Skip to content

DJO SURGICAL 509-02-436Recalls & reported problems

Encore Medical, L.P. · AR, SMALL SOCKET INSERT, 36MM NEUTRAL +4 Eplus · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a shoulder implant made by Encore Medical, a metal-on-plug replacement for the ball-and-socket joint. Most of the 54 reports describe the shoulder joint repeatedly dislocating or feeling unstable, with the plastic socket piece coming apart from the stem, and several patients needing revision surgery; others mention infection, pain, and reduced movement. There are no deaths, and 54 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
54
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 00190446257615), received between Aug 1, 2019 and Apr 22, 2026 — classified as 54 injuries.

The most frequently reported specific device problem is “Unstable” (9 reports).

3 other recalls cover devices cleared under the same FDA submission (K172351); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem39
  • Unstable9
  • Device Dislodged or Dislocated6
  • Appropriate Device Problem Term/Code Not Available5
  • Break1
  • Fracture1
  • Insufficient Device Problem Information1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Joint Dislocation17
  • Joint Laxity17
  • Unspecified Infection9
  • Failure of Implant4
  • Implant Pain4
  • Insufficient Information3
  • Loss of Range of Motion3
  • Medical device site infection2

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 22, 2026Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO THE PATIENT WAS CHRONICALLY DISLOCATING THEIR RSA. AFTER INCISION IT WAS FOUND THE POLY HAD DISASSOCIATED FROM THE STEM.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 24961031 ↗
  • InjuryMar 17, 2026Adverse Event Without Identified Device or Use Problem

    INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 24622295 ↗
  • InjuryFeb 18, 2026Device Dislodged or Dislocated

    REVISION SURGERY - DUE TO DISLOCATION AND DISASSOCIATED.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 24384463 ↗
  • InjuryJan 9, 2026Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO POSSIBLE INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 24031326 ↗
  • InjuryOct 22, 2025Off-Label Use; Off-Label Use

    IT WAS REPORTED THAT THE PATIENT'S REQUIRED REVISION SURGERY DUE TO FAILURE OF GLENOID COMPONENT, DIFFERENT COMPANY.

    Patient problems: Insufficient Information

    MAUDE report 23356077 ↗
  • InjuryOct 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 23297582 ↗
  • InjuryAug 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 22881228 ↗
  • InjuryJun 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 22254133 ↗
  • InjuryJun 16, 2025Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY: DUE TO PATIENT COMPLAINED OF TIGHTNESS IN SHOULDER. PLAN WAS TO REMOVE ANY ADHESIONS AND READDRESS TENSION.

    Patient problems: Loss of Range of Motion

    MAUDE report 22216175 ↗
  • InjuryMay 30, 2025Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY: DUE TO DISSOCIATION.

    Patient problems: Joint Laxity

    MAUDE report 22119868 ↗
  • InjuryMay 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PATIENT PRESENTED TO DOCTOR OFFICE WITH INFLAMMATION IN PREVIOUS REVISED SHOULDER. DOCTOR DID A COMPONENT EXCHANGE WITH THE ADDITION OF A SPACER.

    Patient problems: Swelling/ Edema

    MAUDE report 22007082 ↗
  • InjuryMar 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO SHOULDER PAIN.

    Patient problems: Implant Pain

    MAUDE report 21511120 ↗
  • InjuryJan 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 21120828 ↗
  • InjuryOct 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PATIENT PREVIOUSLY HAD AN APC, STILL EXPERIENCING PAIN DID A POLY AND GLENOSPHERE SWAP.

    Patient problems: Implant Pain

    MAUDE report 20424269 ↗
  • InjuryAug 9, 2024Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY.

    Patient problems: Joint Laxity

    MAUDE report 19956492 ↗
  • InjuryJun 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RIVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 19621245 ↗
  • InjuryJun 8, 2024Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    RIVISION SURGERY - DUE TO DISLOCATION, INSTABILITY AND PAIN.

    Patient problems: Implant Pain; Joint Dislocation; Joint Laxity

    MAUDE report 19495034 ↗
  • InjuryJun 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 19464338 ↗
  • InjuryMay 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DISLOCATION

    Patient problems: Joint Dislocation

    MAUDE report 19348213 ↗
  • InjuryApr 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION & INSTABILITY.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 19102521 ↗
  • InjuryMar 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO DISLOCATION.

    Patient problems: Joint Dislocation

    MAUDE report 18992039 ↗
  • InjuryMar 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO PATIENT COMPLAINED OF TIGHT SHOULDER.

    Patient problems: Loss of Range of Motion

    MAUDE report 18874130 ↗
  • InjuryJan 17, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - MDR - DISLOCATION - PATIENT DISLOCATED SHOULDER AND LEFT IT DISLOCATED FOR MONTHS. POLY INSERT WAS DISLODGED FROM STEM.

    Patient problems: Joint Dislocation

    MAUDE report 18531951 ↗
  • InjuryDec 21, 2023Unstable; Adverse Event Without Identified Device or Use Problem; Unstable; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - DUE TO INSTABILITY AND DISLOCATION.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 18378926 ↗
  • InjuryNov 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - MDR - SIGNIFICANT ADVERSE EVENT/REQUIRED INTERVENTION TO PREVENT IMPAIRMENT/DAMAGE (REVISION).

    Patient problems: Failure of Implant

    MAUDE report 18213158 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K172351. One submission often covers many models, so check the product and lot details in each record.

  • Z-0281-2024 · Sep 25, 2023 · Open, Classified
    RSP Humeral Small Socket Insert Size 32 mm Semi-Constrained +4, e-plus, REF: 509-03-432, orthopedic shoulder implant component.
  • Z-2607-2023 · Aug 4, 2023 · Open, Classified
    DJO Surgical: Altivate Reverse, Humeral Stem, Small Shell, Size 6 x 108mm, REF: 533-06-108
  • Z-0117-2021 · Aug 31, 2020 · Terminated
    AltiVate Reverse Small Socket Insert, 44mm Semi Constrained - E-Plus, Part Number: 509-03-044 - Product Usage: Per IFU, DJO Surgical shoulder devices are intended for treatment of

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page