Skip to content

DJO SURGICAL 540-01-001Recalls & reported problems

Encore Medical, L.P. · DISC ULNA E+ BEARING KIT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a DJO Surgical elbow prosthesis part, a disc ulna bearing kit used in total elbow replacement. The 53 reports mostly describe the implant loosening, becoming unstable or dislocating, wearing out, or infection, leading patients to have revision surgery on the elbow. There were no deaths, but 53 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
55
Deaths reported
0
Injuries reported
55
Recalls naming it
0

Summary

The FDA's MAUDE database contains 55 adverse event reports that list this exact device (DI 00190446254553), received between Jun 2, 2020 and Sep 15, 2026 — classified as 55 injuries.

The most frequently reported specific device problem is “Unstable” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 3 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem33
  • Appropriate Device Problem Term/Code Not Available10
  • Unstable5
  • Degraded2
  • Device Dislodged or Dislocated2
  • Loose or Intermittent Connection2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Naturally Worn2

Patient problems

  • Joint Laxity17
  • Insufficient Information12
  • Unspecified Infection12
  • Failure of Implant7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Joint Dislocation2
  • Fluid Discharge1
  • Implant Pain1

Recent adverse event reports

The 25 most recent of 55 reports, in the reporter's own words as published by the FDA.

  • InjurySep 15, 2026No Apparent Adverse Event

    Patient problems: Insufficient Information

    MAUDE report 50782328 ↗
  • InjurySep 15, 2026Loosening of Implant Not Related to Bone-Ingrowth

    Patient problems: Joint Laxity; Metal Related Pathology

    MAUDE report 50781681 ↗
  • InjuryJun 11, 2026Unstable

    REVISION SURGERY DUE TO LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 25479056 ↗
  • InjuryMay 26, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 25276473 ↗
  • InjuryApr 20, 2026Unstable

    REVISION SURGERY- DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 24933742 ↗
  • InjuryMar 2, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 24492404 ↗
  • InjuryDec 19, 2025Mechanical Problem; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Adverse Event Without Identified Device or Use Problem

    THE HUMERUS NEEDED A POST TRAUMATIC REVISION. IT WAS REVISED. THE NEW POLYETHYLENE IMPLANTS DID NOT FIT.

    Patient problems: Insufficient Information

    MAUDE report 23857984 ↗
  • InjuryDec 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE HUMERUS NEEDED A POST TRAUMATIC REVISION. IT WAS REVISED. THE NEW POLYETHYLENE IMPLANTS DID NOT FIT.

    Patient problems: Insufficient Information

    MAUDE report 23857981 ↗
  • InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - REVISION OF ULNAR AND BEARING COMPONENTS OF DISCOVERY ELBOW, UNKNOWN REASON

    Patient problems: Insufficient Information

    MAUDE report 23700714 ↗
  • InjuryOct 28, 2025Degraded; Degraded

    IT WAS REPORTED THAT THE PATIENT REQUIRED REVISION SURGERY DUE TO POLY WEAR.

    Patient problems: Failure of Implant

    MAUDE report 23410554 ↗
  • InjuryOct 21, 2025Degraded; Degraded

    IT WAS REPORTED THAT THE PATIENT PREVIOUSLY UNDERWENT A TOTAL ELBOW ARTHROPLASTY IN 2007. THE POLYETHYLENE COMPONENT HAD WORN OUT AND WAS SUBSEQUENTLY REPLACED ALONG WITH THE BEARING.

    Patient problems: Insufficient Information

    MAUDE report 23347026 ↗
  • InjuryJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PATIENTS POLY BEARING WAS WORN.

    Patient problems: Joint Laxity

    MAUDE report 22254198 ↗
  • InjuryMar 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO BROKEN ULNA.

    Patient problems: Failure of Implant

    MAUDE report 21612871 ↗
  • InjuryMar 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO BROKEN ULNA.

    Patient problems: Failure of Implant

    MAUDE report 21612863 ↗
  • InjuryFeb 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 21309232 ↗
  • InjuryJan 7, 2025Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 21098247 ↗
  • InjuryNov 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO WEARING.

    Patient problems: Joint Laxity

    MAUDE report 20803016 ↗
  • InjuryNov 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO POLY WEAR.

    Patient problems: Joint Laxity

    MAUDE report 20614716 ↗
  • InjurySep 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO ULNA RING FRACTURED.

    Patient problems: Failure of Implant

    MAUDE report 20153737 ↗
  • InjuryJun 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RIVISION SURGERY - DUE TOFAILED POLY, BROKEN ULNAR EXTRACTOR, BROKEN POLY SWAP INSTRUMENTS.

    Patient problems: Failure of Implant

    MAUDE report 19621226 ↗
  • InjuryJun 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO IMPLANT BROKE

    Patient problems: Unspecified Infection

    MAUDE report 19482491 ↗
  • InjuryMay 13, 2024No Apparent Adverse Event; No Apparent Adverse Event

    REVISION SURGERY - INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 19306682 ↗
  • InjuryApr 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY: DUE TO PAIN AND INFECTION.

    Patient problems: Implant Pain; Unspecified Infection

    MAUDE report 19177640 ↗
  • InjuryMar 11, 2024Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 18882650 ↗
  • InjuryMar 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 18874140 ↗

Other PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED devices

All PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page