Calibrate recallZ-1065-2025 · Alphatec Spine, Inc. · initiated Dec 15, 2024
Class II Open, Classified All recalls posted in January 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alphatec Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Complaints have been received regarding post-operative implant collapse.
FDA root cause category: Under Investigation by firm
What the company is doing
The recalling firm issued letters dated 12/13/2024 via email on 12/15/2024. The letter provided an attachment listing the specific device part numbers and lot numbers, the reason for recall, the potential risks, a listing of mitigating factors that should be considered when determining patient care, and field action (recall) instructions for the consignee, which included (1) Review of inventory to determine if any of the affected devices are in the consignee's possession. If the device is in the consignee's possession, the consignee is to abstain from use and contact the recalling firm immediately for return instructions; (2) Share the notice with all those who need to be aware within their organization or any organization where the affected device has been transferred; (3) Share the notice and review potential risks with doctors and surgical staff that have implanted the affected device; (4) Fill out the last page of the letter to confirm the notification has been read and the consignee has taken all necessary actions described in the notification. A revised letter dated 12/20/2024 was issued via email 12/23/2024 to update the products that were affected, as the original letter inadvertently omitted three of the part and lot numbers and also it did not indicate some of the implants were contained within convenience kits. The remainder of the letter was similar to the original letter.
Recall details
- Recall number
- Z-1065-2025
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Dec 15, 2024
- Posted by the FDA
- Jan 30, 2025
- Recalling firm
- Alphatec Spine, Inc.
- Quantity
- 1,127 loose implants; 10/24-implant kits; and 4/36-implant kits (1,511 total implants)
- Product code
- MAX
- FDA submissions
- K232504
- Distribution
- US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, IA, IL, IN, MA, MD, MI, MN, NC, NJ, NY, OH, OK, OR, PA, SC, SD, TN, and UT.
Product description
Calibrate CCX Interbody System for spinal fusion procedures: Implants: (1) Calibrate CCX-O, 7 x 10 x 25mm 5 degrees, Part #325-07102505, nonsterile; (2) Calibrate CCX-O, 7 x 10 x 25mm 10 degrees, Part #325-07102510, nonsterile; (3) Calibrate CCX-O, 7 x 10 x 25mm 15 degrees, Part #325-07102515, nonsterile; (4) Calibrate CCX-O, 7 X 10 X 30mm 5 degrees, Part #325-07103005, nonsterile; (5) Calibrate CCX-O, 7 x 10 x 30mm 10 degrees, Part #325-07103010, nonsterile; (6) Calibrate CCX-O, 7 x 10 x 30mm 15 degrees, Part #325-07103015, nonsterile; (7) Calibrate CCX-O, 7 x 10 x 35mm 10 degrees, Part #325-07103510, nonsterile; (8) Calibrate CCX-O, 7 x 10 x 35mm 15 degrees, Part #325-07103515, nonsterile; (9) Calibrate CCX-O, 9 x 10 x 25mm 5 degrees, Part #325-09102505, nonsterile; (10) Calibrate CCX-O, 9 x 10 x 25mm 10 degrees, Part #325-09102510, nonsterile; (11) Calibrate CCX-O, 9 x 10 x 30mm 5 degrees, Part #325-09103005, nonsterile; (12) Calibrate CCX-O, 9 x 10 x 30mm 10 degrees, Part #325-09103010, nonsterile; (13) Calibrate CCX-O, 11 x 10 x 30mm 10 degrees, Part #325-11103010, nonsterile; (14) Calibrate CCX-S, 7 x 10 x 22mm 5 degrees, Part #326-07102205, nonsterile; (15) Calibrate CCX-S, 7 x 10 x 22mm 10 degrees, Part #326-07102210, nonsterile; (16) Calibrate CCX-S, 7 x 10 x 22mm 15 degrees, Part #326-07102215, nonsterile; (17) Calibrate CCX-S, 7 x 10 x 26mm 5 degrees, Part #326-07102605, nonsterile; (18) Calibrate CCX-S, 7 x 10 x 26mm 10 degrees, Part #326-07102610, nonsterile; (19) Calibrate CCX-S, 7 x 10 x 26mm 15 degrees, Part #326-07102615, nonsterile; (20) Calibrate CCX-S, 7 x 10 x 30mm 10 degrees, Part #326-07103010, nonsterile; (21) Calibrate CCX-S, 9 x 10 x 22mm 5 degrees, Part #326-09102205, nonsterile; (22) Calibrate CCX-S, 9 x 10 x 22mm 10 degrees, Part #326-09102210, nonsterile; (23) Calibrate CCX-S, 9 x 10 x 26mm 5 degrees, Part #326-09102605, nonsterile; (24) Calibrate CCX-S, 9 x 10 x 26mm 10 degrees, Part #326-09102610, nonsterile; and (25) Calibrate CCX-S, 11 x 10 x 26mm 10 degrees, Part #326-11102610, nonsterile. Convenience kits containing affected implants: (1) Calibrate CCX Oblique Implants, REF CLCCXOIMPA, containing 24 implants/kit, nonsterile; and (2) Calibrate CCX Straight Implants, REF CLCCXSIMPA, containing 36 implants/kit, nonsterile.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Implants: (1) Lot numbers AP02103, AP02374, AP02426, AP02442, AP02766, AP02967, AP03207, UDI-DI 00190376533308; (2) Lot numbers AP02104, AP02375, AP02427, AP02770, AP02780, AP02833, AP02834, AP02845, AP02846, AP02894, UDI-DI 00190376532967; (3) Lot numbers AP02105, AP02376, AP02782, AP02893, AP02957, AP03109, AP03110, UDI-DI 00190376532974; (4) Lot numbers AP02114, AP02377, AP02428, AP02429, AP02767, AP02919, UDI-DI 00190376532998; (5) Lot numbers AP02106, AP02263, AP02378, AP02430, AP02771, AP02783, AP02836, AP02837, AP02838, AP02839, UDI-DI 00190376533001; (6) Lot numbers AP02107, AP02379, AP02431, AP02781, AP02835, AP02847, AP02848, AP02892. UDI-DI 00190376533018; (7) Lot numbers AP02108, AP02380, AP02432, AP02433, UDI-DI 00190376533049; (8) Lot number AP02144, UDI-DI 00190376533056; (9) Lot numbers AP02109, AP02257, AP02381, AP02434, AP02435, AP02436, AP02768, AP03180, UDI-DI 00190376533322; (10) Lot numbers AP02110, AP02382, AP02772, AP02962, UDI-DI 00190376533339; (11) Lot numbers AP02111, AP02259, AP02383, AP02437, AP02438, UDI-DI 00190376533360; (12) Lot numbers AP02112, AP02384, AP02626, AP02711, AP02773, UDI-DI 00190376533377; (13) Lot numbers AP02113, AP02385, AP02439, AP02440, AP02441, AP02774, AP02921, UDI-DI 00190376532653; (14) Lot numbers AP02115, AP02258, AP02283, AP02301, AP02710, AP02925, AP02971, AP03100, UDI-DI 00190376530277; (15) Lot numbers AP02116, AP02302, AP02871, AP02872, AP02896, UDI-DI 00190376530284; (16) Lot numbers AP02117, AP02303, AP02623, UDI-DI 00190376530291; (17) Lot numbers AP02118, AP02304, AP02775, AP02869, AP02875, UDI-DI 00190376530314; (18) Lot numbers AP02119, AP02305, AP02422, AP02777, AP02785, AP02843, UDI-DI 00190376530321; (19) Lot numbers AP02120, AP02306, AP02624, AP02744, AP02895, AP02922, UDI-DI 00190376530338; (20) Lot numbers AP02121, AP02307, AP02423, UDI-DI 00190376530369; (21) Lot numbers AP02122, AP02312, AP02424, UDI-DI 00190376530758; (22) Lot numbers AP02123, AP02306, AP02308, AP02625, UDI-DI 00190376530765; (23) Lot numbers AP02124, AP02309, AP02425, AP02776, AP02786, AP02844, UDI-DI 00190376530796; (24) Lot numbers AP02125, AP02310, AP02778, AP02784, AP02840, AP02841, AP02842, AP02870, AP03095, UDI-DI 00190376530802; and (25) Lot numbers AP02126, AP02311, AP02779, UDI-DI 00190376531281. Convenience kits containing affected implants: (1) REF CLCCXOIMPA - Serial numbers 2000494, 2000495, 2000496, 2000497, 2000498, 2000499, 2000500, 2000501, 2000502, and 2000503; UDI-DI 00190376553535; and (2) REF CLCCXSIMPA - Serial numbers 2000727, 2000728, 2000730, and 2000731; UDI-DI 00190376553542.
Devices named in this recall
27 devices in the FDA device registry whose identifier appears in the recall's code information.
- Calibrate 325-07103005Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall3 reports
- Calibrate 325-07103015Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall3 reports
- Calibrate 325-09103010Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall2 reports
- Calibrate 326-07102615Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall1 report
- Calibrate 325-07103010Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall1 report
- Calibrate 325-07103510Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall1 report
- Calibrate 326-07102205Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-07102210Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-07102215Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-07102605Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-07102610Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-07103010Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-09102205Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-09102210Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-09102605Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-09102610Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 326-11102610Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 325-11103010Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 325-07102510Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 325-07102515Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 325-07103515Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 325-07102505Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 325-09102505Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 325-09102510Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate 325-09103005Alphatec Spine, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall0 reports
- Calibrate CLCCXOIMPAAlphatec Spine, Inc. · TRAY, SURGICAL, INSTRUMENT1 recall0 reports
- Calibrate CLCCXSIMPAAlphatec Spine, Inc. · TRAY, SURGICAL, INSTRUMENT1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page