Skip to content

DJO SURGICAL 803-95-275Recalls & reported problems

Encore Medical, L.P. · EXPRT Revision Hip, Trial Proximal Body Capture Bolt, 75mm · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00190446156031), received between Sep 20, 2021 and Sep 20, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 20, 2021Loosening of Implant Not Related to Bone-Ingrowth

    INSTRUMENT FAILURE - EXPERT HIP WITH SURGEON IMPLANTED DISTAL STEM. WENT TO TRIAL PROXIMAL BODY. INSERTED TRIAL PROXIMAL BODY AND CRANKED ON THE IMPLANTED TRIAL SCREW DISTAL TO MAKE SURE IT WAS TIGHT. THE TRIAL PROXIMAL BODY CAME LOOSE DURING REDUCTION AND SPUN AROUND THE IMPLANTED DISTAL STEM. ONCE PROXIMAL BODY IS LOOSE, SURGEON LOOSES HIS VERSION DURING AND CAN NOT GAUGE STABILITY OF THE CONSTRUCT.

    Patient problems: Insufficient Information

    MAUDE report 12500469 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page