctionNov 18, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE DRIVER HAS A BROKEN TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23585706 ↗
  • MalfunctionNov 4, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23465799 ↗
  • MalfunctionOct 22, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23358474 ↗
  • InjuryOct 16, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF WHEN PLACING A SCREW INTO A PATIENT. THE TIP OF THE DRIVER REMAINS IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23311560 ↗
  • MalfunctionOct 16, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF WHEN PLACING A SCREW INTO A PATIENT. THE TIP WAS RETRIEVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 23311513 ↗
  • MalfunctionSep 17, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE DRIVER IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23076322 ↗
  • MalfunctionAug 25, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22884677 ↗
  • MalfunctionAug 20, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THE TIP WAS NOT LEFT IN THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22851324 ↗
  • MalfunctionJul 7, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE OFF DURING A CASE. THIS IS REPORT 1 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22421287 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →