Calibrate 325-09103010Recalls & reported problems
Alphatec Spine, Inc. · CALIBRATE CCX-O, 9 x 10 x 30mm 10° · Class 2 · Implant
All models of Calibrate (44)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00190376533377), received between Oct 3, 2024 and Jan 24, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Collapse” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Dec 15, 2024Open, Classified
Complaints have been received regarding post-operative implant collapse.
Z-1065-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Collapse2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 24, 2025Collapse; Collapse
ON (B)(6) 2024, A PATIENT UNDERWENT A LUMBAR INTERVERTEBRAL BODY FUSION PROCEDURE. RADIOGRAPHS TAKEN DURING THE 3-MONTH POSTOPERATIVE VISIT INDICATED A PARTIAL EARLY COLLAPSE OF THE CAGE. THE PATIENT IS ASYMPTOMATIC, AND THERE ARE NO PLANS FOR REVISION SURGERY. THE SURGEON WILL CONTINUE TO MONITOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21235567 ↗InjuryOct 3, 2024Collapse; Adverse Event Without Identified Device or Use Problem; Collapse; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024, THE PATIENT HAD A SPINAL PROCEDURE IN WHICH AN EXPANDABLE INTERBODY CAGE WAS PLACED AT THE L3-L4. APPROXIMATELY 2.5 MONTHS POST-SURGERY, THE PATIENT PRESENTED TO THE HOSPITAL WITH PAIN AND CONCERNS ABOUT A POSSIBLE COLLAPSE OF THE CAGE. IMAGING RESULTS WERE INCONCLUSIVE. NO REVISION SURGERY IS PLANNED AT THIS TIME, AND THE SURGEON WILL CONTINUE TO MONITOR THE PATIENT'S CONDITION.
Patient problems: Pain
MAUDE report 20363992 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record